    The accuracy of the TriTrac-R3D accelerometer to estimate energy expenditure.

    Source

    Obesity/Nutrition Research Center and Human Energy Research Laboratory, University of Pittsburgh, School of Medicine, PA 15213, USA. jjakicic@ukans.edu

    Abstract

    PURPOSE:

    This study examined the reliability and validity of the TriTrac-R3D triaxial accelerometer to estimate energy expenditure during various modes of exercise.

    METHODS:

    Twenty subjects (age = 21.5+/-3.4 yr; body mass index = 23.3+/-3.6 kg x m(-2)) performed five exercises (treadmill walking, treadmill running, stepping, stationary cycling, and slideboard), with each lasting 20-30 min and workload increased at 10-min intervals. To test the inter-TriTrac reliability, two TriTrac-R3D accelerometers were worn during each exercise period, and to examine validity, a simultaneous measurement of energy expenditure was made using indirect calorimetry (SensorMedics 2900 Metabolic Cart).

    RESULTS:

    Results showed a significant correlation between the two TriTrac-R3D accelerometers during all exercises. The difference in estimated energy expenditure between the two accelerometers during the walking, stepping, and slideboard exercises was less than 1 kcal x min(-1) but statistically significant (P<0.05). There was also a significant correlation between energy expenditure estimated by each of the TriTrac-R3D accelerometers and indirect calorimetry during walking, running, stepping, and slideboard exercise (P<0.05). The interaction of Method x Workload was significant (P<0.05) for each exercise, indicating that the TriTrac-R3D underestimates energy expenditure and that the magnitude of this underestimation increases as workload increases.

    CONCLUSIONS:

    Therefore, energy expenditure estimated via triaxial accelerometry does not increase with increasing workloads. These results suggest that there are limitations to using triaxial accelerometry to quantify energy expenditure.

    PMID:
    10331898
    [PubMed - indexed for MEDLINE]