Analysis of Tear Glucose Concentration with Electrospray Ionization Mass Spectrometry

Abstract

We have developed a mass spectrometry-based method which allows one to accurately determine the glucose concentration of tear fluid. We used a 1 μL micro-capillary to collect tear fluid from the tear meniscus with minimal irritation of the eye. We analyzed the 1 μL volume of collected tear fluid with liquid-chromatography electrospray ionization mass spectrometry with the use of D-glucose-6,6-d2 as an internal standard. Repeated measurements and a recovery experiment on pooled, onion-induced tears showed that the analysis of the glucose in tears was precise (4% relative standard deviation) and provided 100% recovery. We found the tear glucose concentration of one fasting non-diabetic subject to be 13 to 51 μM while the onion-induced tear glucose concentration of a different non-diabetic subject to be 211 to 256 μM.

Introduction

In the United States, 9.6% of the population over 20 years of age and 20.9% of the population over 60 years of age have Diabetes Mellitus [1]. The Diabetes Control and Complications Trial has shown that close monitoring and tight control of blood glucose concentration is critical in preventing the complications of Diabetes [2]. Current self-monitoring methods rely on invasive finger sticks to directly measure blood glucose levels to provide the critical information required to achieve glycemic control. A number of non-invasive approaches to monitoring blood glucose concentrations are being pursued; however, none have been successfully developed to the point where they have gained widespread clinical acceptance [3]. There has been significant recent activity in exploring non-invasive monitoring by using tear fluid glucose as an indicator for blood glucose [46]. Such an approach will be effective only if the glucose concentrations in tear fluid are a reliable surrogate for blood glucose concentrations.

A survey of the literature over the last 70 years on tear glucose determinations indicates significant disagreement in measured tear glucose concentrations and on the relationship of tear fluid glucose to blood glucose concentrations. Daum and Hill [7] noted in their 1982 review of the human tear glucose literature that reported values of tear glucose in normal individuals range between 0 and 3600 μM with the median values between 110 and 280 μM. More recently, a study of six non-diabetic adults found glucose concentrations in tear fluid ranging between 128 and 166 μM with an average of 139 μM [8].

The differences in tear glucose concentrations between these reports are likely due to the use of different tear fluid collection methods. Van Haeringen and Glasius specifically addressed the dependence of tear fluid glucose on the method of collection [9]. They used a glucose dehydrogenase method to analyze chemically stimulated tears collected with a capillary and mechanically stimulated tear fluid collected by filter paper. Van Haeringen and Glasius found higher glucose concentrations in the tear fluid collected by the filter paper and concluded that the increase is due to the mechanical stimulation of the corneal and conjunctival epithelium [9]. Other studies reported similar findings in experiments on rabbits [10] and in subjects that were tested immediately after swimming [7]. Therefore, it is apparent that to determine a physiologically-relevant, baseline tear glucose concentration, tear fluid must be acquired with minimal tear stimulation and eye irritation.

Several previous studies used mass spectrometry to analyze tears for constituents other than glucose. Fung et al. [11] characterized the protein profile in the human tear fluid with MALDI-ToF and electrospray ionization mass spectrometry. Ham et al. [12] identified the lipid profile in tear fluid and reported on their variations between normal and dry eyes. There have been many reports of glucose measurement by mass spectrometry on bodily fluids other than tears. The majority of these studies use gas chromatography and an electron ionization source [1316]. With gas chromatography, glucose typically needs to be derivatized to make it more volatile. Other studies used liquid chromatography to bypass the volatility issue [17, 18]. Sodium [19], lithium [20, 21], cesium [17], and chloride [22] ions have been used with electrospray ionization to form the cationic adduct of glucose from aqueous solutions.

We discuss here a new, robust method of determining physiological tear glucose concentrations by using electrospray ionization mass spectrometry (ESI-MS). This method enables reliable studies of tear glucose concentrations in tear fluid. We use a tear fluid sampling method that is minimally irritating to the eye to collect 1 μL tear samples. We validate the ESI-MS method by using glucose standard solutions, artificial tear fluid, and large volumes of stimulated tears. We determine tear glucose concentration in a non-diabetic, fasting adult.

