Preparation of enzyme calibration materials

Clin Chim Acta. 1998 Dec;278(2):151-62. doi: 10.1016/s0009-8981(98)00141-7.

Abstract

Standardisation in clinical enzymology needs not only reference methods but also reference materials. While single-enzyme reference enzymes have been developed, a multienzyme certified reference material (MECRM) available in high amount remains to be produced. To transfer trueness from the value of the reference system to patients' results, validated enzyme calibrators (EC) are also needed. Both the MECRM and the ECs must exhibit the same catalytic properties as the corresponding enzymes in human plasma. Moreover, commutability of these materials with patients' samples must be experimentally tested for one or a set of methods defined by an analytical specificity equal to that of the reference method. Various experimental studies have shown that the commutability of an enzyme material depends on the source of enzyme and its purification process, the matrix (including cofactors, effectors, additives, stabilisers... ) and the mode of processing of the final material. To promote intermethod calibration in clinical enzymology, a collaborative programme between the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC), Institute for Reference Materials and Measurements (IRMM, Geel, Belgium) and IFCC corporate members is in progress for the development of a MECRM containing amylase, ALT, AST, ALP, CK, GGT, LDH, and lipase and exhibiting a wide and defined commutability.

MeSH terms

  • Calibration*
  • Catalysis
  • Enzymes / metabolism*
  • Humans
  • Reference Standards*
  • Viscosity