![]() | ![]() |
Formats:
|
||||||||||||||
Copyright © 2000 Current Science Ltd Relationships between volume and pressure measurements and stroke
volume in critically ill patients 1Leiden University Medical Center, Leiden, The Netherlands. Alexander JGH Bindels: abindels/at/worldonline.nl Received April 5, 1999; Revisions requested September 7, 1999; Revised April 25, 2000; Accepted May 2, 2000. Abstract Objective: To evaluate the relationships between the changes in stroke volume
index (SVI), measured in both the aorta and the pulmonary artery, and the
changes in intrathoracic blood volume index (ITBVI), as well as the
relationship between changes in aortic SVI and changes in the pulmonary artery
wedge pressure (PAWP). Design: Prospective study with measurements at predetermined
intervals. Setting: Medical intensive care unit of a university hospital. Patients and methods: One hundred and fifty-four measurements were taken in 45
critically ill patients with varying underlying disorders. Aortic SVI and
pulmonary arterial SVI were determined with thermodilution. PAWP was measured
using a pulmonary artery catheter. ITBVI was determined with thermal-dye
dilution, using a commercially available computer system. Results: A good correlation was found between changes in ITBVI and changes
in aortic SVI. However, this correlation weakened when changes in ITBVI were
plotted against changes in pulmonary arterial SVI, which was in part probably
due to mathematical coupling between ITBVI and aortic SVI. A good correlation
between changes in ITBVI and changes in aortic SVI could also be established in
most of the individual patients. No correlation was found between changes in
PAWP and changes in aortic SVI. Conclusion: ITBVI seems to be a better predictor of SVI than PAWP. ITBVI may
be more suitable than PAWP for assessing cardiac filling in clinical
practice. Keywords: cardiac output, intrathoracic blood volume, pulmonary artery wedge pressure, stroke volume, thermal dye dilution Introduction Assessing the volume status of critically ill patients is a routine
task for intensivists. Clinical assessment by history taking, physical
examination, fluid balance or radiographic findings provides belated or
unreliable information [1,2,3,4]. Apart from clinical skills,
invasive monitoring is widely applied as a tool for assessment of volume
status. In its simplest form, central venous pressure (CVP) can be measured, by
using a central venous catheter. In more complicated cases a pulmonary artery
catheter is often used, with PAWP as the variable for determining cardiac
filling. Because CVP and PAWP depend not only on cardiac filling, but also on
ventricular compliance, these pressures are only poor reflections of a
patient's volume status [5,6,7]. Moreover, CVP and PAWP are absolute
intravascular pressures, meaning that changes in intrathoracic pressures will
influence the recorded values of CVP and PAWP. This applies in particular to
mechanically ventilated patients who are ventilated with positive
end-expiratory pressure. Thus, therapeutic decisions based on CVP and/or PAWP
may be based on inaccurate measures of a patient's volume status. The thermal-dye dilution technique was originally introduced as a
method to measure extravascular lung water (EVLW) [8]. In
recent years the emphasis has moved to the ITBVI as the most important variable
that can be determined using this technique. In a limited number of studies,
ITBVI has been shown to correlate well with the cardiac index (CI), and it
appears to be a better measure of cardiac filling than CVP or PAWP [7,9,10,11]. Lichtwarck-Aschoff et al [7] showed a correlation coefficient of 0.65 between changes in
ITBVI and SVI in 21 patients with acute respiratory failure. Gödje et
al [9] showed a correlation coefficient of 0.87
between changes in ITBVI and changes in CI in cardiac surgery patients.
