U.S. flag

An official website of the United States government

NCBI Bookshelf. A service of the National Library of Medicine, National Institutes of Health.

Berliner LJ, Parinandi NL, editors. Measuring Oxidants and Oxidative Stress in Biological Systems [Internet]. Cham (CH): Springer; 2020. doi: 10.1007/978-3-030-47318-1_10

Cover of Measuring Oxidants and Oxidative Stress in Biological Systems

Measuring Oxidants and Oxidative Stress in Biological Systems [Internet].

Show details

Chapter 10Resonators for Clinical Electron Paramagnetic Resonance (EPR)

, , and .

Author Information and Affiliations

Published online: August 9, 2020.

In pulsed electron paramagnetic resonance (EPR), free-induction decay (FID) or spin echo (SE) signals of unpaired electrons are recorded in the time-domain. In both methods, electromagnetic waves play an important role in the detection of unpaired electrons in EPR spectroscopy. The resonator generates and senses electromagnetic waves and therefore serves as a critical interface between unpaired electrons and the transmit/receive systems of an EPR spectrometer. Since a resonator is a sensitive electrical circuit that can amplify voltages and currents when the electrical circuit of the resonator is on resonance, the resonator is an essential component for EPR detection in continuous wave and pulsed EPR. Without the resonator, EPR signals cannot be detected with sufficient sensitivity. In this chapter, the basics of resonators and some examples of resonators used in preclinical studies with small animals and human subjects are explained.

Keywords:

Resonators for clinical EPR, Clinical electron paramagnetic resonance

Introduction

For EPR detection , a radiofrequency (RF) or microwave resonator is commonly used to apply electromagnetic waves to unpaired electrons in a sample. The frequency of electromagnetic waves is determined by the static magnetic field B0 applied to the sample. In continuous-wave (CW) electron paramagnetic resonance (EPR) , the energy absorption in the resonator is measured as a function of the magnetic field B0. In pulsed EPR, free-induction decay (FID) or spin echo (SE) signals of unpaired electrons are recorded in the time-domain. In both methods, electromagnetic waves play an important role in the detection of unpaired electrons in EPR spectroscopy. The resonator both generates and senses electromagnetic waves and therefore serves as a critical interface between unpaired electrons and the transmit/receive systems of an EPR spectrometer.

In CW-EPR , the energy absorption of unpaired electrons due to the EPR phenomenon is detected through the reflection of incident electromagnetic waves at the resonator. When the energy absorption of unpaired electrons occurs at the resonator, the input impedance of the resonator changes; this results in an impedance mismatch between the resonator and the transmission line, which is usually 50 Ω. To detect EPR absorption, the resonator should be sensitive to the energy absorption due to electron spins by having sufficient quality and filling factors, and by permitting sufficient microwave power to be transmitted to the measured sample [1]. Since a resonator is a sensitive electrical circuit that can amplify voltages and currents when the electrical circuit of the resonator is on resonance, the resonator is an essential component for EPR detection in CW and pulsed EPR . Without the resonator, EPR signals cannot be detected with sufficient sensitivity. This is similar to the detection of nuclear magnetic resonance (NMR).

The requirements of the resonator differ between CW-EPR and pulsed EPR because the methods of detection involve different experimental setups, which measure different aspects of the resonance characteristics at different points in time. Therefore, the resonator is designed to meet specific needs in applications and measurements. In this chapter, the basics of resonators and some examples of resonators used in preclinical studies with small animals and human subjects are explained.

Basics of Resonators for EPR

Resonant Circuit

An RF resonator is an electrical circuit that stores electrical and magnetic energy. The simplest resonant circuit is a series LC circuit and a parallel LC circuit. Figure 10.1 illustrates the RLC series resonant circuit, which is driven by a sinusoidal voltage source V (t). In this circuit, a resistor R satisfies Ohm’s law, a capacitor C stores electric charges, and an inductor L stores magnetic fields.

Fig. 10.1

Fig. 10.1

An RLC series resonant circuit with a sinusoidal voltage source V (t)

Let us consider a sinusoidal signal of frequency f. The impedance Zaa of the RLC circuit at port aa′ is given by

Zaa=R+jωL1ωC,
10.1
where w is the angular frequency, ω = 2 ρf, and j is the imaginary unit, j=1. If you are not familiar with sinusoidal steady-state analysis, textbooks of basic circuit theory will be helpful. By Ohm’s law, current i(t) is expressed as
it=VtZaa=VtR+jωL1ωC.
10.2

When the imaginary part in Eq. (10.1) vanishes, the impedance becomes minimal and consists of only the real part R. At this point, current i(t) is expressed as

it=VtR.
10.3

In Eq. (10.2), current i(t) is frequency-dependent, and is maximal at angular frequency

ω=1LC.
10.4

From Eq. (10.4), the resonant frequency fr is given by

fr=12πLC.
10.5

At the resonant frequency, current i flowing in the circuit is maximized, and magnetic flux in the inductor also becomes maximized. Therefore, RF magnetic fields are efficiently generated at the resonant frequency fr.

Sensitivity of EPR Detection

The sensitivity of EPR signal detection is well documented [2]. In a reflection-type EPR bridge, as illustrated in Fig. 10.2, the RF resonator is connected to the transmission line to apply the electromagnetic waves to a sample. Figure 10.3 illustrates a resonator connected to the transmission line.

Fig. 10.2

Fig. 10.2

General schematic of a simplified reflection-type bridge for CW-EPR

Fig. 10.3

Fig. 10.3

Transmission line and a resonator

The signal intensity VS of an EPR absorption spectrum in CW-EPR is given by

VS=χηQPZ0,
10.6
where η is the filling factor, Q is the quality factor of the resonator loaded with a sample, P is the incident RF power to the resonator, and Z0 is the characteristic impedance of the transmission line connected to the resonator [1]. In Eq. (10.6), χ is the imaginary component of the effective RF susceptibility and depends on the sample. From the viewpoint of the electrical circuit, the product ηQPZ0 should be maximized to obtain the maximum signal intensity. The incident RF power P has a limitation in EPR spectroscopy that is sample-specific, because saturation effects of EPR signals are observed when a spin system being measured is saturated with a higher RF magnetic field. The RF power P also depends on the sample and the conversion efficiency of RF magnetic fields in the resonator. The characteristic impedance Z0 depends on the selection of the transmission lines of the EPR spectrometer. The most commonly used characteristic impedance in RF and microwave commercial products is 50 Ω. Thus, the filling factor η and the quality factor Q are key factors for optimizing the sensitivity of the resonator in EPR spectroscopy. Further descriptions of the filling factor and quality factor are given below.

Filling Factor and Quality Factor

The filling factor is generally defined as the ratio between the stored magnetic energy in the resonator and the stored magnetic energy in the sample that can contribute to EPR phenomena. Figure 10.4 illustrates the resonator and the sample, as well as the magnetic fields.

Fig. 10.4

Fig. 10.4

Geometry of magnetic fields , a resonator, and a sample regarding the filling factor

The RF magnetic field B1⊥ perpendicular to the static magnetic field B0 is taken into account in EPR phenomenon. The density of magnetic energy wm in the space can be calculated as

wm=12HB=μ2HB=12μBB,
10.7
where H is a magnetic field in A/m, B is a magnetic flux density (μH) in T, and μ is the magnetic permeability of the space and is written as μ0 = 4π × 10−7 (N/A2) in free space. When the magnetic field distribution of RF magnetic field B1 in the resonator is known, the stored magnetic energies in the sample and the resonator can be calculated as
η=sampleB12dVresonatorB12dV.
10.8

The quality factor is a dimensionless value that quantifies a resonator’s ability to store and dissipate energy, and can be evaluated as

Q=2π×energystoredenergydissipatedperonecycle.
10.9

However, direct measurements of the stored energy in the resonator and the energy dissipated per cycle are not practical. Another equation for the quality factor is often used:

Q=frΔf=frf1f2,
10.10
where δf is the 3-dB bandwidth of the resonator that is the difference between the frequencies, f1 and f2, in which the stored energy becomes half of the maximum stored energy in the resonator. Presently, frequencies can be measured very precisely through the use of a vector network analyzer. Equation (10.9) is a better definition for understanding the physical meaning of the quality factor. However, Eq. (10.10) is commonly used in practice by measuring the RF resonance characteristics of the resonator.

For CW-EPR , a high quality factor and a high filling factor are both required to ensure sufficient EPR signal intensity according to Eq. (10.6). A high filling factor with a biological sample, which has a loss in terms of electromagnetic waves, results in an increase in energy dissipation in the resonator, which results in a low quality factor. In contrast, when a small part of a biological sample is placed in the resonator, the losses due to the resonator decrease, which results in a low filling factor and a high quality factor. In biological EPR spectroscopy and imaging, the balance between the quality factor and the filling factor should be optimized so that the biological sample can be measured adequately. Therefore, the resonator should be designed for a specific application to obtain the maximal EPR signal intensity with the optimal conditions for the quality factor and the filling factor. For pulsed EPR, the quality factor is significantly lowered to reduce the ring-down time after the RF pulses and to have a broader bandwidth. This aspect of pulsed EPR is explained in another chapter.

