Abstract

Genotyping by high-resolution melting analysis of small amplicons is homogeneous and simple. However, this approach can be limited by physical and chemical components of the system that contribute to intersample melting variation. It is challenging for this method to distinguish homozygous G::C from C::G or A::T from T::A base-pair neutral variants, which comprise ~16% of all human single nucleotide polymorphisms (SNPs). We used internal oligonucleotide calibrators and custom analysis software to improve small amplicon (42–86 bp) genotyping on the LightScanner®. Three G/C (PAH c.1155C>G, CHK2 c.1-3850G>C and candidate gene BX647987 c.261+22,290C>G) and three T/A (CPS1 c.3405-29A>T, OTC c.299-8T>A and MSH2 c.1511-9A>T) human single nucleotide variants were analyzed. Calibration improved homozygote genotyping accuracy from 91.7 to 99.7% across 1105 amplicons from 141 samples for five of the six targets. The average Tm standard deviations of these targets decreased from 0.067°C before calibration to 0.022°C after calibration. We were unable to generate a small amplicon that could discriminate the BX647987 c.261+22,290C>G (rs1869458) SNP, despite reducing standard deviations from 0.086°C to 0.032°C. Two of the sites contained symmetric nearest neighbors adjacent to the SNPs. Unexpectedly, we were able to distinguish these homozygotes by Tm even though current nearest neighbor models predict that the two homozygous alleles would be identical.

INTRODUCTION

Genotyping by melting traditionally relies on the melting temperature (Tm) of a labeled probe hybridized to an amplification product (1–3). The primary advantage of probe-based genotyping is that the Tm between the probe and the target sequences in the two alleles usually varies by 2–8°C and is easily detected by standard methods. This large Tm separation occurs with probes of 15–35 bases that are perfectly matched to one allele.

High-resolution melting analysis of amplicons is an attractive genotyping method because it eliminates the need for oligonucleotide probes (4,5). Only two standard PCR primers are used and no sample processing is required after amplification is begun. While heterozygotes are easily detected, discriminating between homozygous alleles is harder because the homozygotes display similar melting curves and Tm. For example, the homozygote of the common cystic fibrosis mutation F508del could not be distinguished from the wild-type allele using a 277-bp amplicon (6). Melting data are additionally confounded in microtiter plate-based systems by well-to-well variations due to hardware (instrument and plates) and chemistry factors (7–9).

High-resolution melting of small amplicons (~40–90 bp) improves homozygote detection sensitivity because Tm differences are greater compared to larger amplicons (4). Small amplicon melting clearly resolves homozygous alleles in at least 84% of human single nucleotide variants where one allele is an A::T pair and the other is a G::C pair, resulting in Tm differences of 0.8–1.4°C (10). It is more challenging to use small amplicon melting to genotype the ~16% of single base variants that are base-pair neutral when the GC content remains the same. In these G/C or A/T variants, Tm values are predicted to change by 0.4°C or less (10). Furthermore, current nearest neighbor analysis (11) predicts no Tm difference in 4% of human single base variants with symmetric nucleotides flanking a base-pair neutral change, for example,

5′-… TGA …-3′ changing to 5′-… TCA …-3′

3′-… ACT …-5′ 3′-… AGT …-5′.

The homozygous alleles of HFE 187C>G contain these specific nearest neighbor symmetric sequences and were not distinguishable using small amplicon melting without mixing in a known homozygous control sample (12). In such a mixture, if only homoduplexes are observed, the genotype of the test sample is identical to that of the known standard. In contrast, if a heterozygous melting curve is obtained, the genotype of the test sample is different from that of the known sample.

Internal oligonucleotide calibrators can improve genotyping by melting. Molecular beacons have been used to normalize Tm across a microfluidic platform (13) and to minimize SYBR® Green I Tm variation across a 96-well plate (14). The use of two double-stranded oligonucleotide internal calibrators flanking a target amplicon improved genotyping of nonbase-pair neutral variants (i.e. G or C to A or T variants) in amplicons on both plate and capillary platforms (15,16). In each case, custom software normalized the Tm of the internal oligonucleotide calibrators across reactions for more accurate classification of genotypes.

