Copyright © 2009 Molony et al; licensee BioMed Central Ltd. Fluid-structure interaction of a patient-specific abdominal aortic aneurysm treated with an endovascular stent-graft 1Centre for Applied Biomedical Engineering Research (CABER), Department of Mechanical and Aeronautical Engineering and Materials and Surface Science Institute, University of Limerick, Ireland 2Department of Vascular Surgery, Mid-Western Regional Hospital, Limerick, Ireland Corresponding author.#Contributed equally. David S Molony: david.molony/at/ul.ie; Anthony Callanan: anthony.callanan/at/ul.ie; Eamon G Kavanagh: eamon.kavanagh/at/hse.ie; Michael T Walsh: michael.walsh/at/ul.ie; Tim M McGloughlin: tim.mcgloughlin/at/ul.ie Received April 3, 2009; Accepted October 6, 2009. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. Abstract Background Abdominal aortic aneurysms (AAA) are local dilatations of the infrarenal aorta. If left untreated they may rupture and lead to death. One form of treatment is the minimally invasive insertion of a stent-graft into the aneurysm. Despite this effective treatment aneurysms may occasionally continue to expand and this may eventually result in post-operative rupture of the aneurysm. Fluid-structure interaction (FSI) is a particularly useful tool for investigating aneurysm biomechanics as both the wall stresses and fluid forces can be examined. Methods Pre-op, Post-op and Follow-up models were reconstructed from CT scans of a single patient and FSI simulations were performed on each model. The FSI approach involved coupling Abaqus and Fluent via a third-party software - MpCCI. Aneurysm wall stress and compliance were investigated as well as the drag force acting on the stent-graft. Results Aneurysm wall stress was reduced from 0.38 MPa before surgery to a value of 0.03 MPa after insertion of the stent-graft. Higher stresses were seen in the aneurysm neck and iliac legs post-operatively. The compliance of the aneurysm was also reduced post-operatively. The peak Post-op axial drag force was found to be 4.85 N. This increased to 6.37 N in the Follow-up model. Conclusion In a patient-specific case peak aneurysm wall stress was reduced by 92%. Such a reduction in aneurysm wall stress may lead to shrinkage of the aneurysm over time. Hence, post-operative stress patterns may help in determining the likelihood of aneurysm shrinkage post EVAR. Post-operative remodelling of the aneurysm may lead to increased drag forces. Background Abdominal aortic aneurysm (AAA) is a localized disease of the abdominal aorta. This dilatation of the infrarenal aorta has been found to affect 8.9% of the population over age 65 [1]. If the aneurysm continues to expand it may eventually rupture. Currently the decision to operate is based solely on the diameter of the aneurysm. When the maximum diameter exceeds 55 mm or the expansion of the aneurysm is greater than 10 mm/year surgery is normally recommended. The traditional approach to treating this disease is termed open repair, which involves an incision in the abdomen and the exclusion of the diseased aneurysm with a synthetic graft. An alternative treatment is endovascular aneurysm repair (EVAR). This is the minimally invasive technique of inserting a stent-graft into the aneurysm site via the femoral and iliac arteries. The function of the stent-graft is to shield the aneurysm from the systemic blood pressure. Complications associated with EVAR include graft migration and endoleak. The occurrence of these complications may result in continued expansion of the aneurysm and possible rupture after EVAR [2,3]. Clinically it has been shown that aneurysms can still expand after EVAR without the presence of endoleak [4]. This phenomenon has been termed "endotension". Although such complications have been shown to be relatively rare, ruptures have been reported [5]. Numerical modelling of aneurysms and stent-grafts is a useful method for determining the stresses and forces seen in-vivo. Finite element analysis (FEA) allows the stresses on the aneurysm wall to be determined [6,7]. This method has shown that peak wall stress is more commonly associated with aneurysm rupture than maximum diameter [6]. Computational fluid dynamics (CFD) allows for the investigation of flow patterns and drag forces [8]. Drag forces on stent-grafts, which may influence stent-graft migration, have been investigated by several authors [9-11]. Fluid-structure interaction (FSI) of aneurysms, though still an emerging field of study has seen an increasing number of publications in recent years and with improved computer power and software capability this is expected to grow. There have been several patient-specific