Frequency Selection of Preclinical and Clinical EPR

The RF frequency η for EPR spectroscopy and imaging is determined by the static magnetic field B0 applied to a subject, according to the fundamental equation for the EPR phenomenon, hv = BB0, where h is Plank constant (6.626 × 10−34 Js), g is the g-factor of the species to be measured (2.002 for the free electron), and βB is Bohr magneton (9.274 × 10−24 JT1) [3]. In laboratories, an X-band EPR spectrometer is commonly used as an analytical tool for various samples. However, the penetration of microwaves into biological tissues is critical in biomedical applications for small animals and even human subjects. As a measure of the penetration of electromagnetic waves in a subject, the skin depth is considered. The skin depth defines the distance at which the amplitude of electromagnetic waves becomes 1/e = 0.3679, where e is Napier’s constant. At angular frequency w, the skin depth d can be calculated as

d=εμω221+σεω211/2,
10.11
where Ɛ is the permittivity (dielectric constant) of the medium, ε is the conductivity of the medium (biological tissues), and μ is the permeability of the medium [4]. As you can see, the skin depth depends on the angular frequency. The dielectric constant and the conductivity are also frequency-dependent. Only permeability μ in biological tissues is usually considered to be the permeability in free space.

In vivo EPR measurements with mice were initially conducted in the X-band (ν = 8.5 GHz, B0 = 0.3 T) [5], but soon thereafter the frequency of electromagnetic waves was reduced to a lower frequency to allow increased penetration of electromagnetic waves in biological tissues [6]. The EPR studies below the X-band are well documented by Eaton and Eaton [7]. From the 1990s to the present, most in vivo small animal studies have been performed at frequencies from 250 to 1200 MHz (VHF/UHF to L-band). EPR imaging on human skin has been conducted in the S-band (2 GHz) [8, 9], since it does not require a significant penetration of electromagnetic waves in tissues and provides better sensitivity.

Frequency selection is important for biomedical applications of EPR spectroscopy and imaging, since it may limit the detection volume and the overall sensitivity of the EPR measurements. Several factors that affect EPR measurements should be taken into account. The criteria for frequency selection are as follows:

(a)

The penetration depth of electromagnetic waves

A higher frequency may limit the penetration depth (see Eq. (10.11)).

(b)

The sensitivity of EPR signals

A higher magnetic field and a corresponding frequency give stronger EPR signals. If there is no problem regarding the penetration depth, a frequency as high as possible is beneficial.

(c)

Size of the species to be measured

How large is the subject? A mouse, rabbit, pig, or even a human?

(d)

What will be measured, and where will it be measured?

The region of interest, such as superficial tissues or a deep region, determines the required penetration depth.

(e)

Specific absorption rate (SAR) and safety regulations

To avoid unwanted effects in a subject, energy absorption in biological tissues should be below certain levels, which are promulgated in safety regulations by national governments or international organizations.

Types of Resonators

The resonators used for biological EPR applications can be divided into two major categories: volume resonators and surface resonators (similar to clinical magnetic resonance imaging (MRI)).

Samples are typically placed inside a volume resonator, which can accommodate a whole body or part of a subject animal. These types of resonators, such as a loop-gap resonator, benefit from a higher filling factor [10]. Since the sample is placed in the volume of the resonator where the magnetic flux generated by the resonator is the highest, the spins in the sample are stimulated more efficiently. The wavelength of electromagnetic waves may limit the dimensions of the resonator. When a large sample space is needed and the sample is not small in comparison to the wavelength, the resonator structure should be considered to obtain homogeneous RF magnetic fields in the sample space.

Surface resonators are typically placed on the surface of a sample. A surface resonator can measure EPR signals from localized regions of a subject animal, such as the skin or tissues close to the external surface of the animal [11]. As mentioned above, the region of interest is a key factor when choosing the type of resonator. In addition to these major categories, implantable resonators have been developed for small animal experiments, which address the problem of the penetration depth. Surgical procedures may be required for the use of an implantable resonator in animals. Specific examples of EPR resonators are introduced in section “Brief History of Technical Developments for In Vivo EPR Resonators ”.

Other Technical Aspects for RF Resonators in EPR

Previous sections addressed several technical aspects (filling factor, quality factor, and penetration depth). This section briefly explains other important aspects of the resonators used in EPR spectroscopy and imaging.

Conversion Efficiency of RF Magnetic Fields

The conversion efficiency of an RF magnetic field is the ratio of RF magnetic flux density B1 and the square root of the incident RF power to the resonator. This conversion efficiency directly affects the sensitivity of the resonator because the density of magnetic energy in the sample is proportional to the square of the RF magnetic field B1 generated by the resonator. When the input RF power is constant, a resonator with a high conversion efficiency will provide a stronger B1 field than a resonator with a lower conversion efficiency. As a result, the signal intensity is proportional to the square of the conversion efficiency [12]. A resonator with a high conversion efficiency is desirable for all EPR experiments .

Frequency Tuning Adjustment

The microwave frequency of the RF source can be controlled and locked to the resonant frequency of the resonator, which is commonly referred to as automatic frequency control (AFC). With AFC, there is no need to adjust the frequency of the resonator. However, experiments that measure the narrow line-width of a first-derivative EPR absorption spectrum (such as EPR oximetry) require a very stable source frequency that is not always attainable. Another approach to maintaining equal frequency between the resonator and the source is called automatic tuning control (ATC). With ATC, the resonator is designed to accommodate its own frequency adjustment. There are several approaches to adjust the resonant frequency, including the use of an (1) induction motor [13], (2) piezoelectric actuator [14], and (3) varactor diodes [15, 16].

Impedance Matching (Coupling) Adjustment

For the bridge to adequately transfer the reflections of electromagnetic waves due to EPR from the resonator to the detector, the impedance of the loaded resonator should be adjusted to the characteristic impedance of the bridge (usually 50 Ω). Inductive coupling or capacitive coupling networks have been used for this purpose [17]. Traditionally, in X-band cavity resonators, the iris of a waveguide was adjusted by hand. In modern X-band cavity resonators, motorized iris control is available. When samples are measured in vivo where motion of the subject is expected, the resonator should automatically tune and couple to the perturbation due to motion of the subject to ensure high-quality recorded spectra. As with frequency adjustment, there are several approaches to adjusting the impedance of the resonator: (1) induction motor [13], (2) piezoelectric actuator [18], (3) varactor diodes [15], and (4) photo-resistor [16]. This technique is commonly referred to as automatic coupling control (ACC) or automatic matching control (AMC).

Homogeneity of RF Magnetic Fields

As mentioned previously, the EPR signal intensity depends on the square of RF magnetic field B1. To achieve a uniform sensitivity of EPR detection within the resonator, a homogeneous B1 field needs to be established. Three factors alter the RF magnetic field:

(a)

The distance from the currents. In the radial direction, the RF magnetic field is weakest at the center of the resonator. However, as the distance from the current source to the observation point decreases, the RF magnetic field increases.

(b)

In many cases, the physical length of the resonator is not negligibly small in comparison to the wavelength of electromagnetic waves. In this case, the RF current flowing in the conductors is no longer uniform.

(c)

Also, the electrical properties of biological tissues in animals are not the same. These properties affect the RF electromagnetic fields.

Figure 10.5 shows the calculated RF magnetic field for a single-turn loop driven at 300 MHz (55 mm in diameter) [19]. Figure 10.5a, b show the loop segments and the distribution of the RF current flowing in the loop. Since the length of the loop circumference is not negligibly small in comparison to the wavelength at 300 MHz, the RF current in the loop is no longer uniform. Figure 10.5c shows the calculated RF magnetic energy that contributes to EPR. As mentioned above, the RF magnetic energy (square of the RF magnetic field B1, see Eq. (10.7)) at the center of the loop is less than that in the proximity of the loop. To investigate the RF magnetic fields in the resonator, the finite-element method (FEM) or the finite-difference time-domain (FDTD) method can be used, and commercially available simulators for electromagnetic waves have been used to investigate the resonator and its RF electromagnetic fields. These simulations can then guide the manufacture, geometric dimensions and tolerance of resonators and their components to ensure that a given resonator design delivers maximal performance .

Fig. 10.5. Calculated RF magnetic field for a single-turn loop driven at 300 MHz (55 mm in diameter) [19].

Fig. 10.5

Calculated RF magnetic field for a single-turn loop driven at 300 MHz (55 mm in diameter) [19]. (a) Modeling and segments of the loop for the calculation of the current and the magnetic field, (b) RF current distribution on the loop, and (c) the distribution (more...)

Ease of Operation

In animal experiments or a clinical setting, sample/subject handling is an important practical issue. An animal should be properly placed in the resonator or a person must be placed comfortably to allow for proper resonator placement. If ease of operation of the resonator is not well-considered, measurements may be confounded by several factors, such as the resonator not being placed properly in or relative to the magnetic field, inability to tune/couple, etc. A good example is a pop-up retractable resonator for use with mouse tumor-bearing legs [20].