Experimental

All tear samples were collected and transferred with 1 μL “Microcaps” micro-capillaries (Drummond Scientific Company, Broomall, USA). The tear film meniscus on the bottom eyelid was visualized using a slit lamp ophthalmic microscope (American Optics, Burlington, Canada). The micro capillary was gently touched to this tear fluid meniscus and the tear fluid was drawn in by capillary action. During half the collections, there was inadequate tear fluid to immediately fill the capillary with one touch. In these cases, the subject was asked to blink in order to redistribute the tear fluid and the capillary was again touched to the meniscus. This process was repeated until the capillary was completely filled. A few samples required as many as five repetitions of this process. Care was taken throughout tear fluid collection to avoid stimulation of reflex tearing. The aqueous glucose standards were collected and transferred with the same process as a tear sample; the capillary was touched to the solution surface and was filled by capillary action. For studies involving larger volumes of stimulated tears, the subject chopped onions until tearing was induced. These tears were collected by capillary action with glass Pasteur pipettes (2 mL) and pooled for each subject in order to analyze a homogeneous sample.

Immediately after collection, the tear sample was transferred to a glass vial (Agilent, Palo Alto, USA) with a glass insert (Restek, Bellafonte, USA) containing 10 μL of acetonitrile (Mallinckrodt Baker, Inc., Phillipsburg, USA) and 10 μL of an aqueous stock solution of isotopically labeled glucose (D-glucose-6,6-d2, 99% labeled, Cambridge Isotope Laboratories, Andover, USA) used as an internal standard. A typical value for the concentration of the D-glucose-6,6-d2 stock solution was 27.5 μM, corrected for the 99% labeling (leading to a D-glucose-6,6-d2 concentration of 13.1 μM in the sample injected into the mass spectrometer). The vials were then shaken and centrifuged for five minutes to ensure mixing and that the liquid was collected at the bottom of the vial. The samples were refrigerated at 5°C until analyzed.

Glucose concentrations were measured using ESI-MS. Samples were analyzed by flow injection analysis on a HP 1000 LC-MS (Agilent, Palo Alto, USA). Initial studies used an autosampler connected directly to the mass spectrometer. However, our optimized method used a 2 μm filter and a Nova-Pak C18 3.9 × 150 mm column (Waters, Milford, USA) placed in-between the autosampler and the mass spectrometer. In both methods, the mobile phase consisted of acetonitrile and water with 0.1% formic acid (Mallinckrodt Baker, Inc., Phillipsburg, USA) with a constant flow rate of 0.2 mL/min. The ratio of water to acetonitrile was 1:1 in the absence of the column and 17:3 with the column. The ion chromatograms of sodiated glucose isotopes (m/z 203 for natural glucose and m/z 205 for D-glucose-6,6-d2) were selected from the total ion chromatogram using the ChemStation software (Rev. A.06.03 [509]). These selected ion chromatograms were integrated to provide peak areas for the two glucose signals. During integration, the software option “baseline hold” was used. The peak areas of m/z 203 and 205 were used to determine the glucose-to-D-glucose-6,6-d2 ratio. When calculating the peak area ratio, we took into account the 1.3% m+2 natural isotopic contribution of Na+(glucose), m/z 203, to m/z of 205; 1.3% of the observed glucose peak area of m/z 203 was subtracted from the observed m/z 205 peak area to determine the area due solely to the internal standard. Each sample was injected three times at 6.5 μL per injection and the average glucose concentration recorded. All human samples were collected through protocols approved by the University of Pittsburgh Institutional Review Board.

Results and Discussion

The cationization of glucose with H+, Na+, K+, or Pb2+ was examined. An initial survey comparing negative ion to positive ion ESI suggested greater promise with positive ion methods. With the exception of sodium, millimolar concentrations of the acid or salt were needed to observe the glucose cation adducts as the base peak. The sodium-glucose adduct was present in all of the spectra and in the majority of them was the most abundant ion. Since sodiated glucose dominated even without the addition of sodium and since tear fluid contains ~160 mM of sodium [23], our glucose analysis is based on the sodiated glucose ion. Further experiments were conducted to ascertain the solvent composition (water, acetonitrile, and methanol) that provided the maximum signal for sodiated glucose. Based on the results of these experiments, a 1:1 ratio of water to acetonitrile for the sample solvent was selected. For the experiments without the use of the Nova-Pak C18 column, the 1:1 solvent ratio was also used for the mobile phase with the addition of 0.1% formic acid to the aqueous phase to produce more stable baseline ion currents. In our initial work, 13C6-glucose was used as an internal standard but an unidentified peak at m/z 209 (which would interfere with the detection of sodiated 13C6-glucose) was observed in the absence of the internal standard. Therefore, D-glucose-6,6-d2 was a superior alternative since no signal was observed at m/z 205 in the absence of the internal standard.