Herein, we report the genotyping of six base-pair neutral human variants using small amplicon melting with two internal oligonucleotide calibrators on the 96-well LightScanner. Three of the variants consist of A::T to T::A exchanges located in the carbamyl phosphate synthetase gene (CPS1 c.3405-29A>T), the ornithine transcarbamylase gene (OTC c.299-8T>A) and the mutator homolog gene (MSH2 c.1511-9A>T). The variant sites in CPS1 and OTC are nearest neighbor symmetric. The other three variants consist of G::C to C::G variants located in the phenylalanine hydroxylase gene (PAH c.1155C>G), the checkpoint 2 gene (CHK2 c.1-3850G>C) and the candidate BX647987 gene (c.261+22,290C>G).

MATERIALS AND METHODS

Human genomic DNAs

One hundred and forty-one human genomic DNA samples were obtained from healthy donors and de-identified according to an institutional review board approved protocol #7275. The DNA was isolated from whole blood using the PUREGENE® Genomic DNA Purification Kit from Gentra systems (Minneapolis, Minnesota, MN USA). After extraction, samples were adjusted to a final concentration of 10–15 ng/ul.

The genotypes of all DNA samples were confirmed using probe-based assays (3). In addition, representative samples of each genotype were sequenced from longer OTC and CPS1 PCR products to confirm genotype.

Calibrators

Amplicon melting data were aligned (i.e. calibrated) relative to internal oligonucleotide calibrators of low (Tm ~62°C) and high (Tm ~92°C) Tm, henceforth referred to as low and high calibrators. Calibrators included in each reaction vessel provided PCR-independent melting signatures. The sequence of the low calibrator was 5′-TTAAATTATAAAATATTTATAATATTAATTATATATATATAAATATAATA-C3-3′. The sequence of the high calibrator was 5′-GCGCGGCCGGCACTGACCCGAGACTCTGAGCGGCTGCTGGAGGTGCGGAAGCGGAGGGGCGGG-C3-3′. These oligonucleotides, and their reverse complements, were included at 0.05 μM in each amplification reaction.

Importantly, all calibrators were blocked on their 3′-hydroxyl termini with a three-carbon (C3) alkyl group during synthesis to prevent extension by Taq polymerase. Fluorescent probes blocked with such alkyl moieties resist nucleotide extension by Taq polymerase after multiple freeze–thaw cycles and/or long-term storage better than phosphorylated probes (17).

Calibrators were obtained from IDT (Integrated DNA Technologies, Coralville, IA, USA) using standard phosphoramidite synthesis. After standard synthesis, PAGE purification was used to obtain higher oligonucleotide purity. Cartridge purification and reverse HPLC purification were also assessed with less optimal results.

Primer synthesis and primer design

Amplification primers were obtained from the University of Utah (Salt Lake City, UT, USA) Core Oligonucleotide/Peptide Synthesis Facility using standard phosphoramidite synthesis.

Primer sequences were analyzed (http://genome.ucsc.edu/cgi-bin/hgPcr) to minimize the likelihood that undesired products would co-amplify and interfere with the target sequence melting curves. Table 1 shows the variants, PCR primer sequences, concentrations and amplicon lengths. Some of the small amplicon PCR primers were designed to tightly flank the variants of interest leaving only 3–7 bases, including the polymorphism, between the primers. This is optimal for homozygote detection. However, the forward primer of the 68-bp OTC amplicon does not tightly flank the variant. This is due to the 7-bp long polythymidine tract followed by a cytidine and another 10-bp polythymidine tract immediately preceding the c.3405-29A/T variant. Other primers were constructed with up to 43 bp between each primer, for a total amplicon length reaching 98 bp.

Table 1.
Genetic variants assessed and amplification primers utilized

PCR conditions and high-resolution melting

PCR was performed in 10 μl using 1X LightScanner® High Sensitivity Master Mix (Idaho Technology), containing LCGreen® Plus, and 0.05 μM of each internal oligonucleotide calibrator. In addition, 0.10 μM or 0.15 μM of primers (depending on target) and 10–15 ng of human genomic DNA were included in each reaction. PCRs were performed on iCyclers (Bio-Rad, Hercules, CA, USA) with a final MgCl2 concentration of 2 mM. PCR conditions were as follows: an initial denaturation of 95°C for 2 min, followed by 44–50 cycles (depending on the assay) of 94°C for 30 s and annealing at 63–67°C (depending on the assay) for 30 s. The exact thermocycling conditions are shown in Table 2. After completing amplification, a final denaturation and reannealing protocol was performed by raising the temperature to 94°C for 30 s followed by a 28°C hold for 30 s.