Specific applications have different needs and technical considerations. For example, in EPR-based tooth dosimetry, the external loop resonator should be placed on the surface of the incisor of the human subject. Since placement of the loop is critical for good reproducibility of the measurements, the mechanical holder of the resonator should be stable and the other mechanical elements in the measurement setup should be considered in the context of ease of operation. In such a case, the resonator used has to be considered in terms of operation and sample handling.

Brief History of Technical Developments for In Vivo EPR Resonators

In vivo EPR spectroscopy and imaging of small animals have a history of almost four decades. This section will briefly look back at the development of resonators for in vivo EPR with small animals.

First In Vivo EPR Experiment Using a Helix Coil

In a report by Feldman et al. in 1975, a helix coil was implanted in the liver of a rat, and EPR spectra from exogenously injected free radical spin probes were observed [5]. Helices were made of gold-coated brass wire and enclosed in thin-wall Teflon tubing. The diameter of helices was in the range 1.6–2.4 mm. The helix coil was connected to a Varian X-band spectrometer via a waveguide-to-coax adapter. EPR spectra were recorded at a magnetic field of 0.3 T, which corresponds to a microwave frequency of 8.5 GHz.

Surface Coils

A helix coil and a flat coil were used as resonators at 1.86 GHz for animals in vivo [11]. A single-turn flat loop coil was used for in vivo EPR imaging of a living murine tumor (Cloudman S-91 melanoma in the tail of a mouse) [6]. An electronically tunable surface coil operating at 3 GHz was used for in vivo skin measurements in a human forearm [21]. A surface coil (4 mm in diameter) at 2.4 GHz was used for EPR spectroscopy and imaging of the surface domain of a large subject [22]. In that report, Herrling et al. explained two imaging approaches using a surface coil. One approach uses mechanical scanning of the surface coil and the other uses a magnetic field gradient as a conventional spatial imaging. Skin measurements are a good application for EPR spectroscopy and imaging using surface coils. Another skin study was performed by Takeshita et al. with a surface coil at 1.1 GHz [23, 24]. Another surface coil resonator, called an external loop resonator, was used for small animal experiments [15]. The single-turn loop was placed on the surface of tissue within which the oxygen-sensitive crystal LiPc was implanted.

Loop-Gap Resonators

In 1982, Froncisz and Hyde reported a loop-gap resonator (LGR) for EPR in the range of 1–10 GHz. They demonstrated EPR spectroscopy with LGRs at 3 and 9 GHz [10]. Figure 10.6 shows the structure of the LGR in the literature [10]. It is considered to be a lumped circuit when the dimensions of the LGR are smaller than 1/4 wavelength. Also, Nardy and Whitehead reported a split-ring resonator in the context of NMR in a frequency range from 200 to 2000 MHz [25]. These resonators are lumped circuits and are considered to be an LC resonant circuit. The LGR has the advantage of a high filling factor, which results in a greater EPR signal intensity. In the context of in vivo small animal EPR, a part of or the whole body of a subject animal can be placed in the loop. As previously mentioned, the resonant frequency should be appropriately selected to give enough penetration depth for electromagnetic waves in biological tissues.

Fig. 10.6. Structure of the loop-gap resonator (two-gap one-loop) reported in 1982 [10].

Fig. 10.6

Structure of the loop-gap resonator (two-gap one-loop) reported in 1982 [10]. (Reprinted from Journal of Magnetic Resonance, Vol. 47, Froncisz W, Hyde JS, The loop-gap resonator: A new lumped circuit ESR sample structure, pp. 515–521, Copyright (more...)

A modified structure of the LGR, called a bridged loop-gap resonator (BLGR), was reported in 1988 [26]. Curved conductive plates (or thin foils) called bridges are placed near the gap(s). In Pfenninger’s BLGR, the bridge was located outside of the loops, and they intended to use BLGR for pulsed EPR at an X-band frequency. In the context of in vivo EPR, Ono et al. reported another structure for BLGR at an L-band frequency [27], in which electric shields are located inside the gaps to prevent the heating of biological tissues due to the electric field. LGR and BLGR have been used in small animal EPR spectroscopy and imaging [28, 29].

The design of an LGR is important for specific applications. Diodato et al. reported optimization of the axial RF field distribution in an LGR [30]. They showed that the coupling loop influenced the axial RF magnetic field and proposed the combination of two LGRs, with the coupling loop between the two LGRs. Ono et al. reported that electric shields influenced the RF magnetic field distribution in a BLGR [31]. The RF magnetic and electric fields inside the BLGR were measured. Their experiments clarified that appropriate angles of the electric shields can improve the homogeneity of the RF magnetic field .

Reentrant Resonators

Another type of resonator is a reentrant resonator . In 1983, Giordano et al. reported a reentrant cavity with a gap and a path for magnetic flux, which operated at 2 GHz [32]. They intended to use their reentrant resonator in studies on anisotropic line broadening. Later, Sotgiu and coworkers reported the designs of reentrant resonators that operated in the range of 2–10 GHz [33, 34]. Since reentrant resonators can operate at a low frequency, they have been used in small animal experiments. Figure 10.7 shows a cross-section of the reentrant resonator reported by Sotgiu in 1985 [35]. Chzhan et al. also developed reentrant resonators that operated at 1.2 GHz [36]. Chzhan et al. improved their reentrant resonator by making it capable of frequency tuning using piezoelectric actuators [14]. After that, Zweier and colleagues continued to develop reentrant resonators for EPR imaging in small animals at 750 MHz and 1.2 GHz [37, 38].

Fig. 10.7. Cross-section of a reentrant resonator reported in 1985 [35].

Fig. 10.7

Cross-section of a reentrant resonator reported in 1985 [35]. (Reprinted from Journal of Magnetic Resonance, Vol. 65, Sotgiu A, Resonator design for in vivo ESR spectroscopy, pp. 206–214, Copyright (1985), with permission from Elsevier)

Surface Coils for Teeth or Fingernails

In human subjects, resonators can be used for the retrospective measurement of radiation dose, called EPR dosimetry. Stable free radicals in enamel in teeth and keratin in fingernails can be measured by EPR [39]. In measurements of the teeth, the surface coil is placed on an incisor or a molar.

Regarding EPR tooth dosimetry , Hochi et al. reported an X-band EPR spectrometer that used a cavity resonator with a small aperture [40]. In their approach, a human subject bit the cavity resonator and the electromagnetic waves from the aperture of the cavity resonator excited free radicals in the enamel of a tooth [41]. While a cavity resonator in the X-band can provide highly sensitive EPR signals because of greater Zeeman energy splitting, it requires a significantly larger magnet to establish the necessary fields, as compared to a magnet for low-field EPR, such as for L-band measurements. A lower magnetic field generated by an air-core magnet is suitable in a portable EPR spectrometer for triage dosimetry [42]. A 1.1-GHz surface loop resonator was used for EPR-based human tooth dosimetry. This application is explained in another subsection of this chapter.

For EPR dosimetry with fingernails, EPR measurements require the shallow penetration of electromagnetic waves in the fingernail. This is because fingernails are approximately 1 mm thick, and deeper penetration leads to a loss of electromagnetic waves in tissue in the fingertips. To satisfy this requirement, surface coil resonators in the X-band (9.5 GHz) were developed for in vivo EPR fingernail dosimetry. Sidabras et al. developed a microwave surface resonator array (SRA) in the X-band that is suitable for EPR-based fingernail dosimetry [43]. Grinberg et al. also reported a dielectric-backed aperture resonator in the X-band for in vivo nail dosimetry [44].

Resonators for Preclinical (Small Animal) Studies

For EPR spectroscopy and imaging in small animals such as mice and rats, a variety of resonators have been reported. This section introduces some of the more commonly used resonators in preclinical studies involving small animals.

Volume Resonators

Loop-Gap Resonator

LGRs can accommodate the whole body or part of a mouse or a rat. An early work on an LGR in 1986 involved the measurement of the partial pressure of oxygen in the peritoneal cavity of a mouse. Subczynski et al. used a 1 GHz LGR (25 mm in diameter and 30 mm long) to accommodate the whole body of a mouse [45]. A BLGR (43 mm in diameter and 30 mm long) operating at 800 MHz was used for measurement of the rat head. Ishida et al. reported the time-course of EPR signal intensities for the nitroxyl radical spin probes CTPO and TEMPOL [28]. A whole-body resonator based on an LGR (30 mm in diameter) at 1200 MHz was used for EPR spectroscopy in mice and rats [16]. For cancer studies using mouse tumor models, the LGR has been intensively used for 250 MHz CW and pulsed EPR imaging. For example, an LGR (16 mm in diameter and 15 mm long) was used for four-dimensional (4D) spectral-spatial EPR imaging to visualize the oxygen partial pressure in tumor-bearing mouse legs [4648]. In addition to the CW protocol, the LGR has been used in pulsed EPR at 250 MHz for imaging of oxygen in tumor-bearing mouse legs [49, 50]. This LGR operating at 250 MHz was specifically designed for tumor-bearing mouse legs . Therefore, the sample space is rather small in comparison to other LGRs.

