• We are sorry, but NCBI web applications do not support your browser and may not function properly. More information
Logo of nihpaAbout Author manuscriptsSubmit a manuscriptNIH Public Access; Author Manuscript; Accepted for publication in peer reviewed journal;
J Magn Reson. Author manuscript; available in PMC Feb 1, 2011.
Published in final edited form as:
PMCID: PMC2818327
NIHMSID: NIHMS154679

A Bioreactor for In Cell Protein NMR

Abstract

The inside of the cell is a complex environment that is difficult to simulate when studying proteins and other molecules in vitro. We have developed a device and system that provides a controlled environment for Nuclear Magnetic Resonance (NMR) experiments involving living cells. Our device comprises two main parts, an NMR detection region and a circulation system. The flow of medium from the bottom of the device pushes alginate encapsulated cells into the circulation chamber. In the chamber, the exchange of oxygen and nutrients occurs between the media and the encapsulated cells. When the media flow is stopped, the encapsulated cells fall back into the NMR detection region, and spectra can be acquired. We have utilized the bioreactor to study the expression of the natively disordered protein α-synuclein, inside Escherichia coli cells.

Keywords: bioreactor, in-cell nuclear magnetic resonance (NMR)

1. Introduction

Most biophysical experiments on proteins are conducted in dilute solution. Most proteins, however, serve their physiologically relevant function in cells, which have a complex crowded environment that affect several protein properties compared to dilute solution [1; 2; 3; 4]. For this reason, there is an increasing interest in studying proteins inside living cells. Nuclear magnetic resonance spectroscopy (NMR) has become a popular tool for experiments on living cells because it provides atomic-level information about cellular components and is nondestructive [5].

A disadvantage of NMR spectroscopy is its low sensitivity. Selective isotopic enrichment or labeling of the targeted species with an NMR active nucleus (e.g., 15N, 13C or 19F) is one way around this problem. Serber et al. suggest that the minimum concentration of the protein under study should be at least ~150 μM for 15N enrichment or ~50 μM for 13C side chain enrichment [6]. Therefore, the target protein must be overexpressed or introduced into the cell by other means (e.g., micro injection, cell penetrating peptides [5]). To increase sensitivity further, high cell densities (109–1011 cells/mL) are used, and the data are time averaged.

Current experimental setups for protein in cell NMR have several drawbacks. First, the lack of aeration and the high cell density create an anaerobic environment. Second, metabolites and waste products accumulate. These characteristics can decrease cell viability, limiting the cell types that can be used, and make it difficult to monitor temporal changes. Overcoming these challenges requires an NMR compatible device that provides a controlled environment for living cells.

Devices with these characteristics have been developed [7; 8; 9]. One type is an in-magnet bioreactor that enables growth of microorganisms to a high density [10; 11]. Another type is a perfusion system that flows media down through immobilized cells [12; 13]. These devices tend to be complex and difficult to fabricate. Furthermore, they are designed to study metabolism.

Here, we describe a circulating encapsulated cells (CEC) bioreactor and accessories for in-cell protein NMR. The instrument comprises parts that are commercially available or easily fabricated. The expression of the natively disordered human protein α-synuclein in Escherichia coli is used to demonstrate its capabilities. α-Synuclein is a 14.5 kDa protein implicated in the pathogenesis of Parkinson’s disease [14]. The expression of the plasmid borne α-synuclein gene is controlled by a lactose inducible, phage T7 promoter. We investigate the bioreactor’s ability to measure the accumulation of α-synuclein with time in living E. coli cells.

2. Results

2.1 The CEC bioreactor

The bioreactor is made from Teflon to facilitate NMR experiments involving 1H detection (We have made a similar version from Plexiglass for 19F detection). We designed the bioreactor for an 8 mm probe to facilitate fabrication with conventional machine tools. The bioreactor comprises three main parts; an 8 mm diameter tube, a circulation chamber, and an adjustable threaded cap (Fig. 1). An inlet for an Upchurch Scientific Super Flangeless Fitting is located on the upper part of the cap (Fig. 1E). A 1/16″ OD inlet for Upchurch Scientific perfluoro alkoxy alkane tubing is located at the bottom (Fig. 1A).

