Regularized higher-order in vivo shimming.

Author information

  • 1Radiological Sciences Laboratory, Department of Radiology, Stanford University, California, USA. dhkim@stanford.edu

Abstract

A regularized algorithm is presented for localized in vivo shimming. The technique uses first- (X,Y,Z), second- (Z(2), ZX, ZY, X(2)-Y(2), XY), and third-order (Z(3)) shim coils, and is robust when applied to arbitrarily-shaped, as well as off-center, regions of interest (ROIs). A single-shot spiral pulse sequence is used for rapid field map acquisition, and a least-squares calculation of the shim currents is performed to minimize the root-mean-square (RMS) value of the B(0) inhomogeneity over a user-selected ROI. The use of a singular value decomposition (SVD) in combination with a regularization algorithm significantly improves the numerical stability of the least-squares fitting procedure. The fully automated shimming package is implemented on a 3 T GE Signa system and its robust performance is demonstrated in phantom and in vivo studies.

Copyright 2002 Wiley-Liss, Inc.