Reentrant Resonator

Many biomedical EPR studies have used reentrant resonators. Alecci et al. visualized nitroxyl radical probes in the rat tail by using three-dimensional EPR imaging and a 1.2-GHz reentrant resonator (sample space of 12 mm in diameter and 24 mm in length) [51]. He et al. reported EPR imaging in the beating heart of a mouse with a reentrant resonator operating at 1.2 GHz [37]. This resonator has varactor-based tuning and impedance-matching capabilities to compensate for the motion caused by the heartbeat of the mouse. A reentrant resonator operating at 750 MHz was used for EPR imaging of the whole-body of a mouse [52]. A 1.3-GHz ceramic three-loop two-gap reentrant resonator (for a sample 20 mm in diameter) [36] was used for EPR spectroscopy and imaging of nitric oxide generation in the mouse [53]. The same reentrant resonator was also used for oxygen mapping in the rat tail [54]. The gastrointestinal tract of a living mouse was visualized using EPR imaging and a 750-MHz tunable reentrant resonator [55].

Parallel Coil Resonator

In low-field pulsed EPR , the ring-down time of the resonator is critical [56]. After an RF pulse is transmitted to the resonator and a subject, the RF energy in the resonator remains stored for a period and disturbs signal detection at the receiver of a pulsed EPR instrument. This signal disturbance after the RF pulse is called dead time and is related to the ring-down time of the resonator. To overcome the problem regarding ringing after an RF pulse, the quality factor of the resonator is significantly reduced. A parallel coil resonator at 300 MHz has been reported by Devasahayam et al. [57]. The parallel coil resonator has been used in mouse whole-body imaging [58] and pO2 mapping for tumor-bearing legs [59] at 300 MHz. A low quality factor (20–25) was achieved by over-coupling of the resonator [58]. The parallel coil resonator operating at 300 MHz could be used for EPR and 7 T proton MRI, which is important for EPR/NMR co-registration imaging of an animal subject.

Figure 10.8 shows a schematic of the parallel coil resonator [57]. The coils are connected to the coupling and tuning capacitors. Trimmer capacitors are adjusted to obtain the desired resonance frequency and a degree of coupling for the resonator. Since pulsed EPR does not require magnetic field modulation, a modulation coil is no longer needed. Thus, the structure of the parallel coil resonator is simple in comparison to a resonator with a modulation coil.

Fig. 10.8. Circuitry and schematic of a parallel coil resonator used for 300-MHz pulsed EPR [57].

Fig. 10.8

Circuitry and schematic of a parallel coil resonator used for 300-MHz pulsed EPR [57]. (Reprinted from Journal of Magnetic Resonance, Vol. 142, Devasahayam N, Subramanian S, Murugesan R, Cook JA, Afeworki M, Tschudin RG, Mitchell JB, Krishna MC, Parallel (more...)

Alderman-Grant Resonator

The Alderman-Grant resonator (AGR) has been used in EPR and NMR. An important characteristic of the AGR is that the RF magnetic field is perpendicular to the axis of the AGR. The AGR was first reported by Alderman and Grant in 1979 [60]. In EPR-related studies, AGR has been used to excite electron spins in proton-electron double resonance imaging (PEDRI), which is also called Overhauser-enhanced magnetic resonance imaging (OMRI). Petryakov et al. developed an AGR at 590 MHz for exciting electron spins and demonstrated experiments with a high input power (up to 60 W for 3 min and 10 W for 10 min) [61]. Figure 10.9 illustrates the structure of the original AGR. This resonator can be used with other types of resonator such as a solenoid-type resonator because the RF magnetic fields are orthogonal to each other. This is necessary for double-resonance techniques (PEDRI/OMRI) [62].

Fig. 10.9. Schematic of the Alderman-Grant resonator reported in 1979 [60].

Fig. 10.9

Schematic of the Alderman-Grant resonator reported in 1979 [60]. (Reprinted from Journal of Magnetic Resonance, Vol. 36, Alderman DW, Grant DM, An efficient decoupler coil design which reduces heating in conductive samples in superconducting spectrometers, (more...)

Surface Resonator

Surface coil resonators have been used for measurements of the superficial regions of subject animals. A surface coil can be located at the region of interest for a measurement, even if a volume coil such as an LGR cannot accommodate a large subject. The simplest surface coil is just a combination of a single-turn loop and a capacitor to form a resonant circuit. However, additional circuitry is usually added to the coil to make it both possible and practical to adjust the resonant frequency and enable impedance matching. This section introduces several concepts regarding surface coil resonators.

Surface Coil/Loop Resonators

An electronically tunable resonator is useful for low-field EPR because perturbation due to motion of the subject should be accommodated. Figure 10.10 shows a circuit diagram of an electronically tunable surface coil resonator at 1.1 GHz [63]. This resonator uses varactor diodes for impedance matching and frequency adjustment. Also, a half-wavelength balun was used to ensure that the resonant circuit was properly balanced. A pair of varactor diodes that are connected in the opposite direction can increase the ability of the resonator to tolerate an incident RF power. This reduces the generation of the harmonics of incident RF signals. This type of surface coil resonator can be used for in vivo EPR spectroscopy .

Fig. 10.10. Schematic of a tunable surface coil resonator reported in 2003 [63].

Fig. 10.10

Schematic of a tunable surface coil resonator reported in 2003 [63]. (Reprinted from Journal of Magnetic Resonance, Vol. 164, Hirata H, Kuyama T, Ono M, Shimoyama Y, Detection of electron paramagnetic resonance absorption using frequency modulation, pp. (more...)

A surface coil resonator is also called an external loop resonator (ELR). Figure 10.11 shows a circuit diagram of an ELR at 1.1 GHz [16, 64]. Varactor diodes are used to adjust the resonant frequency, and a photo-resistor and a light-emitting diode (LED) are used to control the quality factor. A shift in the quality factor can control impedance-matching between the 50-Ω transmission line and the resonator. Thus, the voltage applied to the LED controls the impedance of the ELR. An inductive coupling scheme is also used in the ELR. The length of a twisted wire connected to the inductive coupling loop is adjusted to minimize the shift of the resonant frequency due to the adjustment of impedance-matching [65].

Fig. 10.11. Circuit diagram of the external loop resonator (ELR) reported in 2005 [16].

Fig. 10.11

Circuit diagram of the external loop resonator (ELR) reported in 2005 [16]. (Reprinted from Walczak T, Lesniewski P, Salikhov I, Sucheta A, Szybinski K, Swartz HM. 2005. L-band electron paramagnetic resonance spectrometer for use in vivo and in studies (more...)

An ELR has been used for in vivo L-band EPR spectroscopy. However, the details of the ELR were published in 2005. More recently, the ELR was modeled and analyzed by the FEM [66]. Sugawara et al. reported a FEM analysis and optimization of a 1.1-GHz surface coil resonator for EPR tooth dosimetry [67].

Surface Coil Array

To overcome the limited sensitive region of the surface coil resonator, an array of surface coil resonators can be used. This approach is also used in clinical MRI. Due to the higher RF frequency in EPR than in NMR/MRI, the size of the surface coil is usually limited. If the circumference of the loop is not negligibly short in comparison to the wavelength, i.e., <1/4 wavelength, of the RF electromagnetic waves, the fields generated by the loop are no longer in the correct phase relative to each other. This leads to an inhomogeneous RF magnetic field in the loop. This section briefly introduces a surface coil array for in vivo EPR imaging.

There are two approaches to image acquisition using a surface coil array : (1) sequential acquisition and (2) simultaneous (parallel) acquisition. Sequential acquisition refers to when one of the surface coil resonators within the array is selected for EPR signal recording. This is a simple approach to EPR imaging. However, the interaction between the neighboring coils should be suppressed to avoid interference while an image is being acquired. This is because inductive coupling between coils that are located in close proximity affects the resonance characteristics of the surface coil resonator to which the RF electromagnetic waves are fed. Enomoto et al. used PIN-diode switches to suppress the interaction between the neighboring resonators in a CW-EPR detection protocol [68, 69]. Figure 10.12 shows a four-channel surface coil array operating at 750 MHz. This array was tested with a living mouse [69]. The array was placed on the back of a subject mouse. After the mouse was injected with spin probes, EPR image-acquisition was performed with each surface coil in sequence. The reconstructed images were then combined. In pulsed EPR, the quality factor of the resonator is significantly decreased. This can reduce the influence of the interaction between the neighboring coils. Enomoto et al. demonstrated a surface coil array in pulsed EPR imaging at 300 MHz. As in CW-EPR, the resonant frequencies of unexcited resonators in the array were shifted to reduce the mutual inductive coupling of the coils [70]. In addition, passive decoupling of the coils was also tested for a four-channel surface coil array [71]. The feasibility of the simultaneous acquisition of multiple coils has also been reported [72]. Development of a surface coil array for EPR imaging is an important direction for preclinical and clinical applications .

Fig. 10.12. Photograph of a four-channel surface coil array operating at 750 MHz for EPR imaging in mice [69].

Fig. 10.12

Photograph of a four-channel surface coil array operating at 750 MHz for EPR imaging in mice [69]. (Reprinted from Journal of Magnetic Resonance, Vol. 234, Enomoto A, Emoto M, Fujii H, Hirata H, Four-channel surface coil array for sequential CW-EPR image (more...)