Fig. 1
The CEC bioreactor. On the left, the pump is off and the encapsulates are settled. On the right, the pump is on and the encapsulates circulate at a steady state in the upper chamber. A: tubing inlet, B: NMR detection region, C: circulation chamber, D: ...

Most of the components (Fig. 2) are commercially available. The liquid media is contained in a 1 L Corning three neck spinner flask with a tubing adaptor on one side arm (Fig. 2A). A peristaltic MasterFlex pump (Fig. 2B), moves the media at a rate of 45 ml/min from the flask, through the tubing adaptor, into the PFA tubing (Fig. 2, yellow lines). The tubing runs from the pump, into the bottom of an 8 mm Varian triple resonance z gradient probe through an opening created by removing the heater (Fig. 2D). The temperature is controlled with the spectrometer’s FTS Systems heating apparatus (Model TC-84). PFA tubing between the pump and the bioreactor is placed in a Fisher Scientific Isotemp water bath to warm the media (Fig. 2C), which flows from the bottom of the bioreactor to the top (Fig. 1, arrows on right side). The PFA tubing at the top of the bioreactor returns the media, via a pH probe, to the 1 L the Corning three neck spinner flask. The pH probe, pump and NMR spectrometer are connected to a laptop computer (Fig. 2, green lines).

Fig. 2
The experimental set-up. A: Corning spinner flask fitted with a vented cap on one side arm and tubing adaptors on the other, B: peristaltic pump, C: water bath, D: 8 mm probe with heater removed, E: bioreactor, F: magnet, G: pH probe, H: computer, I: ...

The 1H-15N band-selective optimized flip-angle short transient (SOFAST) heteronuclear multiple quantum coherence (HMQC) [15] pulse program provided in the Varian Biopak suite of pulse sequences was modified to send voltage outputs to the computer. The perfusion pump is controlled by one of the spare output lines on the Varian Inova console. The pump is switched on before the first steady state scan, and remains on during the perfusion delay. It is then then switched off during the bead settling delay. After the first steady state scan, the pulse sequence skips the pump control code. The signals are interpreted by LabView (National Instruments) software, which controls the pump. The software also records the pH value at 1 min intervals.

Cells are electrostatically encapsulated into 1 mm diameter Ca2+ alginate spheres to keep them in the bioreactor [16]. The circulation of the encapsulated cells facilitates the delivery of nutrients and waste removal. The bioreactor (Fig. 1) has two states: pump off, and pump on. When the pump is off, encapsulated cells settle into the 8 mm diameter tube for data acquisition (Fig. 1, left panel). When the pump is on, the encapsulates travel from the 8 mm diameter tube into the wider circulation chamber. The movement of the encapsulates from a narrow to wider tube results in a reduction in pressure causing the encapsulates to circulate in the chamber (Fig. 1, right panel). The pulsed, upward motion also prevents the encapsulates from clogging the outlet of the cell chamber. Settling of the encapsulates requires approximately 90 s from the time the pump is turned off.

2.2 CEC bioreactor without flowing media

To assess the bioreactor’s suitability for in cell NMR experiments, the 1H-15N HSQC spectrum of encapsulated E. coli in the bioreactor is compared to the HSQC spectrucm of 15N enriched α-synuclein obtained in a conventional 5 mm NMR probe (Fig. 3A,B). The in vitro spectrum of purified α-synuclein (Fig. 3C) is shown as a reference. The in vitro spectra were acquired at a lower temperature because we have shown that the in vitro spectrum acquired at 10 °C is equivalent to in cell spectra acquired at 37 °C [1; 17]. The similarity of the spectra in Fig. 3 indicates that α-synuclein can be detected in the bioreactor.

Fig. 3
Comparing α-synuclein spectra. A) In cell HSQC spectrum of alginate encapsulated E. coli expressing α-synuclein in the bioreactor. B) In cell HSQC spectrum of E. coli expressing α-synuclein. C) In vitro HSQC spectrum of 200 μM ...

To assess the bioreactor’s effect on spectral quality, the in cell HSQC spectrum of encapsulated E. coli expressing 15N enriched α-synuclein in the bioreactor is compared to the HSQC spectrum of the same encapsulates in a 5 mm tube (Fig. 3A,D). The spectrum of the encapsulated cells in the 5 mm tube is consistent with the published spectrum [18]. The crosspeaks broaden when encapsulates are placed in the bioreactor, but the quality of spectra is only slightly degraded.