FSI simulations of aneurysms. These have taken into account non-linear wall models [12], iliac bifurcations [13] and intraluminal thrombus (ILT) [14,15]. However, there has been little work published on the effect of a stent-graft on aneurysm wall stress. Furthermore, only a few authors have studied patient-specific cases, two of these were flow studies only [10,16], while another studied the effect of the stent-graft on the wall mechanics [17]. Thus, these studies ignored fluid structure interactions effects. Previous fluid-structure interaction in 3D representative models have ignored the presence of ILT and considered the sac to be composed of stagnant blood [18-20]. This assumption may be imprecise as it has been reported that 75% of all AAAs contain ILT [21]. Also, this work did not assess the impact of remodelling of the aneurysm over time. Nonetheless, these studies have shown that the stent-graft can reduce peak aneurysm wall stress 20 fold. The objective of this work was to investigate the biomechanics in a Pre-op, Post-op and Follow-up patient-specific abdominal aortic aneurysm with the use of fluid-structure interaction. Non-linear material models were used for the aneurysm and thrombus and the resulting wall mechanics were calculated. Finally the drag force on the stent-graft was also determined. Methods Geometry CT scan data from a single patient was obtained from the Midwestern Regional Hospital, Limerick in DICOM format. The study conformed to the Declaration of Helsinki, and was approved by the local research ethics committee. The patient gave written informed consent. The CT data consisted of a pre-op scan and a 6 month follow-up scan. Using Mimics 12.0 (Materialise, Belgium) 3D models were reconstructed from this data. Images were segmented from just below the infrarenal aorta to the bifurcation of the iliac arteries. The reconstruction technique has been described in previous work from our group [22]. Three models in total were created, and will be termed Pre-op, Post-op and Follow-up from here on (figure (figure1).1
Due to difficulties encountered during smoothing of the model in Mimics and also in order to achieve a tight contact between the stent-graft and aneurysm surfaces, the geometry in the form of polylines was exported as an iges file to Pro/Engineer (PTC, NH). Next, surfaces were created along the polylines and the outer aneurysm surface was offset 1.5 mm in order to give the aneurysm a suitable wall thickness. These surfaces were then exported to Gambit (Ansys, Canonsburg, PA) for meshing. Here both the fluid (lumen) and structure (aneurysm wall, thrombus and graft wall) domains are meshed. The meshes are then imported to Fluent 6.3.26 (Ansys, Canonsburg, PA) where the solid region can be exported to Abaqus 6.7-1 (Abaqus Inc, Pawtucket, RI). FSI FSI is normally achieved either through a monolithic (full coupling) or partitioned approach (loose coupling) [23]. The commercial software MpCCI 3.0.6 (Fraunhofer SCAI, Germany) was used in this work. The software is based on the loose coupling of two chosen softwares. This approach allows the use of familiar and mature solvers for each domain. The benefits of using mature solvers are advanced capabilities that may not be available in monolithic solvers such as non-linear material models, contact and non-Newtonian fluid models. For this research Abaqus, for the structural component, and Fluent, as the fluid solver, were coupled. The fluid time step was set to 0.001 s and data exchange occurred every 0.005 s, with Fluent sending the pressure to Abaqus, and Abaqus sending the deformed nodal co-ordinates to Fluent. Fluent uses the Arbitrary Lagrangian Eulerian (ALE) method to deal with the deforming mesh. A remeshing technique is used where cells are remeshed based on whether they violate a user specified size and skewness criteria. Specifically, the tetrahedral cells were remeshed if their size varied by more than 3% of the lower and upper size criteria. Both codes share a common boundary where the data exchange occurs. MpCCI identifies nodes or elements near each other based on an association scheme and data is then transferred from one node to the other. The software allows for non-matching meshes. Further information is available in the MpCCI documentation [24]. Structural Model The ILT and AAA wall were assumed to be hyperelastic, homogenous, incompressible and isotropic. The AAA wall and ILT were modelled using the constitutive model proposed by Raghavan and Vorp [25] and Wang et al. [26] respectively. The strain energy functions are given as:
where C01 = 0.174 MPa and C02 = 1.881 MPa are population mean values for the aneurysm wall; C10 = 0.026 MPa and C20 = 0.026 MPa are population mean values derived for the ILT. IB and IIB are the first and second invariants of the left Cauchy-Green tensor B respectively. It has been shown that the use of population mean values does not significantly affect the prediction of stresses on the aortic wall [25]. The aneurysm and ILT were assigned values of 1120 kg/m3 and 1121 kg/m3 for the structural density respectively [27]. The stagnant blood in the aneurysm sac was assigned the same density as the lumen blood. The stent-graft was modelled as one whole body due to the difficulty in accurately reconstructing the nitinol stents from the CT data. A Young's Modulus of 10 MPa and a density of 6000 kg/m3 were assumed for the stent-graft [27]. The graft and artery were assumed to be tied together, simulating attachment of the stent-graft to the artery wall, thus ignoring the possibility of endoleaks and local dislodgement. The Pre-op and Post-op models had the same mesh for the aneurysm and thrombus. This consisted of 75,849 tetrahedral elements while the Post-op model had an additional 19,024 hexahedral elements for the stent-graft. The space between the ILT and stent-graft was modelled with 21,974 hydrostatic pressure elements in Abaqus. These simulated the stagnant blood in the aneurysm sac. The Follow-up model contained 26,368 hexahedral elements for the stent-graft and 64,350 tetrahedral elements for the aneurysm and ILT. For all models the aneurysm inlet and outlets were constrained in all degrees-of-freedom as shown in figure figure2.2
Fluid Model At the inlet a velocity boundary condition was assigned while at the outlets a pressure boundary was assigned (figure (figure2).2 Mesh independence Mesh independence was carried out separately on the solid and fluid meshes. The solid and fluid meshes were declared independent when the peak stress and wall force integral did not change by more than ± 2% between successive meshes. Pulse cycle independence was achieved after four cardiac cycles. CFD convergence criteria for mass and momentum residuals were 1 × 10-4 and 1 × 10-5 respectively. Simulations were performed on a 64 bit Dell Precision T7400 (2.99 GHz with 16 GB RAM) using 2 processors and one cardiac cycle took 40 hours. In order to achieve further confidence in the accuracy of the results comparisons were made to clinical observations. Vorp et al. [30] found a compliance value of 4 × 10-4/mmHg in a clinical study of aneurysms. A similar value of 3.6 × 10-4/mmHg was seen in our study for the Pre-op model. Clinical investigations have shown reduced wall motion in aneurysms after EVAR Malina et al. [31] noted that the pulsatile wall motion was reduced by 75% after EVAR. Results in this study followed a similar trend with a reduction in compliance of between 80 and 96%. Results Wall Stress The von Mises wall stress on the Pre-op, Post-op and Follow-up models are shown in figure figure3.3
A histogram illustrates the effect of EVAR in redistributing the wall stress over the aneurysm volume (figure (figure5).5
Compliance Compliance is a tool used to describe the distensibility of an artery. The compliance of the aneurysm (figure (figure7)7
where Amax and Amin are the measured area of interest corresponding to systolic pressure (Pmax) and diastolic pressure (Pmin) respectively. Compliance of the aneurysm was reduced post-operatively (figure (figure7).7
Velocity pathlines and pressure During peak systolic pressure the flow of blood has begun to decelerate, which can lead to vortex formation [16]. The removal of a vortex after the graft has been implanted can be seen in figure figure8.8
Drag Force The drag force is caused by pressure and viscous forces acting on the stent-graft [10]. Peak drag force in the Post-op and Follow-up models occurred prior to the peak systolic pressure with a value of 4.85 N and 6.37 N respectively. In the Post-op model the drag force varied from a value of 2.85 N to 4.85 N over the cardiac cycle, while in the Follow-up model the drag force varied between 3.75 N and 6.37 N. Both the viscous force and pressure force followed the same trend as the velocity and pressure waveform respectively (figure (figure9).9
Influence of stent-graft Young's Modulus In order to determine the influence of the stent-graft stiffness the Post-op simulations were performed again for a Young's Modulus of 5 MPa and 15 MPa (table 4). The peak aneurysm wall stress decreased for increasing graft stiffness. Similarly the compliance of the aneurysm decreased for an increase in graft stiffness. No significant changes were seen in the fluid flow. There was a small reduction in drag force for increasing graft stiffness.