Surface Detection Using LGR

Another approach to detecting EPR signals from a superficial region in biological subjects uses the fringe magnetic field of a volume coil resonator such as the LGR instead of a surface coil. A surface probe using a double split-ring resonator operating at 1 GHz was reported by Nilges et al. [73]. In a more clinically related application, EPR imaging for human skin was performed with a surface probe using a BLGR in the S-band (2.2 GHz) [8, 9]. With a higher frequency in the S-band, nitroxide probes in human skin were successfully detected and the distribution of nitroxide radicals was visualized. While the frequency of 2 GHz limits the penetration depth of electromagnetic waves in biological tissues, measurements in human skin do not require deep penetration by electromagnetic waves .

Implantable Resonator

As mentioned above, surface coils have a limited sensitive volume, although there are some advantages of a comparatively high sensitivity and no limitation to the subject size. When a region of interest is deep in tissue that is a part of a large object, neither a surface coil nor a volume coil, e.g., the LGR, can be applied. In particular, the internal organs of human subjects are difficult to measure with either a surface coil or an LGR with a commonly used low-field EPR spectrometer/imager. One possible solution is an implantable resonator.

Li et al. demonstrated an implantable resonator for an EPR oximetry study in rat brain [74]. The implantable resonator has oxygen-sensitive materials in a small loop or multiple small loops that are located at the end of a twisted pair of transmission lines [75]. Figure 10.13 shows a photograph of an implantable resonator for rabbit brain [76]. The length of the twisted wires can be adjusted to reach the region of interest. The other end of the twisted wires has a larger loop that can be coupled with the loop located on the surface of the object.

Fig. 10.13. Schematic of implantable resonators for rabbit brain [76].

Fig. 10.13

Schematic of implantable resonators for rabbit brain [76]. (Reprinted from Stroke, Vol. 46, Khan N, Hou H, Eskey CJ, Moodie K, Gohain S, Du G, Hodge S, Culp WC, Kuppusamy P, Swartz HM, Deep-tissue oxygen monitoring in the brain of rabbits for stroke research, (more...)

RF signals can be delivered to the oxygen-sensitive probe in a small loop (or multiple loops) of the implantable device. After the implantable device is placed in the subject, EPR spectroscopy can be repeatedly conducted noninvasively. The small loop can increase the sensitivity of EPR detection because a strong RF magnetic field is generated in a small loop. This phenomenon is a primary reason why an implantable resonator can work effectively in deep tissues. With an implantable resonator with multiple loops, the partial pressure of oxygen at multiple sites can be measured, and, for example, it is possible to compare oxygenation at several regions of interest in rat brain. These measurements could be useful for studies in an animal model of ischemic stroke [75, 76] or in a tumor xenograft model [77, 78].

From an engineering point of view, the materials used to make an implantable resonator should be biocompatible if the resonator is to be placed in biological tissues for a long time. An implantable resonator made of copper can be coated with a biocompatible material. For oxygen measurements, the coating materials should be oxygen-permeable.

Resonators for In Vivo EPR in Human Subjects

Surface Resonators for Human Skin/Surface

EPR spectroscopy and imaging on human skin have been reported by He et al. [8]. They reported the distribution and metabolism of nitroxide radicals in human skin using a loop-gap surface resonator operating at 2.2 GHz [9]. A penetration depth of electromagnetic waves of approximately 2 mm is enough for skin EPR measurements. Reports of in vivo EPR measurements in human skin are still limited. For in vivo studies in human skin, the microwave frequency can be higher than conventional in vivo frequencies (L-band or lower RF frequencies), since the limited penetration depth is not a problem, but rather a desirable feature, for studies in skin. Successful applications of EPR spectroscopy and imaging in human skin help foster the further development of surface resonators. Wolfson et al. developed an LGR-based surface-detection resonator operating at 2.3 GHz [79]. This resonator was intended for transcutaneous oxygen-monitoring using the oxygen-sensitive crystal LiNc-BuO. The dimensions of this resonator are 13 mm in outer diameter, 3.6 mm in inner diameter, and 19 mm in height. The gap in the loop is 0.1 mm. This LGR was installed in an array of small magnets that generate a static magnetic field .

L-Band Surface Loop Resonators for Tooth Dosimetry

When teeth receive ionizing radiation in medical treatments, or because of other reasons such as an accident in a nuclear power plant, very stable CO2 radical is generated in the enamel [39]. EPR spectroscopy can detect these radiation-induced free radicals in human teeth after they have been exposed to ionizing radiation. Surface coil resonators have been developed for EPR-based in vivo tooth dosimetry.

An ELR at 1.15 GHz was developed and used for human EPR tooth dosimetry [80]. To improve the detection capabilities to more accurately estimate the radiation dose, surface coil resonators were further developed [66, 67, 81]. The loop was placed on an incisor to detect free radicals in the enamel. Recently, a unique EPR resonator was developed for tooth dosimetry. It consisted of a combination of an inductively coupled ELR printed on a flexible substrate which was driven by a non-resonant quarter-wavelength-loop antenna [82]. The printed ELR can be affixed to an incisor for an in vivo measurement. The advantages of this wireless flexible surface coil resonator are as follows:

(a)

A printed ELR can be mass-produced to have a low unit cost, and therefore be disposable. This is important for biomedical applications and high throughput if a large number of people are to be measured in a short period of time, e.g., a triage operation in the event of a nuclear incident.

(b)

A printed ELR can be easily affixed to a tooth via suitable medical double-sided tape outside of the magnet, reducing the need for cumbersome placement inside the restricted space of the magnet.

(c)

The field generated by the antenna coupler encompasses both of the front incisor teeth, and therefore precise placement of the coupler inside the magnet during the measurement is not necessary—the coupler can simply be placed universally for all measurement subjects, reducing the burden on the operator and increasing the potential for automated actions of the EPR dosimeter.

Resonators for tooth dosimetry were tested in human subjects [83]. Since tooth dosimetry is a topical EPR measurement, the small surface loop operating at 1.1–1.2 GHz is not limited by the penetration depth of electromagnetic waves, and loss of the tooth itself is not so significant in comparison to the loss of other soft biological tissues [84].

X-Band Resonators for Fingernail Dosimetry

Several types of resonators have also been developed for another EPR-based dosimetry application: in vivo fingernail dosimetry. Fingernail dosimetry is commonly performed in the X-band, as radiation-induced signals in the fingernails are significantly weaker than those in tooth enamel. This increased frequency provides a high sensitivity of EPR detection due to a greater energy gap of Zeeman splitting. Since the region of EPR spectroscopy detection in the X-band is superficial and is suitable only for nails.

The radiation-induced EPR signal in human fingernails has been reported [85]. Radiation-induced signals in chipped fingernails were initially measured with a conventional X-band cavity resonator to qualify and generally characterize the dosimetry technique as a suitable candidate for physically based dosimetry [86]. This approach of using a microwave cavity is not suitable for in vivo human fingernails, as the conventional cavity resonator is designed for glass capillary tubes. Therefore, X-band resonators have been developed for in vivo studies of fingernails.

A microwave surface resonator array (SRA) has been developed for in vivo EPR nail dosimetry as a practical in vivo replacement for a conventional microwave cavity [43]. Figure 10.14 shows a photograph of an SRA for EPR fingernail dosimetry. This metallic arrayed structure was designed to limit the depth of penetration of the stimulating magnetic field so as to focus the field generated by the SRA into the fingernail itself rather than the underlying tissue. The intent of this is to help ensure that no energy applied to the resonator is then deposited into tissues where there is not expected to be any signal, i.e., the tissue under the nail. For a given nail dosimetry measurement, the resonator is simply placed on the fingernail within (or outside) the magnet, and then the finger and resonator assembly are then mated to an ergonomic receptacle within the EPR dosimeter .

Fig. 10.14. Photograph of an SRA for EPR fingernail dosimetry [43].

Fig. 10.14

Photograph of an SRA for EPR fingernail dosimetry [43]. (Reprinted from Sidabras JW, Varanasi SK, Mett RR, Swarts SG, Swartz HM, Hyde JS. 2014. A microwave resonator for limiting depth sensitivity for electron paramagnetic resonance spectroscopy of surfaces. (more...)

Several resonators that use dielectric resonant structures have also been developed for the purpose of surface measurements at an X-band frequency, e.g., in vivo nail dosimetry. These dielectric resonator assemblies seek to more closely match the dielectric of the resonant structure to the dielectric of the tissue being measured to maximize the amount of RF energy transferred into the tissue of interest, i.e., the fingernail. Proper high-dielectric substrates also provide the opportunity for a higher quality factor for a resonator, given a suitable selection of dielectric resonator for the application. For the dielectric resonator in Ref. [44], RF power is driven to the dielectric resonator via the use of a modified resonant microwave cavity, and Ref. [87] describes a resonator topology where RF power is driven to the dielectric resonator via the use of a non-resonant antenna.

Implantable Resonators

Another direction for resonator development in human subjects is to measure EPR signals from tissues that are not on the surface of the body or superficial to the surface, i.e., at depths >10 mm below the skin. This is primarily of concern when attempting to measure oxygen levels in a given organ in vivo in a noninvasive fashion. While resonator implantation is invasive, repeated measurements can be performed by placing an ELR over the coupling loop of the implanted resonator [88] and collecting EPR data as in a usual surface EPR oximetry measurement.