Table 2.
Thermal cycling and premelting protocols

Following PCR, high-resolution melting was performed in a 96-well plate LightScanner (Idaho Technology) collecting data from 55–97°C at a ramp rate of 0.10°C per second. Calibration algorithm and genotype prediction.

Melting profiles were calibrated by first using a smoothing spline to approximate fluorescence data. Next, the Tm values were computed for each internal oligonucleotide calibrator present using the interpolated apex of the derivative spline data. Shift and linear scale factors were applied to align each calibrator. This shifting process aligns the amplicon melting profiles and minimizes variation. Data were re-sampled with a cubic spline fit and analyzed with commercial LightScanner software. The temperature regions surrounding the amplicon Tm were analyzed before and after prior alignment of the calibrator (+/– calibration) and displayed as derivative peaks. The probability that the observed Tm separation of alternative homozygotes was obtained by chance was assessed using the nonparametric Mann–Whitney U-test in Matlab® (The Mathworks, Inc., Natick, MA, USA).

Homozygous genotypes were automatically predicted by software before and after calibration based on the Tm values and standard deviations within each plate. We assumed that Tm values were normally distributed, that the estimated means were representative of the true population means, and that the variation was equal and representative for both populations. Homozygotes were classified by determining the distance from each sample to the mean Tm value of each homozygote population within the plate. Since the genotypes of all samples were pre-determined by probe-based methods and/or sequencing, we could determine whether each sample's Tm was closer to the correct or incorrect homozygous group. If the sample's Tm was closer to the mean Tm of the correct group, a successful classification occurred.

RESULTS

We designed and developed small-amplicon melting assays to genotype three G/C and three T/A variants in at least 47 unique human DNA samples and compared the results to probe-based assays and Sanger sequencing. The melting curves for all six base-pair neutral variants are shown in Figures 1–3. Figure 1 displays a complete derivative melting profile of the 42-bp CPS1 c.3405-29A>T small amplicon before (A) and after (B) calibration. Melting signatures from the low and high oligonucleotide calibrators are observed at ~62°C and 92°C flanking the small amplicon melting peaks. The insets show magnified views of the homozygote peaks that are only well separated after calibration. Figure 2 shows similar small amplicon melting data for the OTC, MSH2, PAH and BX647987 variants, while CHK2 variant genotyping is shown in Figure 3. With the exception of the BX647987 variant, calibration resolves the genotypes of the homozygous alleles. Five of the small amplicons, including BX647987, had two melting peaks for the heterozygous genotypes. Only the heterozygous 68-bp amplicon of OTC exhibited a single peak.

Figure 1.
Derivative melting curves showing internal oligonucleotide calibrators and amplicon melting peaks for a target in CPS1. Ninety-four profiles, each generated from independent PCR reactions of 47 human genomic DNAs performed in duplicate, are shown before ...
Figure 2.
Calibration improves small amplicon genotyping of base-pair neutral changes. The apexes of the derivative melting peaks are shown, before (left panels) and after calibration (right panels), for the OTC (A and B), MSH2 (C and D), PAH (E and F) and ...
Figure 3.
Effects of amplicon size on genotyping the CHK2 gene. Derivative melting curves for the 42-bp amplicon (A and B) and the 86-bp amplicon (C and D) are presented before (left panels) and after (right panels) calibration. The peaks for the low and high ...

A summary of the Tm results for all variants is shown in Table 3. Predicted Tm values are listed for all homozygous alleles based upon nearest neighbor calculations (11,18). Tm values for heterozygotes are not included because they are, in general, easily genotyped without calibration. For our MSH2 (49 bp), PAH (59 bp), CHK2 (42 bp) and BX647987 (58 bp) small amplicons, nearest neighbor analysis predicted Tm differences of 0.19°C, 0.18°C, 0.31°C and 0.04°C, respectively, between amplicons containing alternative homozygotes. For our CPS1 and OTC PCR products, nearest neighbor analysis predicts no Tm difference between homozygous T/T and A/A amplicons because of nearest neighbor symmetry. However, for both gene targets, the observed mean Tm values are different between homozygote groups, even without calibration. Observed Tm values were 5°C to 7°C higher than theoretical Tm values, caused in part by the presence of LCGreen Plus dye that stabilizes double-stranded DNA hybrids. A nonparametric Mann–Whitney U-test of the uncalibrated and calibrated mean Tm values for the alternate homozygotes indicated significant differences (P < 0.001) for all six variants. The Tm variation was lower by 37–90% after calibration. This reduction in variation after calibration is reflected in the Tm ranges for the CPS1 and PAH small amplicons that exhibit overlapping homozygous peaks before calibration that separate into two groups post-calibration (Figures 1 and 2). The single variant where the homozygous alleles were not resolved by calibration was rs1869458 in the candidate gene BX647987. For this variant, small Tm differences prevented resolution of the G/G and C/C genotypes (Figure 2). Calibration using both high and low calibrators showed better genotype resolution than using only one calibrator (data not shown).

