Display Settings:

Format

Send to:

Choose Destination
We are sorry, but NCBI web applications do not support your browser and may not function properly. More information
    Resuscitation. 2001 Aug;50(2):189-204.

    Rapid (0.5 degrees C/min) minimally invasive induction of hypothermia using cold perfluorochemical lung lavage in dogs.

    Source

    Critical Care Research, Inc. 10743 Civic Center Drive, Rancho Cucamonga, CA 91730-3806, USA. sbharris@ix.netcom.com

    Abstract

    OBJECTIVE:

    Demonstrate minimally invasive rapid body core and brain cooling in a large animal model.

    DESIGN:

    Prospective controlled animal trial.

    SETTING:

    Private research laboratory.

    SUBJECTS:

    Adult dogs, anesthetized, mechanically ventilated.

    INTERVENTIONS:

    Cyclic lung lavage with FC-75 perfluorochemical (PFC) was administered through a dual-lumen endotracheal system in the new technique of 'gas/liquid ventilation' (GLV). In Trial-I, lavage volume (V-lav) was 19 ml/kg, infused and withdrawn over a cycle period (tc) of 37 s. (effective lavage rate V'-lav=31 ml/kg/min.) Five dogs received cold (approximately 4 degrees C) PFC; two controls received isothermic PFC. In Trial-II, five dogs received GLV at V-lav=8.8 ml/kg, tc=16 s, V'-lav=36 ml/kg/min.

    MEASUREMENTS AND MAIN RESULTS:

    Trial-I tympanic temperature change was -3.7+/-0.6 degrees C (SD) at 7.5 min, reaching -7.3+/-0.6 degrees C at 18 min. Heat transfer efficiency was 60%. In Trial-II, efficiency fell to 40%, but heat-exchange dead space (VDtherm) remained constant. Lung/blood thermal equilibration half-time was <8 s. Isothermic GLV caused hypercapnia unless gas ventilation was increased. At necropsy after euthanasia (24 h), modest lung injury was seen.

    CONCLUSIONS:

    GLV cooling times are comparable to those for cardiopulmonary bypass. Heat and CO(2) removal can be independently controlled by changing the mix of lavage and gas ventilation. Due to VDtherm of approximately 6 ml/kg in dogs, efficient V-lav is >18 ml/kg. GLV cooling power appears more limited by PFC flows than lavage residence times. Concurrent gas ventilation may mitigate heat-diffusion limitations in liquid breathing, perhaps via bubble-induced turbulence.