2.3 CEC bioreactor with flowing media

The expression of α-synuclein was monitored with the 1H-15N SOFAST HMQC pulse sequence, rather than the HSQC sequence, to obtain better time resolution. Spectra as a function of time are shown in Fig. 4A–C. The spectrum of the encapsulates before induction (Fig. 4A) has few crosspeaks and no unambiguous α-synuclein crosspeaks. After induction new crosspeaks begin to appear. With each successive spectrum, the crosspeaks increase in volume as seen at 4 h and 18 h (Figs. 4B and C). Using methods described by Slade et al. [19] we determined the intracellular concentration of α-synuclein to be 0.8 mM at 18 h.

Fig. 4
In-cell SOFAST 15N-1H HMQC [12] spectra (37°C) of E. coli expressing α-synuclein in the bioreactor. A) Spectrum collected before induction. B) 4 h post induction. C) 18 h post induction. D) Spectrum of the spent medium.

As a control, the encapsulates were removed after the experiment and a spectrum was acquired of the surrounding media (Fig. 4D). The spectrum shows only a weak crosspeak, indicating that the bulk of the signal comes from the encapsulated cells. The viability of the E. coli in the bioreactor experiments was determined by plating serial dilutions of dissolved encapsulates before and after each experiment. The viability was 95%. The pH of the medium perfused around the encapsulated cells remained at 7.00 for the duration of the experiment.

Although the CEC bioreactor provides an environment where encapsulated E. coli cells express α-synuclein, our sacrifice of spectral resolution for increased time resolution (see the Experimental section) made it difficult to distinguish between metabolites and protein crosspeaks. To determine which crosspeaks corresponded to α-synuclein, we collected spectra of fresh media (Fig 5A) and of E. coli containing a pUC18 plasmid [20] without the α-synuclein gene (Fig 5B). Overlaying these spectra with the spectrum of α-synuclein expressed in the bioreactor (Fig 5C), shows that most of the crosspeaks are from α-synuclein (Fig 5D).

Fig. 5
Determining which crosspeaks correspond to α-synuclein. A) In-cell SOFAST 15N-1H HMQC spectrum of the defined phosphate-free minimal media. B) Spectrum of 15N encriched encapsulated E. coli cells containing the control pUC18 plasmid. C) Spectrum ...

The overlay allowed us to quantify temporal changes in crosspeak volumes (Fig 6). The crosspeak from the defined minimal media is the only crosspeak detectable at 30 min (Fig 6H). Although induction occurred at 30 min, there was a lag phase of approximately 4 h before crosspeaks could be detected (Fig 6A–B, E–G). Some crosspeaks are not detectable until approximately 7 h (Fig 6C,D),

Fig. 6
Temporal changes in crosspeak volume after inducing α-synuclein expression in the bioreactor. A–E) α-Synuclein crosspeaks, F–G) Metabolite crosspeaks. H) Crosspeaks from the defined phosphate free minimal media. Crosspeak ...

The volumes of α-synuclein crosspeaks increased with time, beginning with a lag phase before growing exponentially to a plateau (Fig 6A–E). One crosspeak deviated from this trend (Fig 6D), most likely because the poor resolution in this area of the spectrum (Fig 5D). Temporal changes in crosspeak volumes for two metabolites showed different trends. One metabolite remained constant (Fig 6F), while the other metabolite showed a time dependence that resembled the α-synuclein crosspeaks (Fig 6G). The crosspeak from the defined minimal media, is the only crosspeak that showed a slight decrease in intensity with time (Fig 6H).

3. Discussion

The CEC bioreactor (Fig. 1) is designed to provide a controlled environment for NMR experiments involving living cells. It allows media to deliver nutrients and remove waste from encapsulated cells contained in a circulation chamber (Fig. 1, right panel). When the flow of media is stopped, the encapsulated cells settle allowing data acquisition (Fig. 1, left panel).

In the experimental setup, the CEC bioreactor is the only component located inside the spectrometer (Fig. 2). This configuration allows the external components to be altered without removing the bioreactor before or during the experiment, facilitating studies requiring different conditions in one experiment. The setup is also versatile. Different solution probes and sensors can be inserted between the external components. The tubing can be rerouted, for example, to send the media to a waste container. In addition, the material used to make the bioreactor can be changed for experiments requiring different isotopic nuclei detection. Here we use Teflon for 15N detection, but a Plexiglass bioreactor can be used for 19F NMR.

