Fig. 1. From: Radiographic Classification of Complications of Instrumentation in Adolescent Idiopathic Scoliosis.

The anteroposterior radiograph demonstrates an example of a serious radiographic adverse event based on our classification scheme. The white arrow indicates a broken rod at the T10–11 disc in a patient who was treated for adolescent idiopathic scoliosis using open anterior instrumented spinal fusion. This was defined as a serious radiographic adverse event, as it was a failure of an implant indicating a likely pseudarthrosis.