Discussion Fluid-structure interaction of a Pre-op, Post-op and Follow-up AAA was simulated. The remodelling of the aneurysm over a 6 month period was taken into account. Previously reported parameters such as wall stress and drag force were examined as well as the additional parameter of compliance. Most previous work in this field has investigated numerous parameters in a representative model based on CT scans [18-20,27]. It was observed that peak wall stress is reduced 20 fold in a pre- and post-op AAA [18]. In a patient specific FEA model a 90% decrease in aneurysm wall stress was reported [17]. Our results compare favourably to both these studies. The peak von Mises wall stress on the aneurysm was reduced by 92% (12 fold) after EVAR in the Post-op model. In the Pre-op case higher stresses are seen in the aneurysm region than in the neck region. The opposite is seen in the Post-op and Follow-up cases. This could be seen as an indication of the success of EVAR i.e. low post-operative stress on the aneurysm sac wall. The high stress on the aneurysm neck is due to the contact of the stent-graft with the aneurysm wall. The higher stress here may cause the aneurysm neck to expand which can lead to an increase in drag force [10]. From the Follow-up model it was noted that the proximal neck diameter increased from 24.8 mm to 27.2 mm. Neck enlargement after EVAR has been reported extensively [32]. Arterial remodelling occurs when local wall stress deviates from a reference value [33]. After EVAR there is normally a reduction in aneurysm volume and diameter [34]. Previous research has shown the peak wall stress in the normal abdominal aorta to be in the region of 0.225 MPa [35]. As can be seen in (figure 6a, b Qualitatively, the wall stress in the Post-op and Follow-up models were very similar (figure (figure3).3 The incorporation of the ILT played an essential role in the stress distribution across the model. The area between the stent-graft and aneurysm bulge is encased by a large volume of thrombus (figure (figure4).4 The compliance of the aneurysm refers to the compliance of the overall post-operative environment, which is dependent on the stent-graft, ILT, stagnant blood and aneurysm. If the aneurysm alone was to be considered, the elasticity and hence compliance have not actually altered from the pre-operative case. The only alteration has been to the overall environment, with the resulting reduction in compliance. An alternative method would have been to use the sac pressure but this would not have allowed us to make a direct comparison between the pre-op, post-op and follow-up models. Due to the ILT being a solid entity, there is no sac pressure in the follow-up model. Furthermore, clinically, intra-sac pressure measurements are not routinely recorded. The percentage difference between Pre-op and Post-op compliance was greatest at levels 2-4 (table 2). At level 1 the difference was much smaller. At this location the stent-graft pushes on the aneurysm neck causing it to deform, resulting in a post-op compliance closer to the pre-op compliance. There was little difference in compliance between the Post-op and Follow-up models. The overall reduction in aneurysm volume in the Follow-up case is not reflected by any significant alterations in aneurysm compliance. The shape of the pre-op lumen and post-op stent-graft are quite similar. This is due to the large volume of ILT narrowing the pre-op lumen. Despite this similarity the stent-graft had smoother pathlines and noticeably in one region vortices were removed after the implantation of the stent-graft. Greater flow disturbances are likely in pre-op patients with less ILT, as when the lumen expands into the aneurysm sac vortices will develop in the pockets created. In the Follow-up model, the proximal neck has straightened due to remodelling and this may further improve the flow of blood. As has been seen in previous studies the vast majority of the drag force magnitude is due to the pressure component [11]. The downward axial force is more likely to be associated with caudal graft migration than forces in the radial direction. The maximum axial drag force for the Post-op model was 4.85 N. The drag force increased in the Follow-up model to 6.37 N. This may be attributed to the greater neck diameter of the follow-up aneurysm, as it has been shown that grafts with greater proximal neck diameters result in a larger drag force. Previously Morris et al. found that in similar sized grafts with neck diameters of 25-27 mm and without significant curvature the drag force was approximately 5.5-6.5 N [11]. Bench-top testing has shown that forces ranging from 4.5 N to 40 N can dislodge stent-grafts [36,37]. The stiffness of the stent-graft was found to have little influence on the fluid domain. A negligible change in drag force of less than 1% was noted, which suggests that modelling the stent-graft as rigid may be an acceptable assumption if just the fluid forces are