To our knowledge, the CEC bioreactor is the first bioreactor suitable for protein in cell NMR experiments (Fig. 3A,B). The design provides an environment where encapsulated cells can express protein while maintaining reasonably high quality in cell NMR spectra (Fig 3C–D). Furthermore, the bioreactor can be used to quantify temporal changes in crosspeak volumes during the experiment (Fig. 4 and and55).

Our data allow us to draw several conclusions. We showed that α-synuclein was present at an intracellular concentration of 0.8 mM at 18 h. Using information from Figure 6, we conclude that the detection limit for in-cell NMR is approximately 0.14 mM. This finding is consistent with other work on the minimal intracellular protein concentration needed for in-cell NMR [6]. For most residues, the detection limit is achieved after 3 h. Two protein crosspeaks do not follow this trend in that they are not detectable until approximately 7 h (Fig. 6C,D). The crosspeaks from glycine residues that comprise the ear shaped pattern in the upper left region of α-synuclein Figure 4 (15N ppm 108–113, 1H ppm 8.3–8.7), follow a similar trend. The delay in detectability may be due to differential binding of α-synuclein to other intracellular components, which broadens their crosspeaks. Another possibility for the delay is differential relaxation, because in vitro models for α-synuclein dynamics show that certain residues experience less mobility [17]. Decreased mobility produces broader, weaker signals, which would explain the longer time required to detect them.

In summary, the CEC bioreactor provides a controlled environment where protein NMR spectra data can be acquired in living E. coli cells. Our next goal is to show that the CEC bioreactor is compatible with other cell types, and is versatile enough for metabolomic, as well as protein NMR experiments. We specifically want to focus on eukaryotic cells whose viability is adversly affected by current methods. Our long-term goal is to monitor temporal changes in protein structure and metabolism due to perturbations, such as drug interactions, in human cells, and so increase the understanding of intracellular components under physiological conditions.

4. Experimental

4.1 Purification of wild type α-synuclein for in vitro experiments

The pT7-7 plasmid containing the α-synuclein gene was transformed into E. coli Bl-21 (DE3) Gold cells (Strategene). Plasmid containing cells were selected with 0.1 mg/mL ampicillin. A 5 mL overnight culture was grown from a single colony and used to inoculate a 50 mL culture of Spectra 9 15N-enriched media (Cambridge Isotope Laboratories) at 37°C in a rotary shaker (225 rpm, New Burnswick Scientific, Model I-26). The saturated overnight culture was used to inoculate 1 L of M9 minimal media [21] containing 1 g/L 15NH4Cl. After reaching an absorbance at 600 nm (A600) of 0.8–1.0, the culture was induced with isopropyl β-D-thiogalactopyranoside (IPTG) to a final concentration of 1 mM. The culture was placed in the rotary shaker (225 rpm) at 37°C.

After 5 h the cultures were pelleted using a swinging bucket centrifuge (Sorvall RC-3B, H6000A rotor) at 1600g for 30 min at 4°C and the pellet was stored at −20°C. The pellet was resuspended in 30 mL of lysis buffer (50 mM Tris, 150 mM NaCl, 1 mM phenylmethanesulfonyl fluoride, 0.4 g/L lysozyme, pH 8.0). RNase and DNase were added to a final concentration of 0.02 g/L each. The samples were stirred (250 rpm) at 4°C for 20 min. The lysate was sonicated (Branson Ultrasonics, Fischer Scientific) continuously for 5 min, boiled in a water bath for 20 min, and then centrifuged at 13,000g for 30 min at 4°C (SS-34 rotor). The supernatant was subjected to streptomycin sulfate precipitation (10 g/L) and centrifuged for 30 min at 4°C. The supernatant was subjected to (NH4)2SO4 precipitation (361 g/L) and centrifuged again for 30 min at 4°C. The pellet was resuspended in 20 mM sodium phosphate buffer (pH 7.4) and dialyzed (Thermo Scientific, 3500 MWCO) overnight, with stirring at 4°C, against the same buffer.

