    Clin Chem. 1992 Oct;38(10):2073-7.

    Automated assay of vitamin B-12 by the Abbott IMx analyzer.

    Source

    Department of Physiological Diagnostics, Abbott Laboratories, Abbott Park, IL 60064.

    Abstract

    A nonisotopic assay of vitamin B-12 in human serum or plasma is described, performed with the Abbott IMx analyzer. The sample is first treated at pH > 12.5 to release bound vitamin B-12 and to convert all forms to cyanocobalamin. Next, the analyte is bound, at lower pH, by vitamin B-12-specific binding protein, immobilized to a solid phase of polymeric microspheres. Detection involves monitoring the activity of the tracer enzyme (alkaline phosphatase) coupled to a derivative of cyanocobalamin. Total assay precision is 7.9% for vitamin B-12 at 200 ng/L, 6.6% at 400 ng/L, and 6.7% at 800 ng/L. Assay sensitivity, calculated as 2 SD from the zero calibrator, is 37 (+/- 9) ng/L. The dynamic range extends to 2000 ng/L. Analytical recovery of 300 and 600 ng/L additions of vitamin B-12 to sera with basal concentrations of 30-400 ng/L was 102.5%. Results of the assay correlated well with those of commercially available radioisotope assays. No interference was observed in specimens from patients with pernicious anemia, chronic or acute myelogenous leukemia, or renal failure. Cross-reactivity with cobinamide (1 g/L) was < 0.00003%. Vitamin B-12 measurements for blood specimens drawn into serum, EDTA, or heparinized plasma-collection tubes agreed within 3%.

    PMID:
    1394992
    [PubMed - indexed for MEDLINE]
    Free full text

      Supplemental Content

      Icon for HighWire Press

      Save items

      loading

      Recent activity

      Your browsing activity is empty.

      Activity recording is turned off.

      Turn recording back on

      See more...
      Write to the Help Desk