Recently, Sakka et al [10] showed a correlation
coefficient of 0.67 between changes in ITBVI and changes in SVI during the
early phase of haemodynamic instability in patients with sepsis or septic
shock. In a mixed group of critically ill patients we studied the
correlations between SVI and PAWP, measured using a pulmonary artery catheter,
and the correlations between SVI and ITBVI, measured with a commercially
available computer system using the thermal-dye dilution technique. Methods The data presented in this study were prospectively obtained from 45
critically ill patients. All patients concomitantly participated in four other
studies. In those studies, haemodynamic patterns of specific clinical entities,
with the emphasis on EVLW, were investigated. ITBV, however, was measured
specifically for the present study. The groups consisted of patient with acute
respiratory distress syndrome (ARDS), patients with acute cardiogenic pulmonary
oedema [12], patients with a septic shock [13], and patients with hepatic cirrhosis requiring a
transjugular intrahepatic portosystemic shunt (TIPS). In all of these studies,
patients were monitored using a pulmonary artery catheter (7.5-F Swan
Ganz-catheter, Model VS 1721; Ohmeda, Swindon, UK) and a 4-Fr fiberoptic
catheter (Pulsiocath PV 2024; Pulsion, Munich, Germany), introduced into the
descending aorta through a 6-Fr introducer sheath (Model 616150A; Ohmeda) and
connected to a computer system (COLD Z-021 system; Pulsion) for determination
of ITBVI. Haemodynamic measurements, both with the pulmonary artery catheter and
the thermal-dye dilution technique, were made at regular intervals during the
first 24 h after admission to the intensive care unit. Fluid therapy was given
as long as every seperate fluid bolus (500 ml colloids over 20 min) resulted in
an increase of CI of 10% or more. PAWP was not allowed to exceed 18 mmHg in
patients with acute cardiogenic pulmonary oedema, however, and was not allowed
to exceed 16 mmHg in the other categories. Whenever CI increased by less than
10%, fluid challenges were stopped, regardless of the PAWP at that point, and
inotropes and/or vasopressors were given when appropriate. All study protocols were approved by the Local Ethics Committee, and
informed consent was given by each patient or his/her next of kin. The pulmonary artery catheter was used for measurements of CVP and
PAWP, with the midchest level as zero reference. The heart rate was recorded
continuously with one of the standard leads of the electrocardiogram. PAWP was
measured exclusively by the investigators and not by the nursing staff, taking
problems associated with PAWP measurement and recommendations from the
literature into account [14]. The COLD system was connected both to the pulmonary artery catheter
and to the fibreoptic catheter in the aorta, which enabled us to determine CI
in the pulmonary artery and in the aorta in one measurement. SVI was calculated
by dividing the respective CIs by the accompanying heart rate. The COLD system
was also used for determination of ITBVI. Measurements were done by injecting
10 cm3 of an ice-cold indocyanin green (ICG) solution (2 mg/ml). The
mean value of two measurements was used for analysis. For details concerning the thermal-dye dilution method, see Lewis
et al [8] and Pfeiffer et al [15].
Briefly, the method uses two indicators (ie ice-cold water and ICG). Cold is
distributed throughout both intravascular and extravascular volume, whereas ICG
remains in the intravascular volume. Both indicators are injected into the
right atrium, and concentration changes with time are recorded in the
descending aorta. Thus, dilution curves are obtained for both indicators. From
the thermodilution curve aortic CI is determined. From each indicator's
dilution curve a mean transit time (MTT) can be derived. MTT is composed of the
appearance time, which is the time until the first indicator particle has
arrived at the point of detection, and the mean time difference between the
occurrence of the first particle and all the following particles [15]. The product of CI and MTT is the volume between the site
of injection and the site of detection. ITBVI can be calculated using the
following formula: ITBVI (ml/m2) = CI × MTTaorta
(ICG) The correlations between the variables, as well as correlations
between the changes in these variables, were studied using linear regression
analysis. Changes in the variables were calculated by subtracting the first
from the second measurement, the second from the third, and so on. To reduce
the influence of changes in contractility and afterload, we used only those
values for the analysis for which no supportive adjustments were made with
inotropes and/or vasopressors between the measurements. Both pooled and
intraindividual relationships were studied. The method described by Bland and
Altman [16] was used for assessing differences between
pulmonary arterial CI and aortic CI. Results A total of 283 haemodynamic measurements were made in 45 critically
ill patients (10 patients with ARDS, 10 patients with acute cardiogenic
pulmonary oedema, 15 patients with septic shock and 10 patients with hepatic
cirrhosis requiring TIPS). After discarding the measurements in which
supportive adjustments were made with inotropes and/or vasopressors, 154
changes between measurements were left for analysis. Details concerning the
subgroups are shown in Table 1. Thirty-six patients were mechanically ventilated
throughout the study protocol.
Pulmonary arterial CI and aortic CI correlated well (Fig.