The protein was further purified by anion exchange chromatography (GEHealtcare, Q Sepharose HiPrep 16/10 column) with a 0–1 M linear gradient of NaCl in 20 mM phosphate buffer (pH 7.4). Fractions were subjected to SDS-PAGE on an 18% gel with Commassie blilliant blue staining. Fractions containing α-synuclein were pooled and dialyzed against water overnight, with stirring, at 4°C. The protein was concentrated in a YM-3 Centricon filter (Millipore, MWCO 3500) using centrifugation at 1000g (SS-34 rotor) for 1 h at 4°C. The purity of the protein was determined by SDS-PAGE with Coomassie staining. The pure α-synuclein was lyophilized (Labconco) and stored at −20°C. The yield was 35–60 mg of pure α-synuclein per liter of saturated cell culture.

4.2 Cultivation of E. coli for in-cell NMR experiments

A 5 mL overnight culture was grown from a single colony and used to inoculate a 500-mL Erlenmeyer flask containing 50 mL of isotopically enriched media, as described above. After the culture reached an A600 of 0.8–1.0, the cells were induced with IPTG to a final concentration of 1 mM. Expression was allowed to proceed for 4 h. The cells were gently harvested by using the swinging bucket centrifuge for 30 min at 4°C. The pellet was resuspended in 1 mL of spent media.

4.3 Cultivation and encapsulation of E. coli for NMR bioreactor experiments

A 5 mL overnight culture was grown from a single colony as described above and used to inoculate 150 mL of Luria Broth (10 mg/mL Tryptone, 5 mg/mL yeast extract, 10 mg/mL NaCl) at 37°C. The culture was grown in the rotary shaker (225 rpm) to an A600 of 0.8–1.0. The cells were gently harvested in the swinging bucket centrifuge for 20 min at 4°C and resuspended in 1 mL of spent media. The resuspended cells were mixed with a 2% w/v alginate (Sigma) solution in 20 mM phosphate, 150 mM NaCl (pH 7.4) to give a final concentration of 1% alginate (50:50 mixture alginate:cell slurry).

The electrostatic encapsulation device (Fig. 7) comprised a 1 mL insulin syringe (BD), a 24 gauge winged anigocatheter (0.7 × 19 mm tip, Braun), a 23 gauge needle (BD), a syringe pump (Braintree Scientific 8000), and an adjustable high voltage power supply (Spellman SL10). The insulin syringe, equipped with the needle, was loaded with the cell/alginate mixture. The other needle, which was inserted horizontally through the center of the angiocatheter, was connected to the positive pole of the power supply. The negative pole of the power supply was placed into the 150 mM CaCl2 solution. The syringe containing the mixture was inserted into the top of the angiocatheter and placed onto the pump. The syringe pump was set to a rate of 0.714 mL/min, the power supply voltage to 3.35 kV, and the stir-plate to approximately 300 rpm. The tip of the angiocatheter was centered 1.2 cm above a 250 mL beaker containing 150 mL of 150 mM CaCl2. The mixture was forced through the tip of the angiocatheter and streamed into the CaCl2 solution. The Ca2+ polymerizes the alginate which, in turn, forms encapsulated beads containing the cells. The encapsulated cells were retrieved with suction and placed in a 15 mL Falcon tube containing 150 mM CaCl2 solution for transport to the NMR spectrometer.

Fig. 7
Schematic of the electrostatic encapsulation device. A: stir plate, B: beaker containing 150 mM CaCl2, C: anigocatheter, D: needle, E: insulin syringe, F: alligator clip (connects the positive power supply terminal to the needle) G: high voltage power ...

The CaCl2 solution was removed and the encapsulated cells were washed with the phosphate-free minimal medium. The phosphate-free minimal medium consisted of 100 mM HEPES (pH 7.4), 150 mM CaCl2, phosphate-free M9 salts [1mg/mL 15NH4Cl, 2 mM MgCl2, 1 μg/mL thiamine, 2 % v:v 10x 15N-enriched Bioexpress 1000 media (Cambridge Isotope Laboratories)] and 0.1 mg/mL ampicillin. After washing, the encapsulated cells were placed inside the bioreactor, which was then placed into the spectrometer. After acquiring the initial spectrum, lactose was added to a final concentration of 1% w/v. The lactose acts as an inducer and the sole carbon source. For each spectrum, the pump circulated medium through the system at a rate of 45 mL/min for 30 min. Five min were allotted for the encapsulated cells to settle into the detection region of the bioreactor. As a control the procedure was repeated for E. coli containing the pUC18 plasmid.