Table 3.
Tm values of small amplicon homozygotes before and after calibration

Excluding the unresolved variant of BX647987, genotyping success rates for homozygotes were predicted from the Tm separation and variation at each locus (see Calibration algorithm and genotype prediction section in Materials and methods section). We then determined the observed genotyping error rates by calculating the difference from the measured Tm of each sample to the mean Tm value of either homozygote population within each plate. Since the genotypes of all samples were pre-determined by probe-based methods and/or sequencing, we could determine whether each sample's Tm was closer to either the correct or incorrect homozygous group. If the sample's Tm was closer to the mean Tm of the correct group, genotyping was successful. Excluding BX647987, homozygous genotyping success ranged from 88.3–91.8% before calibration and improved to 99.2–100% after calibration (Table 4).

Table 4.
Homozygous genotyping accuracy before and after calibration

To explore the effects of amplicon size on the ability to genotype a base-pair neutral variant, we designed 42-, 50- and 86-bp amplicons that targeted the CHK2 variant rs9608698. As seen in Figure 3, the separation between the genotypes decreases as the size of the amplicon increases from 42 bp to 86 bp. Derivative melting curves for the 50-bp amplicon are not shown. The results for all three amplicons are summarized in Table 3. The observed mean Tm shifts between G/G and C/C homozygotes in the 42-bp, 50-bp and 86-bp amplicons were 0.6°C, 0.5°C and 0.3°C, respectively, both before and after calibration.

DISCUSSION

Even in an era of high-throughput sequencing, there is still a need for better genotyping and scanning methods. High-resolution melting allows for simple and inexpensive scanning of full exons for sequence variants (19,20). Small amplicon melting is a genotyping method that complements scanning (21) and is attractive for several reasons. First, the method is homogeneous (closed-tube). Second, Tm alterations from base changes are usually greater with smaller amplicons compared with larger amplicons. This difference is because length is inversely correlated with the magnitude of Tm shift resulting from the base-pair change. Third, the sequence being genotyped is limited to the single base or few bases between the 3′ termini of the forward and reverse primers, limiting the possibility of interference from another variant. Fourth, its greatest appeal may be in its ease and low cost of both reagents and development relative to probe-based genotyping methods (10). Small amplicons generally require minimal optimization to produce robust, clean products that have simple melting transitions.

Several factors can confound genotyping by small amplicon melting. One key factor is inter-sample temperature variation due to differences in the physical and chemical components of the system. Inter-sample variation appears dependent on the type of melting instrument and reaction vessels; microtiter plate-based systems displayed more variability than capillary-based systems (9). Differences in extraction kit chemistry can also influence the melting curves of small amplicons by changing the ionic strength (16).

We utilized the dual internal oligonucleotide Tm control approach reported by Liew et al. (15) and Seipp et al. (16) to minimize inter-sample Tm variation. The calibration system we describe herein was able to resolve the homozygotes in five of the six base-pair neutral variants examined that are the most challenging. In fact, the two variants with symmetric nearest neighbors in CPS1 and OTC were predicted by current thermodynamic models to have identical Tm values. Calibration reduced variation within each allele group thereby improving genotyping accuracy. The single variant that was not successfully genotyped by calibration was the C>G variant in BX647987 rs1869458. The failure to discriminate between the G/G versus C/C homozygotes in this 58-bp amplicon appears to be due to the minimal mean Tm differences of 0.09°C before and 0.02°C between these homozygotes before and after calibration, respectively. These small observed mean delta Tm are consistent with the predicted delta Tm of 0.04°C, and are too small to detect given the 0.079–0.092°C and 0.029–0.036°C standard deviations observed before and after calibration, respectively.