With the optimal solvents and internal standard established, the effect of the biological matrix on the glucose signal was examined. The sodiated-glucose signal intensity observed for both artificial and real tear fluid was low compared to the background signal in our initial studies in which samples were directly infused into the mass spectrometer. Addition of a C18 column separated proteins, such as lyzosyme, globulin, and albumin, and excess salts from the elution time of glucose. This in-line purification increased both the sensitivity and signal-to-noise ratio for sodiated-glucose. Analysis of the blank (a protein-free glucose-free artificial tear solution [4]) showed an average glucose concentration of 0.98 ± 0.56 μM. With the use of the column the statistical limit of detection was lowered to 3 μM. These improvements occurred in the presence of the column with the mobile phase optimized at a solvent ratio of 17:3 of water with 0.1% formic acid to acetonitrile which maximized the glucose signal and minimized overlapping peaks.

The precision of the micro-capillaries was measured by gravimetric means. Five individual capillaries were weighed, filled with water at 20°C, and weighed again. The average value for the water contained in the capillaries was 1.008 ± 0.013 mg (1.011 ± 0.013 μL at 20°C). Therefore, in all of our analysis of tears we assume a collection volume of 1.00 μL.

To test the linearity of the method, aqueous solutions with known glucose concentrations (0 – 200 μM) were prepared and analyzed. The linear regression of measured versus added glucose gave an equation of y = 0.99 x − 0.78 (R2 = 0.98). The increase in the relative sodiated glucose signal m/z 203 can be easily seen when the mass spectra are normalized to the sodiated D-glucose-6,6-d2 signal m/z 205 (Figure 1).

Figure 1
Eleven spectra of aqueous solutions of known glucose concentrations (0 – 200 μM) with intensities normalized to the internal standard D-glucose-6,6-d2 signal intensity at m/z 205. The linear regression of the experimental glucose concentration ...

The mass spectral measurement over time of the glucose concentration in the 1:1 ratio of water-to-acetonitrile was examined. A 200 μM glucose aqueous solution was prepared and sampled multiple times. These vials were stored in a refrigerator and tested immediately, a few hours later, three days later, and one week later. The average value for all of the samples was 201 ± 15 μM with no increasing or decreasing trend over storage time (data not shown).

The reproducibility of the method was determined with the use of a pooled sample of onion-induced tears. Six samples were collected from the pool and analyzed. The average glucose concentration was 211 ± 8 μM in this stimulated tear sample (data not shown).

A recovery experiment was performed on another pooled sample of onion-induced tears. The tears were spiked with aqueous glucose solutions to increase the glucose concentration by 0, 107, 233, and 352 μM. Linear regression of measured versus added glucose gave an equation of y = 0.99 x + 256 (R2 = 0.99) (Figure 2). We also determined that the induced tear fluid contained 256 μM glucose.

Figure 2
Duplicate samples of pooled, induced tears at four different concentrations of added glucose were analyzed. Error bars are plus/minus one standard deviation from the average value of each sample. Linear regression analysis gave an equation of y = 0.99 ...

The average relative standard deviation over the three 6.5 μL injections of the diluted tear samples studied with concentrations greater than 50 μM glucose was 4%. For the samples studied with concentrations less than 50 μM glucose, the average relative standard deviation of the three measurements was 15%. The FDA currently requires self-blood-glucose monitors to have an accuracy of within 15%. Within one standard deviation, the calibration curve is indistinguishable from y = x and the recovery experiment shows 100% recovery of added glucose. We therefore feel confident using D-glucose-6,6-d2 as the internal standard and the y = x calibration curve to determine tear glucose concentrations.

We performed two studies on a non-diabetic subject who fasted overnight. During both sessions, which were separated by one week, three samples were taken from each eye. Samples from the same eye were collected at ten minutes intervals. The tear glucose concentration from this subject during the two sessions ranged from 13 to 51 μM with an average of 32 μM (Table 1). The concentration of tear glucose determined in this work for this non-diabetic subject is about four-fold less than the tear glucose concentrations recently reported [8]. Whereas previous studies used collection methods which either touched the eye or chemically stimulated tears [7, 9, 10], we use glass micro-capillaries to collect a single microliter volume of tear fluid from the tear meniscus without touching the eye. This decrease in reported tear glucose concentration is likely due to our less irritating method of tear collection. The onion-stimulated tears (from a different subject) showed a seven to eight fold higher glucose concentration than in the absence of stimulation.

