Proc Natl Acad Sci U S A. May 1986; 83(9): 2879–2883.
PMCID: PMC323410

Contact pressures in the human hip joint measured in vivo.

Abstract

The pressures on human articular cartilage have been measured in vivo. An instrumented femoral head prosthesis that telemeters interarticular pressure at 10 discrete locations 253 times per second was implanted in apposition to natural acetabular cartilage. Data were acquired during surgery, recovery, rehabilitation, and normal activity, for longer than 1 year after surgery. Pressure magnitudes were synchronized with body-segment kinematic data and foot-floor force measurements so as to locate transduced pressure areas on the natural acetabulum and correlate movement kinematics and dynamics with local cartilage pressures. The data reveal very high local (up to 18 MPa) and nonuniform pressures, with abrupt spatial and temporal gradients, that correlate well both in magnitude and distribution with in vitro data and computer simulations of synovial joint mechanics. Peak pressures in vivo are, however, considerably higher than pressures measured in vitro under the putative forces experienced by the joint in life, particularly in normal movements where cocontraction occurs in agonist and antagonist muscles across the hip joint. Thus, extant gait-analysis studies which apply inverse Newtonian calculations to infer joint forces establish the lower limit on such forces, since such analyses include only the net muscular torques about the joint and cannot account for the contribution of the increment in joint force due to muscular cocontraction. Our data also contribute to the understanding of normal synovial joint tribology and the possible role of mechanical factors in the deterioration evident in osteoarthritis. Further, design criteria for both partial and total hip replacement prostheses and specific aspects of rehabilitation protocols following hip surgery (e.g., the extent to which crutches and canes unload the hip joint) warrant reconsideration in light of the extraordinary high pressures measured during the activities of daily living.

Full text

Full text is available as a scanned copy of the original print version. Get a printable copy (PDF file) of the complete article (1.4M), or click on a page image below to browse page by page. Links to PubMed are also available for Selected References.

Images in this article

Click on the image to see a larger version.

Selected References

These references are in PubMed. This may not be the complete list of references from this article.
  • DENHAM RA. Hip mechanics. J Bone Joint Surg Br. 1959 Aug;41-B:550–557. [PubMed]
  • English TA, Kilvington M. In vivo records of hip loads using a femoral implant with telemetric output (a preliminary report). J Biomed Eng. 1979 Apr;1(2):111–115. [PubMed]
  • Lee P, Rooney PJ, Sturrock RD, Kennedy AC, Dick WC. The etiology and pathogenesis of osteoarthrosis: a review. Semin Arthritis Rheum. 1974 Spring;3(3):189–218. [PubMed]
  • Greenwald AS, O'Connor JJ. The transmission of load through the human hip joint. J Biomech. 1971 Dec;4(6):507–528. [PubMed]
  • Day WH, Swanson SA, Freeman MA. Contact pressures in the loaded human cadaver hip. J Bone Joint Surg Br. 1975 Aug;57(3):302–313. [PubMed]
  • Rushfeld PD, Mann RW, Harris WH. Influence of cartilage geometry on the pressure distribution in the human hip joint. Science. 1979 Apr 27;204(4391):413–415. [PubMed]
  • Rushfeldt PD, Mann RW, Harris WH. Improved techniques for measuring in vitro the geometry and pressure distribution in the human acetabulum. II Instrumented endoprosthesis measurement of articular surface pressure distribution. J Biomech. 1981;14(5):315–323. [PubMed]
  • Brown TD, Shaw DT. A technique for measuring instantaneous in vitro contact stress distributions in articular joints. J Biomech. 1982;15(4):329–333. [PubMed]
  • Rushfeldt PD, Mann RW, Harris WH. Improved techniques for measuring in vitro the geometry and pressure distribution in the human acetabulum--I. Ultrasonic measurement of acetabular surfaces, sphericity and cartilage thickness. J Biomech. 1981;14(4):253–260. [PubMed]
  • Tepic S, Macirowski T, Mann RW. Mechanical properties of articular cartilage elucidated by osmotic loading and ultrasound. Proc Natl Acad Sci U S A. 1983 Jun;80(11):3331–3333. [PMC free article] [PubMed]
  • Carlson CE, Mann RW, Harris WH. A radio telemetry device for monitoring cartilage surface pressures in the human hip. IEEE Trans Biomed Eng. 1974 Jul;21(4):257–264. [PubMed]
  • Mann RW, Antonsson EK. Gait analysis--precise, rapid, automatic, 3-D position and orientation kinematics and dynamics. Bull Hosp Jt Dis Orthop Inst. 1983 Fall;43(2):137–146. [PubMed]
  • Antonsson EK, Mann RW. The frequency content of gait. J Biomech. 1985;18(1):39–47. [PubMed]
  • Holmes MH, Lai WM, Mow VC. Singular perturbation analysis of the nonlinear, flow-dependent compressive stress relaxation behavior of articular cartilage. J Biomech Eng. 1985 Aug;107(3):206–218. [PubMed]
  • Kenyon DE. A model for surface flow in cartilage. J Biomech. 1980;13(2):129–134. [PubMed]
  • Patriarco AG, Mann RW, Simon SR, Mansour JM. An evaluation of the approaches of optimization models in the prediction of muscle forces during human gait. J Biomech. 1981;14(8):513–525. [PubMed]
  • Surin V, Hedelin E, Smith L. Degenerative lumbar spinal stenosis: results of operative treatment. Acta Orthop Scand. 1982 Feb;53(1):79–85. [PubMed]

Articles from Proceedings of the National Academy of Sciences of the United States of America are provided here courtesy of National Academy of Sciences

Formats:

Related citations in PubMed

See reviews...See all...

Cited by other articles in PMC

See all...

Links

  • Cited in Books
    Cited in Books
    PubMed Central articles cited in books
  • PubMed
    PubMed
    PubMed citations for these articles

Recent Activity

Your browsing activity is empty.

Activity recording is turned off.

Turn recording back on

See more...