of interest. A change in the stent-graft stiffness had a greater influence on the compliance and aneurysm wall stress. Stiffer stent-grafts will reduce the stress and deformation of the aneurysm wall. Limitations There are several limitations to the current study. The Post-op model assumes that there was no immediate remodelling of the pre-operative aneurysm due to the implantation of the stent-graft. The patient was selected from our patient database as the aneurysm geometry had not changed greatly over the 6 month follow-up period. Despite this, there was some remodelling of the aneurysm as can be seen by the straightening of the aneurysm neck and the formation of thrombus in the stent-graft lumen (figure (figure1).1 Stent-grafts are oversized by 10-20% normally [3], but because the model is reconstructed from CT scans it is already in a state of stress so the unloaded configuration is unavailable. Hence, oversizing of the stent-graft was ignored. The stent-graft was modelled as a single body due to the difficulty in reconstructing the stents. Stents are radiopaque and so show up as high intensity pixels in DICOM images and can contain a lot of noise. This creates difficulties in reconstruction of the geometry as the stent-graft boundary can be difficult to identify. The wall thickness of the aneurysm could not be determined from the CT images. Furthermore, it was not possible to obtain the flow rate and blood pressure for the patient, as these are not routine measurements for AAA patients. The tied contact between the stent-graft and AAA may not simulate the actual contact in-vivo. Stent-graft fixation is due to the radial force and friction exerted by the stent-graft on the aneurysm neck. A non-stick contact would have modelled this more appropriately but would also have significantly increased the simulation time. A loose coupling method was employed for the simulations, though in reality strong coupling exists between the fluid and solid in blood flow in arteries [23]. This is due to similar densities of the fluid and solid. A strong coupling method may better simulate the fluid-structure interactions. It has been shown through biaxial tensile testing that AAA behaves as an anisotropic material. The use of anisotropic properties has been shown to result in higher peak stresses than isotropic properties [28,38]. Similarly, calcifications in the aortic wall were neglected and these have been shown to result in higher peak stresses [39]. Conclusion To the best of the authors' knowledge this is the first fluid-structure interaction study of pre- and post-operative patient-specific AAA biomechanics. Aneurysm remodelling 6 months after implantation of the stent-graft was investigated. The peak wall stress on the aneurysm wall was reduced by 92% after EVAR. Compliance of the aneurysm is significantly reduced after EVAR. Aneurysm remodelling after EVAR may lead to an increase in stent-graft drag force. Hence, despite the success of EVAR in reducing aneurysm wall stress, patient follow-up is essential in order to determine the consequences of remodelling of the aneurysm. Abbreviations AAA: Abdominal aortic aneurysm; CT: Computed tomography; FSI: Fluid-structure interaction; EVAR: Endovascular aneurysm repair; FEA: Finite Element Analysis; CFD: Computational Fluid Dynamics; ILT: Intraluminal thrombus; 3D: Three-dimensional; DICOM: Digital Imaging and Communications in Medicine; ALE: Arbitrary Lagrangian Eulerian; PISO: Pressure Implicit with Splitting of Operators. Competing interests The authors declare that they have no competing interests. Authors' contributions DSM reconstructed the models, conducted the simulations, analyzed the results and drafted the manuscript. AC analyzed the results and revised the manuscript. MTW analyzed the results, revised and gave final approval of the manuscript. EGK acquired and analyzed the CT data and revised the manuscript. TMM analyzed the results, revised and gave final approval of the manuscript. All authors read and approved the final manuscript. Acknowledgements We would like to acknowledge (i) The Midwestern Regional Hospital, Limerick for provision of CT data (ii) Enterprise Ireland grant no. CFTD/05/121 References
|
PubMed related articles
Your browsing activity is empty. Activity recording is turned off. |
|||||||||||||||||||||||||||||||||||
Ann Intern Med. 2001 Feb 6; 134(3):182-90.
[Ann Intern Med. 2001]J Interv Cardiol. 2004 Dec; 17(6):437-45.
[J Interv Cardiol. 2004]J Endovasc Ther. 2008 Aug; 15(4):468-84.
[J Endovasc Ther. 2008]J Endovasc Ther. 2008 Dec; 15(6):702-5.
[J Endovasc Ther. 2008]J Vasc Surg. 2005 Aug; 42(2):194-9.
[J Vasc Surg. 2005]J Vasc Surg. 2002 Sep; 36(3):589-97.
[J Vasc Surg. 2002]J Vasc Surg. 2009 Feb; 49(2):443-54.
[J Vasc Surg. 2009]Med Eng Phys. 2006 Jan; 28(1):19-26.
[Med Eng Phys. 2006]J Endovasc Ther. 2007 Apr; 14(2):138-43.
[J Endovasc Ther. 2007]J Biomech. 2006; 39(12):2264-73.
[J Biomech. 2006]Biomed Eng Online. 2006 May 19; 5():33.