Table 1
Tear Glucose Concentration of a Non-diabetic Subject

Conclusion

In the present study we demonstrate a robust ESI-MS method for determining tear glucose concentration in tear fluid. Liquid chromatography separates glucose from the majority of salts and proteins, such as lyzosyme, globulin, and albumin, in tears reducing interference of the biological matrix on the sodiated-glucose signal. The resulting increase in the sensitivity and signal-to-noise ratio of glucose over data obtained without the column allows measurements of tear glucose concentrations as low as 10 μM (limit of quantification) by relying on an internal standard. We use a 1 μL tear sampling method that is minimally irritating to the eye. This method enables reliable studies of physiological tear glucose concentrations in non-stimulated tears. A preliminary study indicates that the physiological, baseline tear glucose concentration is substantially below previous reports. Further studies are warranted to determine the inter- and intra-individual variation in baseline tear glucose concentration. This method will be used to establish normal values for tear glucose concentration and can monitor changes in tear glucose with changes in blood glucose. Application of this LC-MS method can help us to understand tear glucose physiology. Our future work will determine whether monitoring tear glucose concentration is a viable approach for non-invasive blood glucose monitoring. We are now using this method for tear glucose concentration measurements to examine a population of non-diabetic and diabetic subjects to determine their tear fluid glucose concentrations and their correlations of tear glucose with blood glucose.

Acknowledgments

The authors thank Kasi Somayajula for his consultation. This research was supported of the National Institutes of Health Grant DK-55348 to Sanford A. Asher.

