Determination of inorganic anions in environmental waters with a hydroxide-selective column

J Chromatogr A. 2000 Jul 7;884(1-2):175-84. doi: 10.1016/s0021-9673(99)01315-1.

Abstract

US Environmental Protection Agency Method 300.0 specifies the use of an IonPac AS4A anion-exchange column with a carbonate-hydrogencarbonate eluent and suppressed conductivity detection for the determination of inorganic anions in environmental waters by ion chromatography. Hydroxide eluents have not typically been used for the routine analysis of common inorganic anions due to the lack of an appropriate hydroxide selective column and the difficulty in preparing contaminant free hydroxide eluents. The use of ion chromatography with a hydroxide-selective IonPac AS17 column, automated eluent generation and potassium hydroxide gradient represents a new approach to the routine determination of inorganic anions in environmental waters. This new approach, which is a modification of Method 300.0, allows equivalent method performance with improved linearity, precision, and method detection limits. The AS17 column provides superior retention of fluoride from the column void volume and improved resolution from small organic acids, such as formate and acetate, compared to the AS4A column. Quantitative recoveries were obtained for all the common inorganic anions spiked into typical environmental waters using this new approach, and the Performance Based Measurement System Tier 1 method validation quality control acceptance criteria are well within the acceptable ranges defined by Method 300.0. In addition, the EG40 eluent generator eliminates the need to manually prepare eluents, increasing the level of automation and ease-of-use of the ion chromatography system.

MeSH terms

  • Anions / analysis
  • Hydroxides / chemistry
  • Inorganic Chemicals / analysis*
  • Reproducibility of Results
  • Water Pollutants, Chemical / analysis*

Substances

  • Anions
  • Hydroxides
  • Inorganic Chemicals
  • Water Pollutants, Chemical