Send to:

Choose Destination
See comment in PubMed Commons below
Cardiovasc Radiat Med. 2001 Oct-Dec;2(4):225-30.

Dose-response study of intracoronary beta-radiation with 32P in balloon- and stent-injured coronary arteries in swine.

Author information

  • 1Methodist DeBakey Heart Center and Baylor College of Medicine, Houston, TX 77030, USA.



A dose-response study was performed in swine to investigate the vascular effects of 32P over a broad range of doses in order to define the therapeutic window of intracoronary radiotherapy (ICR) with 32P.


A total of 131 porcine arteries were subjected to balloon injury or stenting followed by 0-36 Gy of ICR from a centered 32P source wire to 1 mm beyond lumen surface or a sham ICR procedure. Animals were euthanized at 4 weeks, and vessels were harvested for histomorphometry.


In the balloon-injured arteries, doses of 7 and 9 Gy did not impact restenosis. At doses of 14-36 Gy, neointima was markedly reduced, with mild dilatation at the highest dose, 36 Gy. In the stent-injured arteries, the lowest dose of 9 Gy failed to reduce neointimal growth, while 14-26 Gy showed the most favorable response.


ICR with 32P features a broad therapeutic window. Doses of 14-26 Gy to 1 mm beyond lumen surface provided an optimal combination of efficacy and safety. Doses of 7 and 9 Gy were generally ineffective, suggesting a minimum threshold for ICR with 32P to effectively inhibit restenosis.

[PubMed - indexed for MEDLINE]