Footnotes

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

References

1. National Diabetes Statistics. NIH Publication No. 06-3892. U.S. Department of Health and Human Services, National of Institutes of Health, National Institute of Diabetes and Digestive and Kidney Diseases, National Diabetes Information Clearinghouse. 2005.
2. Shamoon HDH, Fleischer N, Engel S, Saenger P, Strelzyn M, Litwak M, Wylierosett JFA, et al. The Effect of Intensive Treatment of Diabetes on the Development and Progression of Long-Term Complications in Insulin-Dependent Diabetes-Mellitus. New England Journal of Medicine. 1993;329:977–986. [PubMed]
3. Khalil OS. Non-Invasive Glucose Measurement Technologies: An Update from 1999 to the Dawn of the New Millennium. Diabetes Technology & Therapeutics. 2004;6:660–697. [PubMed]
4. Alexeev VL, Das S, Finegold DN, Asher SA. Photonic Crystal Glucose-Sensing Material for Noninvasive Monitoring of Glucose in Tear Fluid. Clinical Chemistry. 2004;50:2353–2360. [PubMed]
5. Badugu R, Lakowicz JR, Geddes CD. Ophthalmic Glucose Sensing: a Novel Monosaccharide Sensing Disposable and Colorless Contact Lens. Analyst (Cambridge, United Kingdom) 2004;129:516–521. [PubMed]
6. March WF, Mueller A, Herbrechtsmeier P. Clinical Trial of a Noninvasive Contact Lens Glucose Sensor. Diabetes Technology & Therapeutics. 2004;6:782–789. [PubMed]
7. Daum KM, Hill RM. Human Tear Glucose. Investigative Ophthalmology & Visual Science. 1982;22:509–514. [PubMed]
8. Chen R, Jin Z, Colon LA. Analysis of Tear Fluid by CE/LIF: a Noninvasive Approach for Glucose Monitoring. Journal of Capillary Electrophoresis. 1996;3:243–248. [PubMed]
9. Van Haeringen NJ, Glasius E. Collection Method-Dependent Concentrations of Some Metabolites in Human Tear Fluid, with Special Reference to Glucose in Hyperglycemic Conditions. Albrecht von Graefes Archiv fuer Klinische und Experimentelle Ophthalmologie. 1977;202:1–7. [PubMed]
10. Kilp H, Heisig B. Changes of the Glucose and Lactate Concentration in Tears after Mechanical Alteration and Contact Lens-Wearing. Albrecht von Graefes Archiv fuer Klinische und Experimentelle Ophthalmologie. 1975;193:259–267. [PubMed]
11. Fung KYC, Morris C, Sathe S, Sack R, Duncan MW. Characterization of the In vivo Forms of Lacrimal-Specific Proline-Rich Proteins in Human Tear Fluid. Proteomics. 2004;4:3953–3959. [PubMed]
12. Ham BM, Jacob JT, Keese MM, Cole RB. Identification, Quantification and Comparison of Major Non-polar Lipids in Normal and Dry Eye Tear Lipidomes by Electrospray Tandem Mass Spectrometry. Journal of Mass Spectrometry. 2004;39:1321–1336. [PMC free article] [PubMed]
13. Clapperton AT, Coward WA, Bluck LJC. Measurement of Insulin Sensitivity Indices Using 13C-Glucose and Gas Chromatography/Combustion/Isotope Ratio Mass Spectrometry. Rapid Communications in Mass Spectrometry. 2002;16:2009–2014. [PubMed]
14. Dube G, Henrion A, Ohlendorf R, Vidal C. Application of the Combination of Isotope Ratio Monitoring with Isotope Dilution Mass Spectrometry to the Determination of Glucose in Serum. Rapid Communications in Mass Spectrometry. 2001;15:1322–1326. [PubMed]
15. Lee J, Chung BC. Simultaneous Measurement of Urinary Polyols Using Gas Chromatography/Mass Spectrometry. Journal of Chromatography, B: Analytical Technologies in the Biomedical and Life Sciences. 2006;831:126–131. [PubMed]
16. Di Gioia ML, Leggio A, Le Pera A, Liguori A, Napoli A, Siciliano C, Sindona G. Quantitative Analysis of Human Salivary Glucose by Gas Chromatography-Mass Spectrometry. Journal of Chromatography, B: Analytical Technologies in the Biomedical and Life Sciences. 2004;801:355–358. [PubMed]
17. Rogatsky E, Jayatillake H, Goswami G, Tomuta V, Stein D. Sensitive LC MS Quantitative Analysis of Carbohydrates by Cs+ Attachment. Journal of the American Society for Mass Spectrometry. 2005;16:1805–1811. [PubMed]
18. Wan ECH, Yu JZ. Determination of Sugar Compounds in Atmospheric Aerosols by Liquid Chromatography Combined with Positive Electrospray Ionization Mass Spectrometry. Journal of Chromatography, A. 2006;1107:175–181. [PubMed]
19. McIntosh TS, Davis HM, Matthews DE. A Liquid Chromatography-Mass Spectrometry Method to Measure Stable Isotopic Tracer Enrichments of Glycerol and Glucose in Human Serum. Analytical Biochemistry. 2002;300:163–169. [PubMed]
20. Bruggink C, Maurer R, Herrmann H, Cavalli S, Hoefler F. Analysis of Carbohydrates by Anion Exchange Chromatography and Mass Spectrometry. Journal of Chromatography, A. 2005;1085:104–109. [PubMed]
21. Guignard C, Jouve L, Bogeat-Triboulot MB, Dreyer E, Hausman JF, Hoffmann L. Analysis of Carbohydrates in Plants by High-Performance Anion-Exchange Chromatography Coupled with Electrospray Mass Spectrometry. Journal of Chromatography, A. 2005;1085:137–142. [PubMed]
22. Zhu J, Cole RB. Ranking of Gas-Phase Acidities and Chloride Affinities of Monosaccharides and Linkage Specificity in Collision-Induced Decompositions of Negative Ion Electrospray-Generated Chloride Adducts of Oligosaccharides. Journal of the American Society for Mass Spectrometry. 2001;12:1193–1204. [PubMed]
23. Gilbard JP. Human Tear Film Electrolyte Concentrations in Health and Dry-Eye Disease. International Ophthalmology Clinics. 1994;34:27–36. [PubMed]
PubReader format: click here to try

Formats:

Related citations in PubMed

See reviews...See all...

Cited by other articles in PMC

See all...

Links

  • Compound
    Compound
    PubChem Compound links
  • MedGen
    MedGen
    Related information in MedGen
  • PubMed
    PubMed
    PubMed citations for these articles
  • Substance
    Substance
    PubChem Substance links

Recent Activity

Your browsing activity is empty.

Activity recording is turned off.

Turn recording back on

See more...