(Fig.1).1
Figure Figure22
Figure Figure33
Figure Figure44
Discussion The present study shows a good correlation between changes in ITBVI
and aortic SVI. This correlation could also be found in the individual patients
in three of the four disease categories studied. However, the correlation
weakened when, in the pooled data, ITBVI was plotted against pulmonary arterial
SVI. No consistent correlation could be established between PAWP and aortic
SVI. CVP and PAWP are pressures that are used in clinical practice to
assess cardiac filling or cardiac preload. Under experimental conditions, the
so-called ventricular performance curves show a close curvilinear relationship
between the end-diastolic pressure of the ventricle and the stroke volume or
cardiac output, provided that contractility and afterload are held constant. In
clinical practice this relationship may be distorted for several reasons. The first reason is that several assumptions have to be made for PAWP
to reflect the end-diastolic volume of the ventricle. PAWP must be accurately
measured, it must reflect left atrial pressure (LAP), LAP must reflect left
ventricular end-diastolic pressure (LVEDP), and then LVEDP must relate directly
to left ventricular end-diastolic volume to be a true measure of cardiac
filling. In clinical practice there are many doubts about the accuracy of the
PAWP measurement. Accurate measurements are frequently prevented by technical
aspects. There is also an astonishing lack of basic knowledge among clinicians
and nurses on how the measurement should be performed [17,18,19,20]. Apart from the technical factors, there are also clinical
entities that interfere with the reliability of PAWP in reflecting LAP
accurately. Pulmonary venous obstruction (eg tumours, atrial myxomas,
mediastinal fibrosis, pulmonary venous thrombosis) increases PAWP, without an
accompanying increased LAP. Disparity between LAP and LVEDP is found in the
case of mitral stenosis, and, perhaps more often, in the presence of a
decreased left ventricular compliance. A change in ventricular compliance,
often met in critically ill patients, may also distort the assumed relationship
between LVEDP and left ventricular end-diastolic volume. Furthermore,
interventricular dependence also influences the pressure-volume curve of the
left ventricle. Hence, disease states with an increased right ventricular
afterload (eg acute pulmonary hypertension) will also impair left ventricular
compliance. Finally, all intrathoracic pressure changes will affect the
recorded values of CVP and PAWP, because these pressures are measured relative
to ambient air pressure. Therefore, the measured pressures are not transmural
pressures, which is especially true in case the tip of the pulmonary artery
catheter is located outside a West zone III [21]. The second reason for the distorted relationship between the cardiac
filling pressures and the stroke volume in clinical practice is that the
requirement for the contractility and the afterload to be constant is hardly
ever met in clinical practice. Leaving aside the question of whether this
requirement is verifiable, practically all interventions interfere either with
the myocardial contractility (eg inotropes) or with the ventricular afterload
(eg vasoconstrictors, vasodilators). Although we tried to make an approximate
correction for this phenomenon, by leaving out those measurements in which
supportive changes were made with inotropes or vasoactive medications, it
cannot be ruled out that this phenomenon played a role in the results we
found. Taking into account the reasons indicated above, it is not surprising
that we did not find a consistent correlation between PAWP and aortic SVI in
the individual patients. The present results confirm those of earlier studies
[7,9,10,11]. In the patients we studied
there were no major differences in the correlation of PAWP and aortic SVI
between the different disease states, regardless of whether all patients were
ventilated mechanically (ARDS), or only a minority of patients (TIPS) was on
mechanical ventilation. In conclusion, PAWP is influenced by so many factors
other than cardiac filling that it is not a reliable indicator of cardiac
filling in clinical practice. Therefore, the absolute values of these two
variables are not an adequate reflection of the cardiac filling conditions of
an individual patient. Changes in ITBVI showed better correlations with changes in aortic SVI
than did changes in PAWP, which is also in accordance with earlier findings
[7,9,10,11]. From the individual regression
lines (Fig. (Fig.4),4 By connecting the Swan-Ganz catheter to the COLD system, time
differences between pulmonary arterial CI and aortic CI were precluded.
Pulmonary arterial CI and aortic CI were closely correlated, with a mean higher
value of aortic CI of 0.49 l/min per m2. This is in accordance with an
earlier report [11]. However, the difference in the
correlation between ITBVI and pulmonary arterial SVI, and the correlation
between ITBVI and aortic SVI (Fig. (Fig.2)2 The thermal-dye dilution technique was originally developed to
determine EVLW. As a consequence, validation of the method is based on
comparison of measured values of EVLW with gravimetrically determined EVLW.