[Biomed Eng Online. 2006]Med Eng Phys. 2005 Dec; 27(10):871-83.
[Med Eng Phys. 2005]Med Eng Phys. 2001 Nov; 23(9):647-55.
[Med Eng Phys. 2001]Biomed Eng Online. 2005 Nov 4; 4():64.
[Biomed Eng Online. 2005]J Biomech. 2004 Jul; 37(7):1087-95.
[J Biomech. 2004]J Endovasc Ther. 2008 Oct; 15(5):518-29.
[J Endovasc Ther. 2008]Med Eng Phys. 2005 Jun; 27(5):369-82.
[Med Eng Phys. 2005]J Biomech Eng. 2006 Feb; 128(1):59-68.
[J Biomech Eng. 2006]J Ultrasound Med. 1982 Oct; 1(8):315-8.
[J Ultrasound Med. 1982]Biomed Eng Online. 2007 Oct 19; 6():38.
[Biomed Eng Online. 2007]J Biomech. 2000 Apr; 33(4):475-82.
[J Biomech. 2000]J Biomech Eng. 2001 Dec; 123(6):536-9.
[J Biomech Eng. 2001]J Biomech. 2000 Apr; 33(4):475-82.
[J Biomech. 2000]J Biomech Eng. 2005 Aug; 127(4):662-71.
[J Biomech Eng. 2005]J Vasc Surg. 2009 Feb; 49(2):443-54.
[J Vasc Surg. 2009]Ann Biomed Eng. 2008 Jun; 36(6):921-32.
[Ann Biomed Eng. 2008]Cardiovasc Res. 1970 Oct; 4(4):405-17.
[Cardiovasc Res. 1970]J Endovasc Ther. 2008 Oct; 15(5):518-29.
[J Endovasc Ther. 2008]Cardiovasc Surg. 1996 Dec; 4(6):732-9.
[Cardiovasc Surg. 1996]J Vasc Surg. 1998 Apr; 27(4):624-31.
[J Vasc Surg. 1998]Cardiovasc Surg. 1996 Dec; 4(6):732-9.
[Cardiovasc Surg. 1996]J Endovasc Ther. 2008 Oct; 15(5):518-29.
[J Endovasc Ther. 2008]J Biomech. 2004 Jul; 37(7):1087-95.
[J Biomech. 2004]Med Eng Phys. 2005 Jun; 27(5):369-82.
[Med Eng Phys. 2005]J Biomech Eng. 2006 Feb; 128(1):59-68.
[J Biomech Eng. 2006]J Biomech Eng. 2005 Aug; 127(4):662-71.
[J Biomech Eng. 2005]J Biomech. 2004 Jul; 37(7):1087-95.
[J Biomech. 2004]Ann Vasc Surg. 2008 Jul-Aug; 22(4):559-63.
[Ann Vasc Surg. 2008]Med Biol Eng Comput. 2008 Nov; 46(11):1121-7.
[Med Biol Eng Comput. 2008]J Vasc Surg. 2001 Feb; 33(2 Suppl):S1-10.
[J Vasc Surg. 2001]Med Hypotheses. 2005; 65(6):1191-5.
[Med Hypotheses. 2005]Med Eng Phys. 2005 Jun; 27(5):369-82.
[Med Eng Phys. 2005]J Biomech Eng. 2006 Feb; 128(1):59-68.
[J Biomech Eng. 2006]Biomed Eng Online. 2007 Oct 19; 6():38.
[Biomed Eng Online. 2007]J Biomech Eng. 2001 Dec; 123(6):536-9.
[J Biomech Eng. 2001]J Biomech. 2006; 39(12):2264-73.
[J Biomech. 2006]Eur J Vasc Endovasc Surg. 2000 Aug; 20(2):190-5.
[Eur J Vasc Endovasc Surg. 2000]J Endovasc Ther. 2003 Dec; 10(6):1101-9.
[J Endovasc Ther. 2003]J Endovasc Ther. 2008 Aug; 15(4):468-84.
[J Endovasc Ther. 2008]Ann Biomed Eng. 2008 Jun; 36(6):921-32.
[Ann Biomed Eng. 2008]J Biomech Eng. 2009 Feb; 131(2):021009.
[J Biomech Eng. 2009]J Biomech Eng. 2007 Feb; 129(1):1-11.
[J Biomech Eng. 2007]