4.4 NMR

Data were acquired at the UNC Biomeolcular NMR facility on a Varian Inova 600 MHz NMR spectrometer. Data were processed and visualized with NMRpipe and NMRviewJ, respectively [22; 23].

Samples for dilute solution spectra comprised a 90:10 (v:v, pH 7.4) mixture of purified 200 μM α-synuclein solution: D2O in a standard 5 mm NMR tube. 1H-15N HSQC spectra were acquired at 10°C [1] with a 5 mm Varian Triax triple resonance probe (1H sweep width: 11990.40 Hz; 15N sweep width: 2100 Hz, 8 transients, 128 increments). Each spectrum required 35 min.

Samples for simple in-cell NMR experiments comprised 90:10 (v:v) mixture of resuspended cells: D2O in a standard 5 mm NMR tube. 1H-15N HSQC [24; 25] spectra were acquired as described above, except with 12 transients and 128 increments. Each spectrum required 1 h.

Samples for encapsulated in-cell NMR experiments comprised encapsulates in a 90:10 (v:v) mixture of 150 mM CaCl2: D2O in a standard 5 mm NMR tube. For encapsulates in standard 5 mm NMR tubes, 1H-15N HSQC spectra were acquired as described above. In the bioreactor, 1H-15N HSQC spectra were acquired unlocked at 37°C with an 8 mm modified Varian Triax triple resonance z gradient probe as described above. Samples for NMR bioreactor experiments comprised encapsulated cells in phosphate-free media supplemented with Bioexpress 1000. 1H-15N SOFAST HMQC spectra were acquired at 37°C as described above, except with 48 transients and 96 increments. The spectra were acquired unlocked due to the adverse effects of D2O on cell growth and protein expression. Each spectrum required 35 min.

Acknowledgments

We thank Peter Lansbury, Jr., for the α-synuclein expression system, Freddy Pinero of the UNC Chemistry Instrument Shop for fabricating the bioreactor and other small parts, Collin J. McKinney of the UNC Chemistry Electronics Facility for software support, Christopher Seagle for insightful input on the design of the bioreactor and the members of the Pielak laboratory for helpful comments on the manuscript.

