Format

Send to:

Choose Destination
See comment in PubMed Commons below
J Refract Surg. 2003 Mar-Apr;19(2):94-103.

First clinical results with the femtosecond neodynium-glass laser in refractive surgery.

Author information

  • 1Focus Medical Eye Micro-surgery and Laser Center, Budapest, Hungary. traubfri@axelero.hu

Abstract

PURPOSE:

We evaluated four femtosecond laser intrastromal cutting procedures: creation of a corneal flap for laser in situ keratomileusis (LASIK), tunnel and entry cut for intracorneal ring, corneal flap and removable lens for keratomileusis, and intrastromal ablation for myopia and hyperopia.

METHODS:

A clinical trial using a femtosecond surgical laser (IntraLase Corporation) was performed in partially sighted eyes. Femto-LASIK treatment was performed on 46 eyes up to -14.00 D; 16 patients received intracorneal ring segments (Femto-ICRS); 5 patients each with one highly myopic eye had femtosecond laser keratomileusis (FLK), and 13 patients each with one myopic or hyperopic eye had intrastromal ablation (ISPRK). In Femto-LASIK, excimer laser ablation was done under the flap. In Femto-ICRS, ring segments were introduced into the laser-created channels. In femtosecond laser keratomileusis, a lens-shaped block of stroma was removed manually from under the flap.

RESULTS:

No difference was found between the results obtained with Femto-LASIK and a standard microkeratome. No refractive effects occurred when the created flap was not elevated. In cases of Femto-ICRS and conventional ICRS produced the same refractive results. With Femto-ICRS, no intraoperative complications occurred and visual acuity improved immediately after surgery. In femtosecond laser keratomileusis, high myopia was corrected without using excimer laser ablation; centralization of the treatment area was excellent. In intrastromal ablation, 1 to 2 hours after surgery the corneas were highly transparent; refractive results were stable.

CONCLUSIONS:

Femtosecond lasers can produce precise intrastromal cutting, offering significant safety and other advantages (no razor blades, corneal trauma, partial resections, or sterilization issues) over current techniques.

PMID:
12701713
[PubMed - indexed for MEDLINE]
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Loading ...
    Write to the Help Desk