Palmar plate fixation of AO type C2 fracture of distal radius using a locking compression plate--a biomechanical study in a cadaveric model

J Hand Surg Br. 2003 Jun;28(3):263-6. doi: 10.1016/s0266-7681(03)00011-1.

Abstract

The stability of palmar plate fixation using a locking compression T-plate was compared with that of a conventional palmar T-plate and a dorsal T-plate in a cadaveric model of an AO type C2 fracture of distal radius. The wrist axial load transmission through the radius was tested for each fixation. The results show that, under 100N axial load, the palmar locking compression T-plate restores stability comparable to that of the intact radius, and is superior to conventional palmar or dorsal T-plates.