    Heart. 1999 Aug;82(2):128-33.

    High resolution in vivo intra-arterial imaging with optical coherence tomography.

    Source

    Department of Electrical Engineering, Massachusetts Institute of Technology, Cambridge, MA 02139, USA.

    Abstract

    BACKGROUND:

    Optical coherence tomography (OCT) is a new method of catheter based micron scale imaging. OCT is analogous to ultrasound, measuring the intensity of backreflected infrared light rather than sound waves.

    OBJECTIVE:

    To demonstrate the ability of OCT to perform high resolution imaging of arterial tissue in vivo.

    METHODS:

    OCT imaging of the abdominal aorta of New Zealand white rabbits was performed using a 2.9 F OCT imaging catheter. Using an ultrashort pulse laser as a light source for imaging, an axial resolution of 10 micrometer was achieved.

    RESULTS:

    Imaging was performed at 4 frames/second and data were saved in either super VHS or digital format. Saline injections were required during imaging because of the signal attenuation caused by blood. Microstructure was sharply defined within the arterial wall and correlated with histology. Some motion artefacts were noted at 4 frames/second.

    CONCLUSIONS:

    In vivo imaging of the rabbit aorta was demonstrated at a source resolution of 10 micrometer, but required the displacement of blood with saline. The high resolution of OCT allows imaging to be performed near the resolution of histopathology, offering the potential to have an impact both on the identification of high risk plaques and the guidance of interventional procedures.

    PMID:
    10409522
    [PubMed - indexed for MEDLINE]
    PMCID:
    PMC1729132
    Free PMC Article

    Images from this publication.See all images (5)Free text

    Figure 1  
    Figure 3  
    Figure 5  
    Figure 2  
    Figure 4  

      Supplemental Content

      Icon for HighWire Icon for PubMed Central

      Save items

      Recent activity

      Your browsing activity is empty.

      Activity recording is turned off.

      Turn recording back on

      See more...
      Write to the Help Desk