    J Chromatogr Sci. 2001 Jun;39(6):235-8.

    Determination of carboxylic acids, carbohydrates, glycerol, ethanol, and 5-HMF in beer by high-performance liquid chromatography and UV-refractive index double detection.

    Castellari M, Sartini E, Spinabelli U, Riponi C, Galassi S.

    Istituto di Industrie Agrarie, Università degli Studi di Bologna, Italy. cast@foodsci.unibo.it

    A high-performance liquid chromatographic method is proposed for the simultaneous separation of main carboxylic acids, carbohydrates, ethanol, glycerol, and 5-HMF in beer by direct injection. A column packed with a sulfonated divinyl benzene-styrene copolymer and an isocratic elution with 0.0045N sulfuric acid and acetonitrile (6%, v/v) are employed. UV and refractive index detectors connected in series are also used to reduce the matrix interference of phenolic compounds. In conditions described, nine compounds are quantitated in a single chromatographic run without any pretreatment except for sample dilution and filtration before injection. Precision, accuracy, linearity of response, limit of detection, and limit of quantitation are also evaluated for each compound. Satisfactory results are obtained to justify the application of this method to all phases of beer production for process and quality control.

    PMID: 11396687 [PubMed - indexed for MEDLINE]

    Supplemental Content