Biosens Bioelectron. 2002 Apr;17(4):297-302.

Biosensor arrays for simultaneous measurement of glucose, lactate, glutamate, and glutamine.

Author information

  • 1IMTEK, Albert-Ludwigs University Freiburg, Georges Koehler Allee 103, D-79110, Freiburg, Germany. moser@imtek.de

Abstract

For simultaneous measurement of glucose, lactate, glutamine, and glutamate a biosensor array is implemented in a micro flow-system thus giving a microsystem. The microsystem consists of a glass chip with the integrated biosensor array and a bottom part, which comprises a gold counter electrode, a 300 microm thick seal, and electrical interconnection lines. The flow device has a total internal volume of 2.1 or 6 microl when integrated with a mixer on chip. The biosensors with no crosstalking and high long term stability were produced by modifying the electrochemical transducers and utilizing photopatternable enzyme membranes. The use of appropriate miniaturization technology leads to mass producable devices for in vivo and ex vivo applications in whole blood and fermentation broth. Due to a novel glutaminase with an activity optimum in the neutral pH range direct and simultaneous monitoring of glutamine together with glucose, lactate, and glutamate could be performed.