Magn Reson Med. 1999 Aug;42(2):314-23.

Reduced spatial side lobes in chemical-shift imaging.

Author information

  • 1Radiological Sciences Laboratory, Department of Radiology, Stanford University, Stanford, California 94305-5488, USA. elfar@lucas.stanford.edu

Abstract

Density-weighted k-space sampling with spiral trajectories is used to reduce spatial side lobes in chemical-shift imaging (CSI). In this method, more time is spent collecting data at the center of k space and less time at the edges of k space in order to make the sampling density proportional to a given apodization function, subject to constraints imposed by gradient performance and Nyquist sampling. The efficient k-space coverage of spiral-based trajectories enables good control over the sampling density within practical in vivo scan times. The density-weighted acquisition is compared to a conventional, nonweighted spiral sampling without the application of a window function. For a fixed voxel size and imaging time, the noise variance is observed to be the same for both cases, while spatial side lobes are greatly reduced with the variable-density sampling. This method is demonstrated on a normal volunteer by imaging of brain metabolites at 1.5 T with both single slice CSI and volumetric CSI. Magn Reson Med 42:314-323, 1999.

Copyright 1999 Wiley-Liss, Inc.