    Anal Bioanal Chem. 2003 Sep;377(2):287-92. Epub 2003 Jul 9.

    Automated method for the determination of vitamin D3 hydroxymetabolites in serum.

    Source

    Department of Analytical Chemistry, Annex C-3, Campus of Rabanales, University of Córdoba, 14071 Córdoba, Spain.

    Abstract

    An almost automated method for the determination of hydroxymetabolites of vitamin D(3) (cholecalciferol) in human serum is reported. The method consists of three steps: 1) a batch liquid-liquid extraction step with 2-propanol and hexane, and drying of the extract and reconstitution with phosphate buffer. 2) A cleanup and preconcentration step based on solid-phase extraction using Prospekt equipment, with CN group cartridges and elution with the chromatographic mobile phase. 3) A chromatographic step for individual separation of the target analytes starting with a 90:10 methanol-water mixture, then a linear gradient to obtain 100% methanol; followed by photometric detection. The method provides a linear range between 1.0 and 100 ng mL(-1) for 24,25-(OH)(2) vitamin D(3) and for 25-(OH)(2) vitamin D(3), and between 1.5 and 100 ng mL(-1) for 1,25-(OH) vitamin D(3), with correlation coefficients ranging between 0.993 and 0.987, repeatability between 1.9% and 4.8% and within-laboratory reproducibility between 2.8% and 8.8%.