A possible solution is the use of an implantable resonator to detect EPR signals deep in tissue. A clinically acceptable implantable EPR resonator is currently under development [89], and an implantable resonator for a rat brain [75, 76, 78] is a similar model of such an implantable resonator for human subjects. A rabbit brain was also measured with an implantable resonator [77]. Implantable coils have been investigated in proton magnetic resonance imaging (MRI) [90, 91]. The implantable resonator being developed in Ref. [89] seeks to accommodate and address aspects of clinical acceptance, including insertion/removal procedures, biocompatibility, MRI compatibility, and other aspects associated with obtaining an investigation device exemption from the U.S. Food and Drug Administration (FDA) with the goal of bringing implantable resonators closer to reality in the clinic. The development of implantable resonators for use in EPR spectroscopy and imaging in human subjects is a potential direction for future studies .

Future Perspective

From the viewpoint of clinical EPR, the development of resonators has several directions that can contribute to the future clinical use of EPR spectroscopy and imaging.

A Resonator for Superficial Tissues

As mentioned above, this is the most practical target for present EPR resonators. For example, skin, teeth, and fingernails are easy to access. When the RF frequency for EPR decreases to an RF frequency similar to that in clinical MRI, the problem of the penetration depth of electromagnetic waves may be partly solved; however, the absolute EPR amplitude sensitivity will suffer. A small resonator used for endoscope-like EPR measurement may also be considered for measurements of superficial regions.

An Implantable Resonator

Surgery is required to place implantable resonators deep in tissue. This is a drawback in clinical EPR. However, once placed, implantable resonators enable repeated, noninvasive measurements in EPR spectroscopy and imaging. The risks and benefits of implantable resonators should be considered, as with any other clinical treatment. However, technologies regarding implantable resonators should be explored further to achieve clinical EPR, as each implantable resonator should meet the specific needs of the clinical application of EPR spectroscopy and imaging .

A Volume Coil Resonator for Further Low-Frequency Clinical EPR

To overcome the limited penetration depth of electromagnetic waves in tissues, a further low RF frequency might be used. MRIs with magnetic fields of 1.5 T and 3 T operate at 64 MHz and 128 MHz, respectively. By analogy with MRI, such a lower RF frequency provides better penetration of electromagnetic waves in tissue as well as a smaller energy gap in Zeeman splitting, thereby resulting in a decrease in the sensitivity of EPR detection. In contrast to a smaller energy gap, a volume coil operating at a frequency similar to that in clinical MRI can accommodate a larger subject. This might compensate for the weaker EPR signals at a lower magnetic field and a corresponding RF frequency. Both the resonator and other parts of EPR instruments should be further improved to achieve a reasonable sensitivity of EPR detection at a low RF frequency to approach that of clinical MRI. Whole-body human EPR imaging, like proton MRI, is a dream in the application of EPR spectroscopy and imaging.

The future development of resonators for clinical EPR is not limited to these directions. Innovative technologies in RF/microwave electronics and advanced manufacturing techniques such as 3D printing may provide breakthroughs for EPR resonators in the future.