We investigated the impact of fragment length on the CHK2 assay. While the C/C and G/G homozygotes for CHK2 were resolved with 42-, 50- and 86-bp amplicons, the observed delta Tm between the homozygous alleles decreased (~0.6°C, 0.5°C and 0.3°C) as the amplicon size increased (Table 3). This reduced Tm separation is also seen in the derivative melting curves of the heterozygotes that show more distinct peaks in the 42-bp fragment compared to that of the 86-bp fragment (Figure 3).

Internal oligonucleotide calibrators decrease genotyping errors caused by multiple physical and chemistry factors (13–16). Several sporadic samples in our data set displayed Tm far away from other samples of the same genotype before calibration that subsequently grouped within their respective genotype after calibration (e.g. gray heterozygote in Figure 3A versus 3B).

Our calibrators did not adversely affect the eight amplicon melting assays described here with respect to amplification robustness and specificity. We also tested these calibrators with other assays, some in multiplex, and have not observed deleterious consequences (ref. 21 and unpublished data). In theory, internal oligonucleotide calibrators may adversely influence PCR of some targets by either reducing amplification efficiency and/or generating undesired products. Reduction of amplification efficiency may occur if, for example, one of the calibrators can hybridize within the target sequences. The generation of undesired products could occur in two scenarios. In the first case, loss of the 3′ blocks on the calibrators may allow the oligonucleotides to act as primers for undesirable amplification. In the second case, sufficient homology between a primer and one of the calibrator strands may allow primer-to-calibrator products to be made. We have tested calibrators lacking the 3′ blocks and found that high levels of these unblocked oligonucleotides generated additional extension products and melting peaks that obscured analyses of the desired calibrator and target amplicon peaks (data not shown). Decreasing calibrator concentrations or using different calibrators may allow success in cases of initial failure.

In conclusion, internal oligonucleotide calibrators are an attractive and effective way to improve genotyping by amplicon melting. The method is homogenous and simple. Accumulated evidence indicates that calibrators will counter inter-sample variation caused by a variety of physical and chemistry factors. Calibrated small amplicon genotyping provides, for the first time, a resolution high enough to accurately genotype base-pair neutral variants, even ones predicted to have identical Tm by current nearest neighbor models. These findings indicate that further investigation of current thermodynamic nearest neighbor models for predicting DNA duplex stabilities is warranted.

ACKNOWLEDGEMENTS

We are grateful to Kent Moyle and Mark Kessler for assistance with figures. Funding to pay the Open Access publication charges for this article was provided by Idaho Technology.

Conflict of interest statement. High resolution melting analysis is licensed from the University of Utah to Idaho Technology. CTW holds equity interest in Idaho Technology. All of the other authors are employees of Idaho Technology