Footnotes

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

References

1. McNulty BC, Tripathy A, Young GB, Charlton LM, Orans J, Pielak GJ. Temperature-induced reversible conformational change in the first 100 residues of alpha-synuclein. Protein Science. 2006;15:602–608. [PMC free article] [PubMed]
2. Charlton LM, Barnes CO, Li C, Orans J, Young GB, Pielak GJ. Residue-level Interrogation of macromolecular crowding effects on protein stability. Journal of the American Chemical Society. 2008;130:6826–6830. [PMC free article] [PubMed]
3. Ai X, Zhou Z, Bai Y, Choy WY. 15N NMR spin relaxation dispersion study of the molecular crowding effects on protein folding under native conditions. Journal of the American Chemical Society. 2006;128:3916–3917. [PubMed]
4. Homouz D, Perham M, Samiotakis A, Cheung MS, Wittung-Stafshede P. Crowded, cell-like environment induces shape changes in aspherical protein. proceedings of the National Academy of Sciences of the United States of America. 2008;105:11754–11759. [PMC free article] [PubMed]
5. Pielak GJ, Li C, Miklos AC, Schlesinger AP, Slade KM, Wang GF, Zigoneanu IG. Protein nuclear magnetic resonance under physiological conditions. Biochemsitry. 2009;48:226–234. [PMC free article] [PubMed]
6. Serber Z, Straub W, Corsini L, Nomura AM, Shimba N, Craik CS, Ortiz de Montellano P, Dötsch V. Methyl groups as probes for proteins and complexes in in-cell NMR experiments. Journal of the American Chemical Society. 2004;126:7119–7125. [PubMed]
7. Mancuso A, Glickson JD. Applications of NMR spectroscopy and imaging to the study of immobilised cell physiology. In: Hofman M, Anné J, editors. Fundamentals of cell immobilisation biotochenology. Kluwer Academic Publishers; 2004. pp. 439–467.
8. Freyer JP, Fink NH, Schor PL, Coulter JR, Neeman M, Sillerud LO. A system for viably maintaining a stirres suspension of multicellular spheriods during NMR spectroscopy. NMR in biomedicine. 1990;3:195–205. [PubMed]
9. Gillies RJ, Galons JP, McGovern KA, Scherer PG, Lien YH, Job C, Ratcliff R, Chapa F, Cerdan S, Dale BE. Design and application of NMR-compatible bioreactor circuits for extended perfusion of high-density mammalian cell cultures. NMR in Biomedicine. 1993;6:95–104. [PubMed]
10. Hartbrich A, Schmitz G, Weuster-Botz D, de Graaf AA, Wandrey C. Development and application of a membrane cyclone reactor for in-vivo NMR spectroscopy with high microbial cell densities. Biotechnology and Bioengineering. 1996;51:624–635. [PubMed]
11. Majors PD, McLean JS, Scholten JCM. NMR bioreactor development for live in-situ microbial functional analysis. Journal of Magnetic Resonance. 2008;192:159–166. [PubMed]
12. Dahan-Grobgeld E, Livneh Z, Maretzek AF, Polak-Charcon S, Eichenbaum Z, Degani H. Reversible induction of ATP synthesis by DNA damage and repair in Escherichia coli. In-vivo NMR studies. Journal of Biological Chemistry. 1998;273:30232–30238. [PubMed]
13. Hesse SJA, Ruijter GJG, Dijkema C, Visser J. Measurement of intracellular (compartmental) pH by 31P NMR in Aspergillus niger. Journal of Biotechnology. 2000;77:5–15. [PubMed]
14. Lang AE, Lozano AM. Parkinson’s disease- first of two parts. The New England Journal of Medicine. 1998;339:1044–1053. [PubMed]
15. Schanda P, Brutscher B. Very fast two-dimensional NMR spectroscopy for real-time investigation of dynamic events in proteins on the time scale of seconds. Journal of the American Chemical Society. 2005;127:8014–8015. [PubMed]
16. Dulieu C, Poncelet D, Neufeld RJ. Encapsulation and immobiliation techniques. In: Kuthreiber WM, Lanza RP, Chick WL, editors. Cell encapsulation technology and therapeutics. Birkhauser; Boston: 1999. pp. 3–17.
17. McNulty BC, Young GY, Pielak GP. Macromolecular crowding in the Escherichia coli periplasm maintains a-synuclein disorder. Journal of Molecular Biology. 2006;355:893–897. [PubMed]
18. Li C, Charlton LM, Lakkavaram A, Seagle C, Wang G, Young GB, Macdonald JM, Pielak GJ. Differential dynamical effects of macromolecular crowding on an intrinsically disordered protein and a globular protein: implications for in-cell NMR spectroscopy. Journal of the American Chemical Society. 2008;130:6310–6311. [PMC free article] [PubMed]
19. Slade KM, Baker R, Chua M, Thompson NL, Pielak GJ. Effects of Recombinant Protein Expression on Green Fluorescent Protein Diffusion in Escherichia coli. Biochemistry. 2009;48:5083–5089. [PubMed]
20. Vieira J, Messing J. The pUC plasmids, an M13mp7-derived system for insertion mutagenesis and sequencing with synthetic universal primers. Gene. 1982;19:259–268. [PubMed]
21. Serber Z, Ledwidge R, Miller SM, Dötsch V. Evaluation of parameters critical to observing proteins inside living Escherichia coli by in-cell NMR spectroscopy. journal of the American Chemical Society. 2001;123:8895–8901. [PubMed]
22. Delaglio F, Grzesiek S, Vuister GW, Zhu G, Pfeifer J, Bax A. NMRPipe: A multidimensional spectral processing system based on UNIX pipes. Journal of Biomolecular NMR. 1995;6:277–293. [PubMed]
23. Johnson BA, Blevins RA. NMR View: A computer program for the visualization and analysis of NMR data. Journal of Biomolecular NMR. 1994;4:603–614. [PubMed]
24. Bodenhausen G, Ruben DJ. Natural abundance nitrogen-15 NMR by enhanced heteronuclear spectroscopy. Chemical Physics Letters. 1980;69:185–189.
25. Kay L, Keifer P, Saarinen T. Pure absorption gradient enhanced heteronuclear single quantum correlation spectroscopy with improved sensitivity. Journal of the American Chemical Society. 1992;114:10663–10665.