References

1.
Rinard GA, Eaton SS, Eaton GR. Sensitivity (Chap. 1). In: Poole Jr CP, Farach HA, editors. Handbook of electron spin resonance, vol. 2. New York: Springer; 1999.
2.
Feher G. Sensitivity considerations in microwave paramagnetic resonance absorption techniques. Bell System Technol J. 1957;36:449–84.
3.
Weil JA, Bolton JR. Electron paramagnetic resonance. 2nd ed. Hoboken, NY: Wiley; 2007. p. 1–35.
4.
Griffiths DJ. Introduction to electrodynamics. 2nd ed. Englewood Cliffs, NJ: Prentice Hall; 1989. p. 369–72.
5.
Feldman A, Wildman E, Bartolinini G, Piette LH. In vivo electron spin resonance in rats. Phys Med Biol. 1975;20:602–12. [PubMed: 171686]
6.
Berliner LJ, Fujii H, Wan X, Lukiewicz SJ. Feasibility study of imaging a living murine tumor by electron paramagnetic resonance. Magn Reson Med. 1987;4:380–4. [PubMed: 3035320]
7.
Eaton GR, Eaton SS. EPR spectrometers at frequencies below X-band (Chap. 2). In: Bender CJ, Beriner LJ, editors. EPR instrumental methods. Biological magnetic resonance, vol. 21. New York: Kluwer Academic/Plenum; 2004.
8.
He G, Samouilov A, Kuppusamy P, Zweier JL. In vivo EPR imaging of the distribution and metabolism of nitroxide radicals in human skin. J Magn Reson. 2001;148:155–64. [PubMed: 11133289]
9.
Petryakov S, Chzhan M, Samouilov A, He G, Kuppusamy P, Zweier JL. A bridged loop-gap S-band surface resonator for topical EPR spectroscopy. J Magn Reson. 2001;151:124–8. [PubMed: 11444946]
10.
Froncisz W, Hyde JS. The loop-gap resonator: a new lumped circuit ESR sample structure. J Magn Reson. 1982;47:515–21.
11.
Nishikawa H, Fujii H, Beriliner LJ. Helices and surface coils for low-field in vivo ESR and EPR imaging applications. J Magn Reson. 1985;62:79–86.
12.
Eaton GR, Eaton SS, Barr DP, Weber RT. Quantitative EPR (Chap. 7). Wien: Springer; 2010.
13.
Brivati JA, Stevens AD, Symons MCR. A radiofrequency ESR spectrometer for in vivo imaging. J Magn Reson. 1991;92:480–9.
14.
Chzhan M, Kuppusamy P, Zweier JL. Development of an electronically tunable L-band resonator for EPR spectroscopy and imaging of biological samples. J Magn Reson Ser B. 1995;108:67–72. [PubMed: 7627435]
15.
Hirata H, Walczak T, Swartz HM. Electronically tunable surface-coil-type resonator for L-band EPR spectroscopy. J Magn Reson. 2000;142:159–67. [PubMed: 10617447]
16.
Walczak T, Lesniewski P, Salikhov I, Sucheta A, Szybinski K, Swartz HM. L-band electron paramagnetic resonance spectrometer for use in vivo and in studies of aqueous biological samples. Rev Sci Instrum. 2005;76:013107.
17.
Rinard GA, Quine RW, Eaton SS, Eaton GR. Microwave coupling structures for spectroscopy. J Magn Reson Ser A. 1993;105:137–44.
18.
McCallum S, Resmer F. Automatic coupling control system for radio frequency in vivo electron paramagnetic resonance based on a piezoelectric controlled capacitor. Rev Sci Instrum. 1999;70:4706–10.
19.
Hirata H, He G, Deng Y, Salikhov I, Petryakov S, Zweier JL. A loop resonator for slice-selective in vivo EPR imaging in rats. J Magn Reson. 2008;190:124–34. [PMC free article: PMC2615245] [PubMed: 18006343]
20.
Epel B, Subramanian VS, Halpern HJ. Retractable loop-gap resonator for electron paramagnetic resonance imaging with in situ irradiation capabilities. Concepts Magn Reson B. 2011;39B:167–72.
21.
Herrling TE, Groth NK, Fuchs J. Biochemical EPR imaging of skin. Appl Magn Reson. 1996;11:471–86.
22.
Herrling T, Rehberg J, Jung K, Groth N. SURF_ER–surface electron spin resonance (ESR) of surface domain of large subjects. Spectrochim Acta A. 2002;58:1337–44. [PubMed: 11993481]
23.
Takeshita K, Takajo T, Hirata H, Ono M, Utsumi H. In vivo oxygen radical generation in the skin of the protoporphyria model mouse with visible light exposure: an L-band ESR study. J Invest Dermatol. 2004;122:1463–70. [PubMed: 15175038]
24.
Takeshita K, Chi C, Hirata H, Ono M, Ozawa T. In vivo generation of free radical in the skin of living mouse under ultraviolet light measured by L-band EPR spectroscopy. Free Radic Biol Med. 2006;40:876–85. [PubMed: 16520239]
25.
Nardy WN, Whitehead LA. Split-ring resonator for use in magnetic resonance from 200–2000 MHz. Rev Sci Instrum. 1981;52:213–6.
26.
Pfenninger S, Forrer J, Schweiger A, Weiland T. Bridged loop-gap resonator: a resonant structure for pulsed ESR transparent to high-frequency radiation. Rev Sci Instrum. 1988;59:752–60.
27.
Ono M, Ogata T, Hsieh K-C, Suzuki M, Yoshida E, Kamada H. L-band ESR spectrometer using a loop-gap resonator for in vivo analysis. Chem Lett. 1986;15:491–4.
28.
Ishida S, Kumashiro H, Tsuchihashi N, Ogata T, Ono M, Kamada H, Yoshida E. In vivo analysis of nitroxide radicals injected into small animals by L-band ESR technique. Phys Med Biol. 1989;34:1317–23. [PubMed: 2552481]
29.
Ishida S, Matsumoto S, Yokoyama H, Mori N, Kumashiro H, Tsuchihashi N, Ogata T, Yamaga M, Ono M, Kitajima T, Kamada H, Yoshida E. An ESR-CT imaging of the head of a living rat receiving an administration of a nitroxide radical. Magn Reson Imaging. 1992;10:109–14. [PubMed: 1312195]
30.
Diodato R, Alecci M, Brivati JA, Varoli V, Sotgiu A. Optimization of axial RF field distribution in low-frequency EPR loop-gap resonators. Phys Med Biol. 1999;44:N69–75. [PubMed: 10368029]
31.
Ono M, Suenaga A, Hirata H. Experimental investigation of RF magnetic field homogeneity in a bridged loop-gap resonator. Magn Reson Med. 2002;47:415–9. [PubMed: 11810689]
32.
Giordano M, Momo F, Sotgiu A. On the design of a reentrant square cavity as resonator for low-frequency ESR spectroscopy. J Phys E Sci Instrum. 1983;16:774–9.
33.
Momo F, Sotgiu A. Re-entrant resonators for ESR spectroscopy between 2 and 10 GHz. J Phys E Sci Instrum. 1984;17:556–8.
34.
Sotgiu A, Gualtieri G. Cavity resonator for in vivo electron-spin-resonance spectroscopy. J Phys E Sci Instrum. 1985;18:899–901.
35.
Sotgiu A. Resonator design for in vivo ESR spectroscopy. J Magn Reson. 1985;65:206–14.
36.
Chzhan M, Shteynbuk M, Kuppusamy P, Zweier JL. An optimized L-band ceramic resonator for EPR imaging of biological samples. J Magn Reson Ser A. 1993;105:49–53.
37.
He G, Petryakov S, Samoulilov A, Chzhan M, Kuppusamy P, Zweier JL. Development of a resonator with automatic tuning and coupling capability to minimize sample motion noise for in vivo EPR spectroscopy. J Magn Reson. 2001;149:218–27. [PubMed: 11318620]
38.
He G, Dumitrescu C, Petryakov S, Deng Y, Kesselring E, Zweier JL. Transverse oriented electric field reentrant resonator (TERR) with automatic tuning and coupling control for EPR spectroscopy and imaging of the beating heat. J Magn Reson. 2007;187:57–65. [PubMed: 17462925]
39.
Swartz HM, Williams BB, Flood AB. Overview of the principles and practice of biodosimetry. Radiat Environ Biophys. 2014;53:221–32. [PMC free article: PMC5982531] [PubMed: 24519326]
40.
Hochi A, Furusawa M, Ikeya M. Applications of microwave scanning ESR microscope: human tooth with metal. Appl Radiat Isot. 1993;44:401–5. [PubMed: 8386044]
41.
Ishii H, Ikeya M. An electron spin resonance system for in-vivo human tooth dosimetry. Jpn J Appl Phys. 1990;29:871–5.
42.
Williams BB, Flood AB, Salikhov I, Kobayashi K, Dong R, Rychert K, Du G, Schreiber W, Swartz HM. In vivo EPR tooth dosimetry for triage after a radiation event involving large populations. Radiat Environ Biophys. 2014;53:335–46. [PubMed: 24711003]
43.
Sidabras JW, Varanasi SK, Mett RR, Swarts SG, Swartz HM, Hyde JS. A microwave resonator for limiting depth sensitivity for electron paramagnetic resonance spectroscopy of surfaces. Rev Sci Instrum. 2014;85:104707. [PMC free article: PMC4201662] [PubMed: 25362434]
44.
Grinberg O, Sidabras JW, Tipikin DS, Krymov V, Mariani M, Feldman MM, Kmiec MM, Petryakov SV, Brugger S, Carr B, Schreiber W, Swarts SG, Swartz HM. Dielectric-backed aperture resonators for X-band in vivo EPR nail dosimetry. Radiat Prot Dosim. 2016;172:121–6. [PMC free article: PMC5225980] [PubMed: 27412507]
45.
Subczynski WK, Lukiewicz S, Hyde JS. Murine in vivo L-band ESR spin-label oximetry with a loop-gap resonator. Magn Reson Med. 1986;3:747–54. [PubMed: 3023786]
46.
Elas M, Williams BB, Parasca A, Malier C, Pelizzari CA, Lewis MA, River JN, Karczmar GS, Barth ED, Halpern HJ. Quantitative tumor oxymetric images from 4D electron paramagnetic resonance imaging (EPRI): methodology and comparison with blood oxygen level-dependent (BOLD) MRI. Magn Reson Med. 2003;49:682–91. [PubMed: 12652539]
47.
Elas M, Ahn KH, Parasca A, Barth ED, Lee D, Haney C, Halpern HJ. Electron paramagnetic resonance oxygen images correlate spatially and quantitatively with oxylite oxygen measurements. Clin Cancer Res. 2006;12:4209–17. [PubMed: 16857793]
48.
Elas M, Bell R, Hleihel D, Barth ED, McFaul C, Haney CR, Bielanska J, Pustelny K, Ahn KH, Pelizzari CA, Kocherginsky M, Halpern HJ. Electron paramagnetic resonance oxygen image hypoxic fraction plus radiation dose strongly correlates with tumor cure in FSa fibrosarcomas. Int J Radiat Oncol Biol Phys. 2008;71:542–9. [PMC free article: PMC2577780] [PubMed: 18474313]
49.
Epel B, Sundramoorthy SV, Mailer C, Halpern HJ. A versatile high speed 250-MHz pulse imager for biomedical applications. Concepts Magn Reson B. 2008;33B:163–76. [PMC free article: PMC2778030] [PubMed: 19924261]
50.
Elas M, Magwood JM, Butler B, Li C, Wardak R, Barth ED, Epel B, Rubinstein S, Pelizzari CA, Weichselbaum RR, Halpern HJ. EPR oxygen images predict tumor control by a 50% tumor control radiation dose. Cancer Res. 2013;73:5328–35. [PMC free article: PMC3913470] [PubMed: 23861469]
51.
Alecci M, Colacicchi S, Indovina PL, Momo F, Pavone P, Sotgiu A. Three-dimensional in vivo ESR imaging in rats. Magn Reson Imaging. 1990;8:59–63. [PubMed: 2157931]
52.
He G, Deng Y, Li H, Kuppusamy P, Zweier JL. EPR/NMR co-imaging for anatomic registration of free-radical images. Magn Reson Med. 2002;47:571–8. [PubMed: 11870845]