REFERENCES

1. Lay MJ, Wittwer CT. Real-time fluorescence genotyping of factor V Leiden during rapid-cycle PCR. Clin. Chem. 1997;43:2262–2267. [PubMed]
2. Crockett AO, Wittwer CT. Fluorescein-labeled oligonucleotides for real-time pcr: using the inherent quenching of deoxyguanosine nucleotides. Anal. Biochem. 2001;290:89–97. [PubMed]
3. Zhou L, Myers AN, Vandersteen JG, Wang L, Wittwer CT. Closed-tube genotyping with unlabeled oligonucleotide probes and a saturating DNA dye. Clin. Chem. 2004;50:1328–1335. [PubMed]
4. Gundry CN, Vandersteen JG, Reed GH, Pryor RJ, Chen J, Wittwer CT. Amplicon melting analysis with labeled primers: a closed-tube method for differentiating homozygotes and heterozygotes. Clin. Chem. 2003;49:396–406. [PubMed]
5. Wittwer CT, Reed GH, Gundry CN, Vandersteen JG, Pryor RJ. High-resolution genotyping by amplicon melting analysis using LCGreen. Clin. Chem. 2003;49:853–860. [PubMed]
6. Montgomery J, Wittwer CT, Zhou L. Scanning the cystic fibrosis transmembrane conductance regulator gene using high-resolution DNA melting analysis. Clin. Chem. 2007;53:1897–1898. [PubMed]
7. Herrmann MG, Durtschi JD, Bromley LK, Wittwer CT, Voelkerding KV. Amplicon DNA melting analysis for mutation scanning and genotyping: cross-platform comparison of instruments and dyes. Clin. Chem. 2006;52:494–503. [PubMed]
8. Herrmann MG, Durtschi JD, Bromley LK, Wittwer CT, Voelkerding KV. Instrument comparison for heterozygote scanning of single and double heterozygotes: a correction and extension of Herrmann et al., Clin. Chem., 2006;52:494–503. Clin. Chem. 2007;53:150–152. [PubMed]
9. Herrmann MG, Durtschi JD, Wittwer CT, Voelkerding KV. Expanded instrument comparison of amplicon DNA melting analysis for mutation scanning and genotyping. Clin. Chem. 2007;53:1544–1548. [PubMed]
10. Liew M, Pryor R, Palais R, Meadows C, Erali M, Lyon E, Wittwer C. Genotyping of single-nucleotide polymorphisms by high-resolution melting of small amplicons. Clin. Chem. 2004;50:1156–1164. [PubMed]
11. SantaLucia J, Jr, Allawi HT, Seneviratne A. Improved nearest-neighbor parameters for predicting DNA duplex stability. Biochemistry. 1996;35:3555–3562. [PubMed]
12. Palais RA, Liew MA, Wittwer CT. Quantitative heteroduplex analysis for single nucleotide polymorphism genotyping. Anal. Biochem. 2005;346:167–175. [PubMed]
13. Dodge A, Turcatti G, Lawrence I, de Rooij NF, Verpoorte E. A microfluidic platform using molecular beacon-based temperature calibration for thermal dehybridization of surface-bound DNA. Anal. Chem. 2004;76:1778–1787. [PubMed]
14. Nellaker C, Wallgren U, Karlsson H. Molecular beacon-based temperature control and automated analyses for improved resolution of melting temperature analysis using SYBR I green chemistry. Clin. Chem. 2007;53:98–103. [PubMed]
15. Liew M, Seipp M, Durtschi J, Margraf RL, Dames S, Erali M, Voelkerding K, Wittwer C. Closed-tube SNP genotyping without labeled probes/a comparison between unlabeled probe and amplicon melting. Am. J. Clin. Pathol. 2007;127:341–348. [PubMed]
16. Seipp MT, Durtschi JD, Liew MA, Williams J, Damjanovich K, Pont-Kingdon G, Lyon E, Voelkerding KV, Wittwer CT. Unlabeled oligonucleotides as internal temperature controls for genotyping by amplicon melting. J. Mol. Diagn. 2007;9:284–289. [PMC free article] [PubMed]
17. Cradic KW, Wells JE, Allen L, Kruckeberg KE, Singh RJ, Grebe SK. Substitution of 3′-phosphate cap with a carbon-based blocker reduces the possibility of fluorescence resonance energy transfer probe failure in real-time PCR assays. Clin. Chem. 2004;50:1080–1082. [PubMed]
18. Bommarito S, Peyret N, SantaLucia J., Jr. Thermodynamic parameters for DNA sequences with dangling ends. Nucleic Acids Res. 2000;28:1929–1934. [PMC free article] [PubMed]
19. Dobrowolski SF, McKinney JT, San Filippo CA, Giak SK, Wilcken B, Longo N. Validation of dye-binding/high-resolution thermal denaturation for the identification of mutations in the SLC22A5 gene. Inherit. Metab. Dis. 2003;26:S188. [PubMed]
20. Dobrowolski SF, McKinney JT, Amat di San Filippo C, Giak Sim K, Wilcken B, Longo N. Validation of dye-binding/high-resolution thermal denaturation for the identification of mutations in the SLC22A5 gene. Hum. Mut. 2005;25:306–313. [PubMed]
21. Dobrowolski SF, Ellingson C, Coyne T, Grey J, Martin R, Naylor EW, Koch R, Levy HL. Mutations in the phenylalanine hydroxylase gene identified in 95 patients with phenylketonuria using novel systems of mutation scanning and specific genotyping based upon thermal melt profiles. Mol. Genet. Metab. 2007;91:218–222. [PubMed]

Articles from Nucleic Acids Research are provided here courtesy of Oxford University Press
PubReader format: click here to try

Formats:

Related citations in PubMed

See reviews...See all...

Cited by other articles in PMC

See all...

Links

  • Gene (nucleotide)
    Gene (nucleotide)
    Records in Gene identified from shared sequence links
  • MedGen
    MedGen
    Related information in MedGen
  • Nucleotide
    Nucleotide
    Published Nucleotide sequences
  • PubMed
    PubMed
    PubMed citations for these articles
  • SNP
    SNP
    PMC to SNP links

Recent Activity

Your browsing activity is empty.

Activity recording is turned off.

Turn recording back on

See more...