These values correlate well, with an overestimation of the thermal-dye
technique in the lower range and an underestimation in the higher range of EVLW
values [24,25,26]. In a recent study [27],
circulating (total) blood volume measured with the COLD system correlated well
with standard methods for measuring circulating blood volume. From these
results, it has been assumed that measured ITBVI also correlates well with the
actual intrathoracic volume. This has not been validated formally, however. On
the other hand, the correlations we found are those one would expect on the
basis of physiological knowledge. This implies that ITBVI, at least, is a
reflection of the actual intrathoracic volume. In conclusion, the present study shows that the cardiac filling in
critically ill patients may not adequately be predicted by PAWP. ITBVI seems to
be a more reliable predictor of cardiac filling, because changes in ITBVI
closely relate with changes in aortic SVI. Partially, however, this may be due
to mathematical coupling. Whether the use of ITBVI for guidance of fluid
therapy will improve patient outcome should be subject to further studies. Acknowledgements The authors would like to thank Professor Jean-Louis Vincent (Free
University of Brussels, Belgium) for his comments on earlier versions of this
manuscript. References
|
PubMed related articles
Your browsing activity is empty. Activity recording is turned off. |
|||||||||||||
Crit Care Med. 1993 Jun; 21(6):871-7.
[Crit Care Med. 1993]Intensive Care Med. 1995 Feb; 21(2):125-31.
[Intensive Care Med. 1995]Crit Care Med. 1983 Jul; 11(7):498-501.
[Crit Care Med. 1983]Intensive Care Med. 1984; 10(6):309-13.
[Intensive Care Med. 1984]Crit Care Med. 1984 Feb; 12(2):107-12.
[Crit Care Med. 1984]J Surg Res. 1979 Oct; 27(4):250-61.
[J Surg Res. 1979]Intensive Care Med. 1992; 18(3):142-7.
[Intensive Care Med. 1992]Eur J Cardiothorac Surg. 1998 May; 13(5):533-9; discussion 539-40.
[Eur J Cardiothorac Surg. 1998]J Crit Care. 1999 Jun; 14(2):78-83.
[J Crit Care. 1999]Am J Cardiol. 1999 Nov 15; 84(10):1158-63.
[Am J Cardiol. 1999]Intensive Care Med. 1991; 17(3):I-VIII.
[Intensive Care Med. 1991]J Surg Res. 1979 Oct; 27(4):250-61.
[J Surg Res. 1979]JAMA. 1990 Dec 12; 264(22):2928-32.
[JAMA. 1990]Crit Care Med. 1997 Feb; 25(2):213-20.
[Crit Care Med. 1997]Crit Care Med. 1994 Oct; 22(10):1674-8.
[Crit Care Med. 1994]Am J Crit Care. 1996 Jan; 5(1):49-54.
[Am J Crit Care. 1996]Intensive Care Med. 1992; 18(3):142-7.
[Intensive Care Med. 1992]Eur J Cardiothorac Surg. 1998 May; 13(5):533-9; discussion 539-40.
[Eur J Cardiothorac Surg. 1998]J Crit Care. 1999 Jun; 14(2):78-83.
[J Crit Care. 1999]Intensive Care Med. 1992; 18(3):142-7.
[Intensive Care Med. 1992]Eur J Cardiothorac Surg. 1998 May; 13(5):533-9; discussion 539-40.
[Eur J Cardiothorac Surg. 1998]J Crit Care. 1999 Jun; 14(2):78-83.
[J Crit Care. 1999]Intensive Care Med. 1990; 16(2):115-20.
[Intensive Care Med. 1990]J Crit Care. 1996 Dec; 11(4):180-8.
[J Crit Care. 1996]J Appl Physiol. 1985 Sep; 59(3):673-83.
[J Appl Physiol. 1985]Crit Care Med. 1985 May; 13(5):432-5.
[Crit Care Med. 1985]Chest. 1986 Oct; 90(4):588-94.
[Chest. 1986]Crit Care Med. 1995 May; 23(5):885-93.
[Crit Care Med. 1995]