53.
Kuppusamy P, Shankar RA, Roubaud VM, Zweier JL. Whole body detection and imaging of nitric oxide generation in mice following cardiopulmonary arrest: detection of intrinsic nitrosoheme complexes. Magn Reson Med. 2001;45:700–7. [PubMed: 11283999]
54.
Velan SS, Spencer RGS, Zweier JL, Kuppusamy P. Electron paramagnetic resonance oxygen mapping (EPROM): direct visualization of oxygen concentration in tissue. Magn Reson Med. 2000;43:804–9. [PubMed: 10861874]
55.
He G, Shankar RA, Chzhan M, Samouilov A, Kuppusamy P, Zweier JL. Noninvasive measurement of anatomic structure and intraluminal oxygenation in the gastrointestinal tract of living mice with spatial and spectral EPR imaging. Proc Natl Acad Sci U S A. 1999;96:4586–91. [PMC free article: PMC16376] [PubMed: 10200306]
56.
Subramanian S, Matsumoto K, Mitchell JB, Krishna MC. Radio frequency continuous-wave and time-domain EPR imaging and Overhauser-enhanced magnetic resonance imaging of small animals: instrumental developments and comparison of relative merits for functional imaging. NMR Biomed. 2004;17:263–94. [PubMed: 15366027]
57.
Devasahayam N, Subramanian S, Murugesan R, Cook JA, Afeworki M, Tschudin RG, Mitchell JB, Krishna MC. Parallel coil resonators for time-domain radiofrequency electron paramagnetic resonance imaging of biological objects. J Magn Reson. 2000;142:168–76. [PubMed: 10617448]
58.
Subramanian S, Devasahayam N, Murgesan R, Yamada K, Cook J, Taube A, Mitchell JB, Lohman JAB, Krishna MC. Single-point (constant-time) imaging in radiofrequency Fourier transform electron paramagnetic resonance. Magn Reson Med. 2002;48:370–9. [PubMed: 12210946]
59.
Matsumoto S, Hyodo F, Subramanian S, Devasahayam N, Munasinghe J, Hyodo E, Gadisetti C, Cook JA, Mitchekk JB, Krishna MC. Low-field paramagnetic resonance imaging of tumor oxygenation and glycolytic activity in mice. J Clin Invest. 2008;118:1965–73. [PMC free article: PMC2323191] [PubMed: 18431513]
60.
Alderman DW, Grant DM. An efficient decoupler coil design which reduces heating in conductive samples in superconducting spectrometers. J Magn Reson. 1979;36:447–51.
61.
Petryakov S, Samouilov A, Roytenberg M, Li H, Zweier JL. Modified Alderman-Grant resonator with high-power stability for proton electron double resonance imaging. Magn Reson Med. 2006;56:654–9. [PubMed: 16902975]
62.
Lurie DJ, Davis GR, Foster MA, Hutchison JMS. Field-cycled PEDRI imaging of free radicals with detection at 450 mT. Magn Reson Imaging. 2005;23:175–81. [PubMed: 15833609]
63.
Hirata H, Kuyama T, Ono M, Shimoyama Y. Detection of electron paramagnetic resonance absorption using frequency modulation. J Magn Reson. 2003;164:233–41. [PubMed: 14511592]
64.
Salikhov I, Hirata H, Walczak T, Swartz HM. An improved external loop resonator for in vivo L-band EPR spectroscopy. J Magn Reson. 2003;164:54–9. [PubMed: 12932455]
65.
Hirata H, Walczak T, Swartz HM. An improved inductive coupler for suppressing a shift in the resonance frequency of electron paramagnetic resonance resonators. Rev Sci Instrum. 1997;68:3187–91.
66.
Pollock JD, Williams BB, Sidabras JW, Grinberg O, Salikhov I, Lesniewski P, Kmiec M, Swartz HM. Surface loop resonator design for in vivo EPR tooth dosimetry using finite element analysis. Health Phys. 2010;98:339–44. [PMC free article: PMC4086293] [PubMed: 20065703]
67.
Sugawara H, Hirata H, Petryakov S, Lesniewski P, Williams BB, Flood AB, Swartz HM. Design and evaluation of a 1.1-GHz surface coil resonator for electron paramagnetic resonance-based tooth dosimetry. IEEE Trans Biomed Eng. 2014;61:1894–901. [PubMed: 24845300]
68.
Enomoto A, Hirata H. Sequential CW-EPR image acquisition with 760-MHz surface coil array. J Magn Reson. 2011;209:244–9. [PubMed: 21320789]
69.
Enomoto A, Emoto M, Fujii H, Hirata H. Four-channel surface coil array for sequential CW-EPR image acquisition. J Magn Reson. 2013;234:21–9. [PubMed: 23832070]
70.
Enomoto A, Hirata H, Matsumoto S, Saito K, Subramanian S, Krishna MC, Devasahayam N. Four-channel surface coil array for 300-MHz pulsed EPR imaging: proof-of-concept experiments. Magn Reson Med. 2014;71:853–8. [PMC free article: PMC7667490] [PubMed: 23532721]
71.
Enomoto A, Saito K, Subramanian S, Krishna MC, Hirata H, Devasahayam N. Passive decoupling due to low Q-factors of four-channel coils in 300-MHz pulsed EPR imaging. Appl Magn Reson. 2015;46:671–83.
72.
Enomoto A, Hirata H. Parallel image-acquisition in continuous-wave electron paramagnetic resonance imaging with a surface coil array: proof-of-concept experiments. J Magn Reson. 2014;239:29–33. [PubMed: 24374749]
73.
Nilges MJ, Walczak T, Swartz HM. 1 GHz in-vivo EPR spectrometer operating with a surface probe. Phys Med. 1989;5:195–201.
74.
Li H, Hou H, Sucheta A, Williams BB, Lariviere JP, Khan MN, Lesniewski PN, Gallez B, Swartz HM. Implantable resonators—a technique for repeated measurement of oxygen at multiple deep sites with in vivo EPR. In: Takahashi E, Bruley D, editors. Oxygen transport to tissue XXXI. Advances in experimental medicine and biology, vol. 662. Boston, MA: Springer; 2010. [PMC free article: PMC3806631] [PubMed: 20204802]
75.
Hou H, Li H, Dong R, Khan N, Swartz H. Real-time monitoring of ischemic and contralateral brain pO2 during stroke by variable length multisite resonators. Magn Reson Imaging. 2014;32:563–9. [PMC free article: PMC4297608] [PubMed: 24629514]
76.
Khan N, Hou H, Eskey CJ, Moodie K, Gohain S, Du G, Hodge S, Culp WC, Kuppusamy P, Swartz HM. Deep-tissue oxygen monitoring in the brain of rabbits for stroke research. Stroke. 2015;46:e62–6. [PMC free article: PMC4342327] [PubMed: 25613304]
77.
Hou H, Dong R, Li H, Williams B, Lariviere JP, Hekmatyar SK, Kauppinen RA, Khan N, Swartz H. Dynamic changes in oxygenation of intracranial tumor and contralateral brain during tumor growth and carbogen breathing: a multisite EPR oximetry with implantable resonators. J Magn Reson. 2012;214:22–8. [PMC free article: PMC3730127] [PubMed: 22033225]
78.
Hou H, Nemani VK, Du G, Montano R, Song R, Gimi B, Swartz HM, Eastman A, Khan N. Monitoring oxygen levels in orthotopic human glioma xenograft following carbogen inhalation and chemotherapy by implantable resonator-based oximetry. Int J Cancer. 2015;136:1688–96. [PMC free article: PMC4314380] [PubMed: 25111969]
79.
Wolfson H, Ahmad R, Twig Y, Kuppusamy P, Blank A. A miniature electron spin resonance probehead for transcutaneous oxygen monitoring. Appl Magn Reson. 2014;45:955–67.
80.
Miyake M, Liu KJ, Walczak TM, Swartz HM. In vivo EPR dosimetry of accidental exposures to radiation: experimental results indicating the feasibility of practical use in human subjects. Appl Radiat Isot. 2000;52:1031–8. [PubMed: 10836403]
81.
Haga T, Hirata H, Lesniwski P, Rychert KM, Williams BB, Flood AB, Swartz HM. L-band surface-coil resonator with voltage-control impedance-matching for EPR tooth dosimetry. Concepts Magn Reson B. 2013;43B:32–40.
82.
Schreiber W, Petryakov SV, Kmiec MM, Feldman MA, Meaney PM, Wood VA, Boyle HK, Flood AB, Williams BB, Swartz HM. Flexible, wireless, inductively coupled surface coil resonator for EPR tooth dosimetry. Radiat Prot Dosim. 2016;172:87–95. [PMC free article: PMC6287419] [PubMed: 27421470]
83.
Williams BB, Dong R, Nicolalde RJ, Matthews TP, Gladstone DJ, Demidenko E, Zaki BI, Salikhov IK, Lesniewski PN, Swartz HM. Physically-based biodosimetry using in vivo EPR of teeth in patients undergoing total body irradiation. Int J Radiat Biol. 2011;87:766–75. [PMC free article: PMC4086327] [PubMed: 21696339]
84.
Hoshi N, Nikawa Y, Kawai K, Ebisu S. Application of microwaves and millimeter waves for the characterization of teeth for dental diagnosis and treatment. IEEE Trans Microwave Theory Tech. 1998;46:834–8.
85.
Trompier F, Romanyuka A, Kornak L, Calas C, LeBlanc B, Mitchell C, Swartz H, Clairand I. Electron paramagnetic resonance radiation dosimetry in fingernails. Radiat Meas. 2009;44:6–10.
86.
Black PJ, Swarts SG. Ex vivo analysis of irradiated fingernails: chemical yields and properties of radiation-induced and mechanically-induced radicals. Health Phys. 2010;98:301–8. [PMC free article: PMC3684967] [PubMed: 20065698]
87.
Petryakov SV, Schreiber W, Kmiec MM, Williams BB, Swartz HM. Surface dielectric resonators for X-band EPR spectroscopy. Radiat Prot Dosim. 2016;172:127–32. [PMC free article: PMC8444672] [PubMed: 27421472]
88.
Swartz HM, Williams BB, Zaki BI, Hartford AC, Jarvis LA, Chen EY, Comi RJ, Ernstoff MS, Hou H, Khan N, Swarts SG, Flood AB, Kuppusamy P. Clinical EPR: unique opportunities and some challenges. Acad Radiol. 2014;21:197–206. [PMC free article: PMC3921887] [PubMed: 24439333]
89.
Caston RM, Schreiber W, Hou H, Williams BB, Chen EY, Schaner PE, Jarvis LA, Flood AB, Petryakov SV, Kmiec MM, Kuppusamy P, Swartz HM. Development of the implantable resonator system for clinical EPR oximetry. Cell Biochem Biophys. 2017;75:275–83. [PMC free article: PMC5972368] [PubMed: 28687906]
90.
Bilgen M, Elshafiey I, Narayana PA. In vivo magnetic resonance microscopy of rat spinal cord at 7 T using implantable RF coils. Magn Reson Med. 2001;46:1250–3. [PubMed: 11746595]
91.
Rivera DS, Cohen MS, Clark WG, Chu AC, Nunnally RL, Smith J, Mills D, Judy JW. An implantable RF solenoid for magnetic resonance microscopy and microspectroscopy. IEEE Trans Biomed Eng. 2012;59:2118–25. [PMC free article: PMC4497577] [PubMed: 22156945]
Copyright 2020, Springer Nature Switzerland AG.
Bookshelf ID: NBK566443PMID: 33411447DOI: 10.1007/978-3-030-47318-1_10

Views

  • PubReader
  • Print View
  • Cite this Page
  • PDF version of this page (867K)

Similar articles in PubMed

See reviews...See all...

Recent Activity

Your browsing activity is empty.

Activity recording is turned off.

Turn recording back on

See more...