    Clin Chem. 2002 Oct;48(10):1672-80.

    European Concerted Action on Anticoagulation. Evaluation of a method for International Sensitivity Index calibration of two point-of-care prothrombin time (PT) monitoring systems (CoaguChek Mini and TAS PT-NC) with fresh plasmas based on whole-blood equivalent PT.

    Source

    ECAA Central Facility, School of Biological Sciences, The University of Manchester, Manchester M13 9PT, United Kingdom. ecaa@man.ac.uk

    Abstract

    BACKGROUND:

    The International Sensitivity Index (ISI) calibration of whole-blood prothrombin time (PT) monitors for point-of-care testing (POCT) described by Tripodi et al. (Thromb Haemost 1993;70:921-4) has been shown to be dependable but is too complex and demanding. The use of plasma would simplify calibration of whole-blood POCT PT monitors, but important differences may exist between the ISI for whole blood and plasma calibrations.

    METHODS:

    In a 10-center calibration study of two POCT whole-blood monitoring systems (CoaguChek Mini and TAS PT-NC), we characterized the relationship between the log PT for whole blood and fresh plasma with use of single lots of test strips/cards. This relationship (linear) was used to correct the difference between the whole-blood and plasma ISI. The reliability of the correction with different lots of test strips/cards was assessed at three centers. The linear relationship was used to correct the difference in the whole-blood and plasma ISI with four other lots of TAS PT-NC cards and with two additional lots of CoaguChek Mini test strips.

    RESULTS:

    The correction decreased the ISI difference from 13.3% to 0.9% for the TAS PT-NC and from 5.7% to 0.6% for the CoaguChek Mini. In assessments at three centers, which included different lots of test strips/cards, the mean ISI difference was markedly decreased with the TAS PT-NC but not with the CoaguChek Mini, for which the mean ISI difference increased slightly.

    CONCLUSIONS:

    The proposed correction resolves the discrepancy between whole-blood and fresh plasma ISI calibrations with TAS PT-NC test cards. The CoaguChek Mini systems could be calibrated without this correction.

    PMID:
    12324483
    [PubMed - indexed for MEDLINE]