Comparison of iceball diameter and temperature distribution achieved with 3-mm accuprobe cryoprobes in porcine and human liver tissue and human colorectal liver metastases in vitro

Cryobiology. 2000 Jun;40(4):302-10. doi: 10.1006/cryo.2000.2250.

Abstract

We aimed to assess the thermal profile and size of iceballs produced by Accuprobe cryoprobes in fresh porcine and human liver and human colorectal cancer liver metastases in vitro to allow better planning of cryosurgical treatment of liver metastases. Iceballs were produced by a 20-min single freeze cycle using 8-mm cryoprobes in pig liver in a waterbath at 37 degrees C (n = 8) and 3-mm cryoprobes in pig liver (n = 8), human liver (n = 3), and human colorectal cancer liver metastases (n = 8). The iceball diameters and the temperatures at different distances from the cryoprobe were measured. Mean iceball diameters produced by 8-mm cryoprobes in pig liver were 56.3 mm and varied from 38.7 to 39.6 mm for 3-mm cryoprobes in the different tissues used. There was no significant difference in iceball size in the different tissues. The diameter of the zone of -40 degrees C or less was approximately 44 mm using 8-mm cryoprobes in porcine liver and between 27 and 31 mm using 3-mm cryoprobes in the different tissues examined. The results may allow better preoperative planning of the cryosurgical treatment of liver metastases with Accuprobe cryoprobes.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Colorectal Neoplasms
  • Cryosurgery / instrumentation*
  • Cryosurgery / methods
  • Evaluation Studies as Topic
  • Humans
  • Ice
  • In Vitro Techniques
  • Liver / surgery*
  • Liver Neoplasms / secondary*
  • Liver Neoplasms / surgery*
  • Swine
  • Temperature

Substances

  • Ice